METHOD #: for Approval for NPDES (August, 1991) TITLE: Determination Of Dissolved Hexavalent Chromium In DrinkingWater, Groundwater And Industrial Wastewater Effluents ByIon ChromatographyANALYTE:CAS # Chromium 7440-47-3, and method provides procedures for the determinatbn of dissolved hexavalentchromium in drinking water, groundwater and industrial wastewater Method Detection Limits (MDL, defined in Sectin 3) for the above matricesare listed in Table 1. The MDL obtained by an individual laboratory for aspecific matrix may differ from those listed depending on the nature of thesample and the instrumentation containing high levels of anionic species such as suHate and chloridemay cause column overload.
5.0 Safety 5.1 Hexavalent chromium is toxic and a suspected carcinogen and should be handled with appropriate precautbns (3,4). Extreme care should be exercised