GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL …

Back to document page

GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIESPRODUCED BYA WORKING GROUP OF THE SCOTTISH QUALITY ASSURANCE SPECIALIST INTEREST GROUP SECOND EDITION FEBRUARY 2004 1ContentsPage NumberGeneral Introduction2Test Methodologies3Section 1- Physical Tests4Introduction41. Non-viable particle counts52. Pressure differentials93. Airflow velocity114. Air change rate135. Operator protection factor156. Filter Integrity Testing16Section 2 - Microbiological Tests17Introduction17General Information- Sample equipment 17- Sampling conditions17- Sample locations18- Incubation conditions18- Results and reading of samples18- Action levels19- Action to be taken in the event of a result being on or over the action level197.

practice. The physical tests performed are reproducible and accurate and can be used to validate the facilities/equipment's performance against accepted criteria e.g. airborne particulate contamination, pressure differentials, clean room air change rates. Testing microbiological parameters as a means of evaluating

  Tests, Contamination, Particulates, Particulate contamination