Oxford Elevate (Rev A) - mpsource.com
Rev AOxford /Hoyer Professional SeriesSERVICE Rev ACONTENTSPages 3 - 4Inspection Criteria of the Oxford /Hoyer ElevatePage 5Calibration of Weigh ScalePage 6Load Testing of the Oxford /Hoyer ElevatePages 7 - 14Service and Maintenance Schedule for the Oxford /Hoyer ElevatePage 15 - 16Fault Finding on the Oxford /Hoyer ElevatePage 17Exploded View of the Oxford /Hoyer ElevatePage 18Parts List of the Oxford /Hoyer ElevatePage 19LOLER (UK only): Thorough Examination ReportPage 20 - 23AppendixPage 28Blank Notes Rev AINSPECTION CRITERIAJoerns Healthcare recommends a thorough inspection and test of the Oxford /Hoyer Elevate and its lifting accessories, slings carried out every six months including a weight test to the full examination and test should be conducted according to the recommendations and procedures below. Joerns Healthcarerecommends authorised service dealers should carry out maintenance, inspection and certified testing hoist is fitted with the optional weigh scales, the weigh scales require an annual calibration as described on page : These recommendations are in compliance with the requirements of 1998 No2307 Health and Safety: The Lifting Operationsand Lifting Equipment Regulations 1998.
3 294000.10040 Rev A INSPECTION CRITERIA Joerns Healthcare recommends a thorough inspection and test of the Oxford/Hoyer Elevate and its lifting accessories, slings etc.