Laboratory Procedure Manual Analyte: Complete Blood Count Matrix: Whole Blood Method: Complete Blood Count with 5-Part Differential Method No.: Revised: as performed by: Contact: Important Information for Users CDC periodically refines these Laboratory methods. It is the responsibility of the use r to contact the person listed on the title page of each write-up before using the analytical method to find out whether any changes have been made and what revisions, if any, have been incorporated. 7-1. Complete Blood Count using HMX. NHANES 2007-2008. Public Release Data Set Information This document details the Lab Protocol for testing the items listed in the following table: Lab Number Analyte SAS Label White bloo d cell co unt (1000. LBXWBCSI cells/uL). LBXLYPCT Lymphocyte (%). LBXMOPCT Monocyte (%). LBXNEPCT Segmented neutrophils (%).
Methodology: The methods used to derive CBC parameters are based on the Coulter® method of counting and sizing, in combination with an automatic diluting and mixing device for sample processing, and a single beam photometer for hemoglobinometry. The WBC differential uses …