Transcription of ZEISS CONTURA Specifications
1 ZEISS CONTURA Specifications Version: May 2018 ZEISS CONTURA active sensors and accuracyThe CMM Specifications are only valid when using original accessories by ZEISS . The specified parameters are observed in the application of the internal test instructions for acceptance testing and in the use of the released standards in accordance with the ISO 10360 VAST XT goldZEISS VAST XTR goldActive scanning and multipoint sensor. Scanning measuring rate up to 500 measuring force (50-1000 mN) for data VAST XT gold: stylus: max. length = 500 mm, max. weight = 500 g incl. stylus adapter, min. stylus tip diameter = VAST XTR gold: max. length (rigid) = 500 mm, max. length (during rotation) = 350 mm, max. weight = 500 g, including stylus adapter, min. stylus tip diameter = to 7/10/69/12/8 to 9/16/810/12/6 to 10/16/612/18/10 to12/24/10 Length measurement error 1) 2)MPE complies with ISO 10360-2:2009E0 / E15018 C - 22 CE0 / E15018 C - 26 Cin min + + + + + + + + L/350 Repeatability range of E0 MPL complies with ISO 10360-2:2009R0in errorMPE complies with ISO 10360-4:2000 THPin measuring time MPT in s40404040 Form measurement error 3)MPE for roundness complies withISO 12181 (VDI/VDE 2617 sheet )RONt (MZCI) in stylus form probing errorMPE complies with ISO 10360-5:2010 PFTUin form probing errorMPE complies with ISO 10360-5:2010 PFTM 4)in 5) 5)Multi-stylus dimension probing errorMPE complies with ISO 10360-5:2010 PSTM 4)in location probing errorMPL complies with ISO 10360-5.
2 2010 PLTM 4)in 5) 5) ZEISS CONTURA RDS sensors and accuracyThe CMM Specifications are only valid when using original accessories by ZEISS . The specified parameters are observed in the application of the internal test instructions for acceptance testing and in the use of the released standards in accordance with the ISO 10360 RDS-C-CAAD ynamic ZEISS RDS articulating unit for optical and contact and lateral tilt range of 180 , large measuring range, rotation increments of ,CAA correction for automatic qualification of all angular positions for scanning sensors( ZEISS VAST XXT TL3) and multiple-point sensors ( ZEISS XDT TL3). ZEISS VAST XXTZEISS XDT 6) ZEISS VAST XXT TL3 on ZEISS RDS scanning and multiple-point sensorZEISS XDT TL3: multiple-point sensor Scanning measuring rate up to 500 length with module: TL3 = 30-150 mm, maximum stylus weight = 15 gTL3 maximum sensor extension = 100 mm, minimum stylus tip diameter = mm7/7/6 to 7/10/69/12/8 to 9/16/810/12/6 to 10/16/612/18/10 to12/24/10 Length measurement error 2) 7)MPE complies with ISO 10360-2:2009E0 / E4018 C - 22 CE0 / E4018 C - 26 Cin min + + + + + + + + L/200 Repeatability range of E0 MPL complies with ISO 10360-2:2009R0in errorMPE complies with ISO 10360-4.
3 2000 THPin measuring time MPT in s50505050 Form measurement error 3)MPE for roundness complies withISO 12181 (VDI/VDE 2617 sheet )RONt (MZCI)in stylus form probing errorMPE complies with ISO 10360-5:2010 PFTUin ) ZEISS VAST XT gold: acceptance test with 60 mm stylus and 8 mm stylus tip. Also valid for other styli ( 3 x 33 mm, 5 x 50 mm, 8 x 114 mm and 12 x 92 mm were tested).2) Measuring length L in ) Roundness in Scanning Mode for Vscan = 5 mm/s, filter 50 ) Measuring location near the calibration position to document sensor ) Applies to ZEISS VAST XTR ) Only ZEISS CONTURA x700 /X10007) ZEISS VAST XXT: acceptance test with TL3 module; 50 mm stylus and 3 mm stylus tip diameter. E40 acceptance applies only in conjunction with CONTURA direct sensors and accuracy 7)The CMM Specifications are only valid when using original accessories by ZEISS .
