Transcription of Alinity ci-series System Specifications
1 Alinity ci-series System Specifications Alinity | Clinical Chemistry | Immunoassay | Hematology | Transfusion | Molecular | Point of Care | Professional ServicesAchieve measurably better healthcare performance. TRANSFORMATIONTM FEATUREALINITY cALINITY iALINITY ciDimension (H x W x D) x x x x x x ft2 MethodsPhotometric, PotentiometricChemiluminescencePhotometr ic, Potentiometric, ChemiluminescenceMaximum ThroughputUp to 1350 TPHUp to 200 TPHUp to 1550 TPHT hroughput/ft2Up to 90 TPH/ft2Up to 13 TPH/ft2Up to 62 TPH/ft2 ScalabilityUp to 4 modules controlled by one System Control Module (SCM)*Continuous Access of Reagents, Calibrators, Controls and ConsumablesYe sFlexible Stat OptionsPrioritize single rack as needed or configure multiple fixed positionsSample TypesSerum, plasma, urine, cerebrospinal fluid, hemolysate, whole bloodSerum, plasma, whole blood, urine Serum, plasma, urine, cerebrospinal fluid, hemolysate, whole bloodSample Capacity150150300 Sample Bar Code Types Code 128, Standard Code 39, Interleaved 2 of 5, CodabarSample Result Storage200,000 Dead Volume50 L (sample cup)Sample Volume* 35 L2 200 LAlinity c: 35 LAlinity i.
2 2 200 LSample Probe Carryover parts per million Reagent CapacityUp to 70 refrigerated reagent cartridges onboard plus patented ISE (Na+, K+, and Cl-)Up to 47 refrigerated reagent cartridges onboardUp to 117 refrigerated reagent cartridges onboard plus patented ISE (Na+, K+, and Cl-)Reagent Type100% liquid ready-to-useReagent Onboard Stability*5 60 days15 30 daysFor Alinity c: 5 60 daysFor Alinity i: 15 30 daysAutomated Onboard Calibrators and Controls*Ye sYes (controls only) Alinity c: YesAlinity i: Yes (controls only)Calibration Frequency*1 60 days15 30 daysFor Alinity c: 1 60 daysFor Alinity i: 15 30 daysSample, Clot and Bubble DetectionYe sReagent Pressure MonitoringYe sSample Interference MeasurementYes; hemolysis, icterus, and lipemiaNoYes; hemolysis, icterus, and lipemia (CC only)On Board Maintenance RecordsYe sOnline Error Code HelpYe sHost InterfaceHL7 or ASTMR emote DiagnosticsAbbottLinkWeight1566 lbs1373 lbs2552 lbsElectrical RequirementsSCM: 90 264 V, 16 ampEach Instrument: 180 264 V, 16 ampWater RequirementsAverage: 27 L/hr Max : <30 L/hrAverage: <10 L/hr Max : <30 L/hrAverage: 37 L/hr Max : <60 L/hrHeat Output (processing)Average 2005 BtuAverage 1634 BtuAverage 3639 BtuNoise Level (1 m) Alinity c: dBA Alinity i.
3 DBALaboratory Automation ConnectionACCELERATOR a3600 ACCELERATOR a3600In developmentTPH=tests per hour* Assay dependent Excluding whole blood Maximum of two minutes during the prime of the wash buffer dilution assemblyFEATUREALINITY ccALINITY iiDimension (H x W x D) x x x x ft2 MethodsPhotometric, PotentiometricChemiluminesenceMaximum ThroughputUp to 2700 TPHUp to 400 TPHT hroughput/ft2107 TPH/ft216 TPH/ft2 ScalabilityUp to 4 modules controlled by one System Control Module (SCM)*Continuous Access of Reagents, Calibrators, Controls and ConsumablesYe sFlexible Stat OptionsPrioritize single rack as needed or configure multiple fixed positionsSample TypesSerum, plasma, urine, cerebrospinal fluid, hemolysate, whole bloodSerum, plasma, whole blood, urineSample Capacity300300 Sample Bar Code Types Code 128, Standard Code 39, Interleaved 2 of 5, CodabarSample Result Storage200,000 Dead Volume50 L (sample cup)Sample Volume* 35 L2 200 LSample Probe Carryover parts per million Reagent CapacityUp to 140 refrigerated reagent cartridges onboard plus patented ISE (Na+, K+, and Cl-)Up to 94 refrigerated reagent cartridges onboardReagent Type100% liquid ready-to-useReagent Onboard Stability*5 60 days15 30 daysAutomated Onboard Calibrators and Controls*Ye sYes (controls only)
4 Calibration Frequency*1 60 days15 30 daysSample, Clot and Bubble DetectionYe sReagent Pressure MonitoringYe sSample Interference MeasurementYes; hemolysis, icterus, and lipemia NoOn Board Maintenance RecordsYe sOnline Error Code HelpYe sHost InterfaceHL7 or ASTMR emote DiagnosticsAbbottLinkWeight2746 lbs2356 lbsElectrical RequirementsSCM: 90 264 V, 16 ampEach Instrument: 180 264 V, 16 ampWater RequirementsAverage: 54 L/hrMax : <60 L/hrAverage: 20 L/hr Max : 60 L/hrHeat Output (processing)Average 4010 BtuAverage 3268 BtuNoise Level (1 m) Alinity c: dBAAlinity i: dBALaboratory Automation ConnectionACCELERATOR a3600 TPH=tests per hour* Assay dependent Excluding whole blood Maximum of two minutes during the prime of the wash buffer dilution assemblyAlinity, AbbottLink and Choose Transformation are trademarks of Abbott Laboratories in various jurisdictions.
5 2018 Abbott Laboratories. ADD-00063137 Achieve measurably better healthcare performance. TRANSFORMATIONTM