Transcription of Model ZR22G, ZR402G Separate type Zirconia …
1 User sManualModel ZR22G, ZR402 GSeparate type Zirconia oxygen / humidity AnalyzerIM 11M12A01-02 EIM 11M12A01-02E17th EditioniIM 11M12A01-02E17th Edition : Jun. 14, 2021-00u IntroductionThank you for purchasing the ZR22G, ZR402G Separate type Zirconia oxygen / humidity read the following respective documents before installing and using the ZR22G, ZR402G Separate type Zirconia oxygen / humidity related documents are as SpecificationsContentsDocument numberNoteModel ZR22G, ZR402G , and ZR202 GDirect In Situ Zirconia oxygen Analyzers and High Temperature humidity AnalyzersGS 11M12A01-01E* the E in the document number is the language s ManualContentsDocument numberNoteModel ZR22G.
2 ZR402 GSeparate type Zirconia oxygen / humidity AnalyzerIM 11M12A01-02E(This manual) Model ZR22S, ZR402 GSeparate type Explosion-proof Zirconia oxygen AnalyzerIM 11M13A01-02 EWhen you use the ZR22S Explosion-proof detector, read this ZR22A, ZR202 AHeater AssemblyIM 11M12A01-21 EModel EXAxt ZR SeriesHART ProtocolIM 11M12A01-51 EModel AV550G Zirconia oxygen analyzer Averaging ConverterIM 11M12D01-01E* the E in the document number is the language exclusive User s Manual might be attached to the products whose suffix codes or option codes contain the code Z (made to customers specifications).
3 Please read it along with this EXAxt ZR Separate type Zirconia oxygen / humidity analyzer is usually the oxygen analyzer which connected ZR402G converter and ZR22G detector together, but it is to the High Temperature humidity analyzer when the option code /HS (Set for humidity analyzer ) of ZR402G is that, in this manaul, the humidity analyzer refers to the thing which ZR22G combined with ZR402G with /HS .In addition, in this manual, the oxygen analyzer is mainly listed. When there are not mentions such as in the case of humidity analyzer , it becomes same as the oxygen EXAxt ZR Separate type Zirconia oxygen / humidity analyzer has been developed for combustion control in various industrial processes.
4 This analyzer basically consists of a detector and a converter. You can select between several versions, based upon your accessories are also available to improve measurement accuracy and automate calibration. An optimal control system can be realized by adding appropriate instruction manual refers to almost all of the equipment related to the EXAxt ZR. You may skip any section(s) on the equipment which is not included in your the HART communication protocol, refer to IM 11M12A01-51E has been published as Model EXAxt ZR Series HART Protocol.
5 The Integrated type (with sensor and analyzer integrated in one body) is described in IM 11M12A01-04E.<Before using the equipment, please read any descriptions in this manual related to the equipment and system that you have, on appropriate use and operation of the EXAxt ZR.>No. IM 11M12A01-02E 17th Edition : June 2021 (YK)All Rights Reserved Copyright 2000, Yokogawa Electric CorporationiiIM 11M12A01-02E17th Edition : Jun. 14, 2021-00 Models and descriptions in this manual are listed and descriptions in this manualModelProduct NameDescription in this manualSpecificationInstallationOperation MaintenanceCMPLZR22 GGeneral use detector ZR22 GHigh temperature detector ( ) ZR402 GConverter ZO21 RProbe protector ZO21 PHigh temperature probe adapter ZH21 BDust protector (only for humidity analyzer ) ZA8 FFlow setting unit (for manual calibration use) ZR40 HAutomatic Calibration unit -Auxiliary ejector for high temperature use (Part No.)
6 E7046EC, E7046EN) -Calibration gas unit case (Part No. E7044KF) -Check valve (Part No. K9292DN, K9292DS) -Dust filter for the detector (Part No. K9471UA) -Dust guard protector (Part No. K9471UC) ZO21 SStandard gas unit CMPL: Customer Maintenance Parts ListThis manual consists of twelve chapters. Please refer to the reference chapters for installation, operation and of ContentsChapterOutlineRelates toInstallationOperationMaintenance1. OverviewEquipment models and system configuration examplesBCB2. SpecificationsStandard specification, Model code (or part number), dimension drawing for each equipmentABB3.
7 InstallationInstallation method for each equipmentAC4. PipingExamples of piping in three standard system configurationsAC5. WiringWiring procedures such as Power supply wiring , output signal wiring or othersAC6. ComponentsMajor parts and function are described in this manualCBB7. StartupBasic procedure to start operation of EXAxt ZR. Chapter 7 enables you to operate the equipment Detailed Data SettingDetails of key operations and displaysBC9. CalibrationDescribes the calibration procedure required in the course of Other FunctionsOther functions describedBA11.
8 Inspection and MaintenanceHow to conduct maintenance of EXAxt ZR and procedures for replacement of deteriorated partsBA12. TroubleshootingThis chapter describes measures to be taken when an abnormal condition (parts list)User replaceable parts listCBA: Read and completely understand before operating the : Read before operating the equipment, and refer to it whenever : Recommended to read it at least 11M12A01-02E17th Edition : Jun. 14, 2021-00n For the safe use of this equipment WARNINGH andle it with care. Be sure not to accidentally drop it.
9 Handle safely to avoid the power supply cord only after confirming that the supply voltage matches the rating of this equipment. In addition, confirm that the power is switched off when connecting power sample gas is dangerous to people. When removing this equipment from the process line for maintenance or other reasons, protect yourself from potential poisoning by using a protective mask or ventilating the area well. CAUTIONThe cell (sensor) at the tip of the detector is made of ceramic ( Zirconia element). Do not drop the detector or subject it to pressure stress.
10 Do NOT allow the sensor (probe tip) to make contact with anything when installing the detector. Avoid any water dropping directly on the probe (sensor) of the detector when installing it. Check the calibration gas piping before introducing the calibration gas to ensure that there is no leakage of the gas. If there is any leakage of the gas, the moisture drawn from the sample gas may damage the sensor. The detector (especially at the tip) becomes very hot. Be sure to handle it with NOTICEl Specification check When the instrument arrives, unpack the package with care and check that the instrument has not been damaged during transportation.