4 The specified parameters are observed in the application of the internal test instructions for acceptance testing and in the use of the released standards in accordance with the ISO 10360 VAST XXT 1) ZEISS XDT 7) ZEISS VAST XXT TL3: scanning and multiple-point sensorZEISS XDT TL3: multiple-point sensor(See ZEISS CONTURA RDS sensors and accuracies, ZEISS VAST XXT and ZEISS XDT for accuracies) ZEISS ViScan 3)Optical 2D image sensor with autofocus on ZEISS distance (depending on lens): 75-90 to 7/10/69/12/8 to 9/16/810/12/6 to 10/16/612/18/10 to12/24/10 Length measurement error 2)MPE complies with ISO 10360-7: 2011EU(XY)in m10 4) + L/35010 4) + L/35010 4) + L/35010 4) + L/350 MPE probing error of the image editingsystem as per ISO 10360-7:2011 PFV2 Din m10 4)10 4)10 4)10 4) ZEISS LineScan 3) 5)Optical laser triangulation scanner on ZEISS RDS-C.
5 7/7/6 to 7/10/69/12/8 to 9/16/810/12/6 to 10/16/612/18/10 to12/24/108 mm Measuring range 32 mm Working distanceProbing error 6)MPE complies with ISO 10360-8:2013PF (OT)in on sphere1 Sigmain mm Measuring range 63 mm Working distanceProbing error 6)MPE complies with ISO 10360-8:2013PF (OT)in m12121212 Dispersion on sphere1 Sigmain m444450 mm Measuring range 94 mm Working distanceProbing error 6)MPE complies with ISO 10360-8:2013PF (OT)in m20202020 Dispersion on sphere1 Sigmain m5555100 mm Measuring range 220 mm Working distanceProbing error 6)MPE complies with ISO 10360-8:2013PF (OT)in m50505050 Dispersion on sphere1 Sigmain m12121212 XDT1) Acceptance test with TL3 module; stylus length of 50 mm and stylus tip diameter of 3 ) Measuring length L in ) The use of optical probes requires calibration with contact probe ( ZEISS VAST XXT) Temperature range of 18 C - 26 ) Measured with ZEISS ViScan 1x lens5) Laser class 2M: the accessible laser beam lies in the visible spectral range that is safe for the eye at a short exposure time ( s) as long as the cross section is not reduced by optical instruments ( magnifiers, lens elements, telescope).
6 6) Probing error in the center of the measuring range on suitable sphere (30 mm diameter) with matte surface. P[ :Tr:ODS]. The information on the working distance is based on the center of the measuring ) Only ZEISS CONTURA x700 /X1000 Sensor overviewVAST XT goldVAST XTR goldVAST XXTXDT 5)VAST XXTXDT 5)ViScanLineScanMultipoint Passive scanning Active scanning Optical scanning Rotatable / tiltable Max. stylus length 1)500 mm350/500 mm250 mm 2)150 mm 4)250 mm 2)150 mm 4)Max. stylus weight 1)500 g500 g15 g 2)15 g 4)15 g 2)15 g 4)Navigator Technical featuresLength measuringsystemPhotoelectric reflected light system, m resolutionControllerType: ZEISS C99 LProtection type: IP53 Accessories (optional)Multi-sensor Rack for storage of stylus systemsEnvironmental requirements 3)Relative humidity30 % - 60% (without condensation)Measuring referencetemperatureX700/X900/X1000/X120 0X700/X900/X1000/X1200 (Option HTG)18 C - 22 C18 C - 26 CPer K/dPer K/ K/mFloor vibrationsZEISS CONTURA is equipped with standard vibration damping (limits upon request).
