6100a . 6180A . MODELS Paperless Graphic Recorders Ideal for : specification sheet 21 CFR Part II Applications Furnaces, NADCAP The 6000 Series offer unrivalled input accuracy with a 125ms total sample rate for up to 48 input channels. Input channels are freely Water/waste water configurable to suit your process requirements. Each instrument has an intuitive, touch screen display to enable operators to clearly view Power generation process data in varying formats. All have onboard Flash data storage Sterilizers, autoclaves capability, Ethernet communication and choice of Compact Flash or SD Card. Data is stored in a tamper-resistant binary format that can be used for secure, long term records of your process. The 6000 Series is truly designed for todays networked world and can be accessed via a Features : Local Area Network, dial-up connection, Intranet or Internet.

Serial Communications Option No of Ports 2 Protocol ASCII (typical applications: Input of ASCII string inputs from Barcode readers, Credit card readers etc.)

Tags:

  Sheet, Specification, Issue, 6100a, 6180a specification sheet, 6180a, Issue 2

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of 6100A, 6180A specification sheet, issue 2 - McGoff …

1 6100a . 6180A . MODELS Paperless Graphic Recorders Ideal for : specification sheet 21 CFR Part II Applications Furnaces, NADCAP The 6000 Series offer unrivalled input accuracy with a 125ms total sample rate for up to 48 input channels. Input channels are freely Water/waste water configurable to suit your process requirements. Each instrument has an intuitive, touch screen display to enable operators to clearly view Power generation process data in varying formats. All have onboard Flash data storage Sterilizers, autoclaves capability, Ethernet communication and choice of Compact Flash or SD Card. Data is stored in a tamper-resistant binary format that can be used for secure, long term records of your process. The 6000 Series is truly designed for todays networked world and can be accessed via a Features : Local Area Network, dial-up connection, Intranet or Internet.

2 Colour touchscreen display USB plug & play' Available Features Up to 48 Universal Inputs Up to 96MB non-volatile Flash memory 6100a 6180A . Display 1/4 VGA XGA. 125ms Parallel sampling Channels 18 48. Compact Flash or Secure Relays 16 36. Digital Card Events Inputs 24 (6 per option card). Groups 6 standard (12 optional). Modbus master Auditor Features Lite or Full available Ethernet TCP/IP Virtual channels* 36, 96, 128. Timers Fitted as standard Alarms 4 per channel Batch Optional Bridge- Remote Lite as standard (Full optional). viewing software Screen builder 24 (optional). Security Unlimited unique user names with configurable access permissions and passwords Configuration software Standard Review/Quickchart Lite Standard software Standard views Vertical and horizontal trending, Vertical and Horizontal Bargraphs, Circular Trend and Numeric Values * Virtual channels can be configured as Maths, Totalisers, Counters or Comms EUROTHERM FLEXIBLE SOLUTIONS.

3 Data Logging and Archiving Dynamic Host Configuration Protocol (DHCP). The 6000 Series recorders have internal Flash memory for secure data Dynamic Host Configuration Protocol, the successor to BootP, allows a 6000. storage. They are also able to accept various removable media types Series host to obtain Network parameters, such as IP address, Subnet Mask, (compact flash, SD card or USB memory stick). Data stored within the default gateway and DNS server address dynamically. The implementation internal memory can be archived to the removable media on demand or at of DHCP on the 6000 Series significantly reduces the overhead for preset intervals. The 6000 will give indication of how long its internal maintaining a network of instrumentation. memory and that of the removable media installed will last according to the configuration of the recorder.

4 All 6000s have Ethernet capability. The 6000 can be configured to archive TECHNICAL specification . to the removable media and / or over Ethernet. Archiving files over Recorder Ethernet effectively gives a secure, infinite archiving capacity. Environmental performance Temperature limits Operation: 0 to +50 C. Approximate duration for continuous recording of one Group of six Storage: 20 to 60 C. channels, high compression: Humidity limits Operation: 5% to 80% RH. Storage: 5% to 90% RH. Archive Sample Rate Protection Bezel and display: IP65. Media 1s 5s 10s 30s 60s Sleeve: IP20. 32Mb Internal days days days 113 days 226 days yrs yrs 6100a Portable case option: IP21. Flash (approx. 4 Shock BS EN61010. million samples). Vibration (10 to 150Hz) BSEN60873, Section 9,18. 96Mb Internal days days days 339 days yrs yrs 11 yrs Altitude <2000 metres.

5 Flash (approx. 12. million samples). Approvals 64Mb CF/SD Card days days days 226 days yrs yrs yrs or USB memory Electromagnetic compatibility CE, cUL (EMC). stick (approx. 8 UL file number e57766. million samples). Emissions and immunity BS EN61326. 256Mb CF/SD Card days days 181 days yrs yrs yrs 20 yrs or USB memory Electrical safety stick (approx, 32 (BS EN61010) Installation cat. II; Pollution degree 2. million samples). 1Gb CF/SD Card 88 Days 354 days yrs yrs 19 yrs 58 yrs 116 yrs INSTALLATION CATEGORY II. or USB memory The rate impulse voltage for equipment on nominal 230V mains is 2500V. stick (approx. 125. million samples) POLLUTION DEGREE 2. Ethernet (FTP Server) Infinite Normally, only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation shall be expected Time Synchronisation (SNTP).

