NIOSH Manual of Analytical Methods (NMAM), Fourth EditionLEAD by Flame AAS 7082Pb (Pb)CAS:7439-92-1 (Pb)RTECS:OF7525000 (Pb) (PbO)1317-36-8 (PbO)OG1750000 (PbO)METHOD: 7082, Issue 2 EVALUATION: FULLI ssue 1: 15 February 1984 Issue 2: 15 August 1994 OSHA mg/m3 NIOSH:< mg/m3; blood Pb 60 g/100 mg/m3 PROPERTIES:soft metal;d g/cm3; MP Cvalences +2, +4 in salts SYNONYMS:elemental lead and lead compounds except alkyl leadSAMPLINGMEASUREMENTSAMPLER:FILTER( m cellulose ester membrane)FLOW RATE:1 to 4 L/minVOL-MIN: 200 L @ mg/m3 -MAX:1500 LSHIPMENT:routineSAMPLE STABILITY:stableBLANKS:2 to 10 field blanks per setTECHNIQUE:ATOMIC ABSORPTIONSPECTROPHOTOMETER, FLAMEANALYTE:leadASHING:conc.
NOTE 2: The Appendix gives a microwave digestion procedure which may be necessary for complete recovery of lead from some matrices, especially epoxy-based paint. 3. Open the cassette filter holders and transfer the samples and blanks to clean beakers. 4. Add 3 mL conc. HNO 3, and 1 mL 30% H 2O 2 and cover with a watchglass. Start reagent blanks at