7 Upon request, we can provide assistance for vibration for operationRelative humidity30 - 60% (without condensation)Ambient temperature17 C - 35 CPower ratingZEISS C99L100-240V VAC ~ ( 10 %); 50-60 Hz ( %)Max. power consumption: 800 VATypical power consumption: 200 WCompressed air supplySupply pressure min. 6 bar, max. 8 bar, consumption 50Nl/min for ZEISS CONTURA X900/1200. Max. consumption 120Nl/min for ZEISS CONTURA x700 use of the AirSaver included with delivery ensures that compressed air is not used during ZEISS CONTURA downtimes, thus enabling environmentally friendly quality complies with ISO 8573 Part 1: Class 4, : max. particle size 15 m, max. dirt particle concentration 8 mg/m3 Paragraph : max. compressed air dew point +3 CParagraph : max. oil concentration of 5 mg/m3If the air supply does not comply with the above requirements, an additional air filter unit and, if necessary,a membrane dryer must be inserted in the compressed air 5)RDS1) Depending on the application, limiting the parameters for a stylus system may be ) ZEISS VAST XXT: depending on model.
8 TL 4: 125-250 mm 10 g. TL 3: 30-150 mm 15 ) To ensure specified ) Only TL35) Only for ZEISS CONTURA x700 /X1000 ZEISS VAST XT gold ZEISS VAST XTR ZEISS VAST XXT direct3) ZEISS XDT direct3) ZEISS RDS-CZEISS CONTURA sizesDimensions in mmWeight in kgMeasuring rangeOverall CMM dimensionsWorking range (Max. workpiece size)Table heightAs-sembly spaceTrans-port height 2)CMMMax. workpieceX axisY axisZ axisLengthWidthHeightLengthWidthHeightHe ightHeightHeightHeightG104G105G106B5B6B7 B16B17B18hB23c1T7/7/6700700600 1)1670148127971039910716 1)824850 200220013005602577 4)630 4)1980 4)1270 4)7/10/67001000600 1)1975148127971344910716 1)824850 200220015707302577 4)630 4)1980 4)1540 4)9/12/89001200800 1)22471867 3410 16001227885 1)950850 2001850290012009/16/89001600800 1)2647 1867 3410 2000 1227 885 1)950 850 2001850 3400120010/12/610001200600 1)21801783279715441215716 1)824850 2002200231011502577 4)630 4)1980 4)2300 4)10/16/610001600600 1)25791783279719441215716 1)824850 2002200284015002577 4)630 4)1980 4) 2790 4)12/18/10120018001000 1)2847 2197 36602200 1557 1086 1)1151 700 2002100 4300 1200 12/24/10120024001000 1)3447 2197 3660 2800 1557 1086 1)
9 1151 700 2002100 6050 12001581471791119741391613B18188197 Mounted to CMM ramMeasuring range start ofZEISS VAST XT gold62 Note: the given dimensions and weights are approximate values. Subject to change. Actual appearance of specific sizes may vary from based on DIN 4000-167 B7c1B6G104B23hG106B161) Applies to ZEISS VAST XT gold. The measuring range (G106) and the maximum workpiece height (B18) are reduced by at least 50 mm when other probes are ) Transport height of the secured machine group without pallet or Z mast. When transporting without base, deduct 600 mm from the transport height. Only ZEISS CONTURA 7/7/6, 7/10/6, 10/12/6 und 10/16/63) Only ZEISS CONTURA x700 /10004) CONTURA with reduced base Printed in Germany CZ-IV/2018 Subject to change in design and scope of delivery and as a result of ongoing technical on chlorine-free bleached paper.
10 Carl ZEISS Industrielle Messtechnik GmbHCarl ZEISS Industrielle Messtechnik GmbH73446 Oberkochen / Germany Sales: +49 7364 20-6336 Service: +49 7364 20-6337 Fax: +49 7364 20-3870 Email: ZEISS Industrial Metrology, LLC6250 Sycamore Lane North Maple Grove, MN 55369/USAP hone: +1 763 744-2400 Fax: +1 763 533-0219 Email: descriptionType according to ISO 10360-1:2000 Moving bridge CMMO perating modemotorized / CNCS ensor mountsFixed installationSoftwareZEISS CALYPSO, ZEISS GEAR PRO, ZEISS HOLOST ravel speedMotorizedAxes0 to 70 mm/sCNCV ectormax. 465 mm/sAccelerationVectormax. m/s2 Scanning speed 1)max. 125 mm/sApprovalsRegulationsZEISS CONTURA complies with EC machine directive 2006/42/EC and EMC directive 2014/30 products and packaging returned to us are disposed of in accordance with applicable legal management systemISO 9001:2008,VDA 6, Parts 4, 2.