Model: HDM- 136a - pH Analyzer/Transmitter- Issue: HDM- 136a -0107-R0 Page 1 FEATURES 4-wire pH transmitter with built-in microcomputer: Field installation type pH analyzer / transmitter featuring isolatedoutput, digital indication and rainproof construction. Simplified calibration with standard solution: Data of five different pH standard solutions can be stored in theinternal memory for single-action calibration. Stability judgmentfunction provides accurate calibration with standard solution, freefrom operator error. Automatic determination of electrode quality: The transmitter judges the electrode quality from it s characteristicsduring calibration with standard solutions. Degradation ofelectromotive force at pH7, degradation of electromotive force perpH and other information is displayed in the form of error data of each electrode can be called out to determinethe extent of degradation as required.

Model: HDM-136A - pH Analyzer/Transmitter- Issue: HDM-136A-0107-R0 Page 3 WIRING DIAGRAMS OPTIONAL ACCESSORIES • Sun Shade (hood) Recommended when the instrument is installed at a …