And GC/MSSolvent Retention DataSolvent Retention DataColumn:DB-624125-133430 m x mm, mCarrier:Helium, constant pressure at 30 cm/sec (40 C)Oven:40 C for 5 min40-260 C at 10 min260 C for 3 minInjection:Split 1:10, 250 CDetector:FID, 300 CColumn:DB-1125-103430 m x mm, mCarrier:Helium, constant pressure at 30 cm/sec (40 C)Oven:40 C for 5 min40-260 C at 10 /minInjection:Split 1:10, 250 CDetector:FID, 300 CColumn:DB-WAX125-703230 m x mm, mCarrier:Helium, constant pressure at 34 cm/sec (40 C)Oven:40 C for 5 min40-230 C at 10 /min230 C for 7 minInjection:Split 1:10, 250 CDetector:FID, 300 CThese Solvent tables have many uses, not least of which is to determine impurities in bulk solvents . All of the columns wereselected for their capacity, selectivity and :Other manufacturers' look-a-like columns do not have the same selectivity as Agilent J&W GC Columns.
354 www.agilent.com/chem/supplies GC and GC/MS Solvent Retention Data Solvent Retention Data Column: DB-624 125-1334 30 m x 0.53 mm, 3.00 μm Carrier: Helium ...