6 The 6000 Series support Simple Network Time Protocol which, when Physical enabled, updates the instrument time every 15 minutes from the Panel mounting DIN43700. configured SNTP server. The unit can also act as a Unicast SNTP server on Panel mounting angle 45 . the network, allowing client instruments to synchronise with the 6000 to a 6100a Bezel size: 144 x 144mm. resolution of one millisecond. Panel cutout dimensions: 138 x 138mm (both 0/+1mm). Depth behind bezel rear face: (284 LTC). Weight: 3kg max. (5kg if fitted in portable case). Batch Recording 6180A Bezel size: 292 x 292mm Up to six user-defined fields can be used to enter batch specific data. Panel cutout dimensions: 281 x 281mm (both 0/+1mm). Depth behind bezel rear face: 261mm Field Descriptor Operator entered batch information Weight: 7kg max. up to 20 characters up to 60 characters Operator interface Type Colour TFT LCD with cold cathode The user can choose to log any number of the given fields on start and / or backlight, fitted with resistive, analogue, stop of a batch.

7 The information will appear on the chart as a message and Touch-Panel cannot be separated from the process data to which it relates. Size and resolution Model 6100a : 1/4 VGA (320 x 240 pixels) . Model 6180A : XGA (1024 x 768 pixels) . Auditor Features Power requirements Designed to meet the requirements of the FDA Regulation 21 CFR Part 11 Supply voltage Standard: 85 to 265V ac; 47 to 63Hz or for Electronic Records and Signatures, this software option provides the 110 to 370V dc Low voltage option: 20 to 42V RMS;45 to 400Hz or 6000 Series with additional security such as password ageing, electronic 20 to 54V dc signatures and time stamped audit trail. Power (Max) 60VA (Inrush current 36A). Fuse type None Modbus Master Interrupt protection: Standard: Holdup >200msec, at 240V ac, Allows users to view data from multiple instruments connected either by a with full load Low voltage option: 20msec at 20V dc or RMS, with full load local Network connection using Modbus TCP, or a Serial connection using Modbus RTU, Back-up Battery Type Poly-carbonmonofluoride/lithium (BR2330) Part No.

8 PA261095. Event Input Support time (RTC) 1 year min. with recorder unpowered The Event Input option offers six isolated event input circuits per board Replacement period 3 years fitted. Triggered externally these discrete inputs can be used to initiate Stored data Time; date; values for totalisers, counters internal actions within the 6000 Series Paperless Graphic Recorder. For and timers; batch data; Fvalue, example they could be used to remotely start or stop a Batch. Rolling average, Stopwatch etc. Ethernet communications Type 10/100baseT Ethernet. ASCII Printer Output (Reports) ( ). Fitted as standard the ASCII text printer option provides the 6000 Series Protocols TCP/IP, FTP, DHCP, BOOTP, SNTP, with the ability to generate up to 10 simple reports that can be directed to a MODBUS, SMTP, ICMP. Serial ASCII text printer.

9 Reports, triggered by an event/job can be Cable Type: CAT5. configured to contain parameters such as time and date, batch names, Maximum length: 100 metres process values and user defined messages. Termination: RJ45. Thermocouple data Serial Communications Option Temperature scale ITS 90. No of Ports 2 Bias current nA. Protocol ASCII (typical applications: Input of Cold junction types Off, internal, external, remote ASCII string inputs from Barcode CJ error 1 C max with inst. at 25 C. readers, Credit card readers etc.) CJ rejection ratio 50:1 minimum ASCII printer support Upscale/downscale drive High, low or none selectable for Modbus RTU Master and Slave each thermocouple channel Isolation (dc to 65Hz BS EN61010) Installation category II;. Additional error: C (typ.) if high or low Pollution degree 2. selected Terminals to ground 50V RMS or dc (basic insulation).

10 Types and ranges See Table 3. Transmission standard EIA232 or EIA485 (software selectable). T/C Type Overall range Standard Max linearisation Input Board ( C) error General B 0 to +1820 IEC 0to400 C= C. Input types dc Volts, dc millivolts, 400to1820 C= C. dc milliamps (with shunt), C 0 to +2300 Hoskins C. Thermocouple, 2/3-wire RTD D 0 to +2495 Hoskins C. Contact closure (not Channels 1, E 270 to +1000 IEC C. 7, 13, 19, 25, 31, 37, 43) >60 ms G2 0 to +2315 Hoskins C. Input type mix Freely configurable. J 210 to +1200 IEC C. Maximum number of inputs 6 per board K 270 to +1372 IEC C. A/D conversion method >16 bits, 2nd order delta sigma L 200 to +900 DIN43710:1985 (To IPTS68) C. Input ranges See Table1 and Table 2 below. N 270 to +1300 IEC C. Termination Edge connector / terminal block R 50 to +1768 IEC C. Noise rejection (48 to 62 Hz) Common mode: >140dB (channel to S 50 to +1768 IEC C.)