Atrium EXPRESS

Atrium EXPRESS . Dry Seal Chest Drain K. L. Set Up M Atrium EXPRESS Chest Drain is manufactured by Atrium Medical Corporation, 5 Wentworth Drive, Hudson NH 03051 603-880-1433. Protected by the following international and patent(s): Y CAUTION: Federal (US) law restricts this device to sale by or on the order of a physician. Maquet is a Trademark and/or Registered Trademark of Maquet GMBH, its subsidiaries, and/or affiliates in the United States F Step 1. and/or other countries. Maquet is registered in Patent and Trademark Office. EXPRESS and Atrium are Trademarks and/or Registered Trademarks of Atrium Medical Corp., its subsidiaries, and/or affiliates in the United States and/or other countries. EXPRESS and Atrium are registered in Patent and Trademark Office. All other company or product names are the trademarks or registered trademarks of their respective holders.

Atrium EXPRESS Dry Seal Chest Drain Verifying suction operation via the suction monitor bellows The bellows located in the suction monitor will expand only when suction is …

Tags:

  Express, Atrium express, Atrium

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Atrium EXPRESS

1 Atrium EXPRESS . Dry Seal Chest Drain K. L. Set Up M Atrium EXPRESS Chest Drain is manufactured by Atrium Medical Corporation, 5 Wentworth Drive, Hudson NH 03051 603-880-1433. Protected by the following international and patent(s): Y CAUTION: Federal (US) law restricts this device to sale by or on the order of a physician. Maquet is a Trademark and/or Registered Trademark of Maquet GMBH, its subsidiaries, and/or affiliates in the United States F Step 1. and/or other countries. Maquet is registered in Patent and Trademark Office. EXPRESS and Atrium are Trademarks and/or Registered Trademarks of Atrium Medical Corp., its subsidiaries, and/or affiliates in the United States and/or other countries. EXPRESS and Atrium are registered in Patent and Trademark Office. All other company or product names are the trademarks or registered trademarks of their respective holders.

2 H J Connect patient tube to patient I A Dry Suction Regulator G Connect chest drain to patient prior to initiating suction. B Vacuum Indicator C Air Leak Monitor Step 2. A B Connect suction to chest drain D D Collection Chamber E Suction Monitor Bellows Attach suction line to suction port on top of chest drain. F Easy-to-Grip Handle Step 3. G Positive Pressure Turn suction source on E Release Valve Increase suction source vacuum to -80 mmHg or higher. H Suction Port Suction regulator is preset to -20 cmH2O. Adjust as required. I Filtered Manual High Step 4. Negativity Vent Fill air leak monitor to fill line J Multi-position Hangers Fill air leak monitor to the fill line by syringe (no K In-line Connector needle) with 30 ml of sterile water or sterile saline L Patient Tube C via the needleless Luer port located on the back M Patient Pressure Float Ball of the drain.

3 For models available with sterile N Swing Out Floor Stand fluid, twist top off syringe and insert tip into Pre-packaged needleless Luer port. Depress syringe plunger into water syringe Luer port and squeeze contents into air leak moni- M N tor until fluid reaches fill line. What to check during system operation Verifying suction operation via the suction Manual high negativity vent monitor bellows To manually vent the system of high negative The bellows located in the suction monitor will pressure, depress the manual vent located on top expand only when suction is operating. The of the drain until bubbling occurs in the air leak monitor bellows will not expand when suction is monitor. Do not use manual vent when suction not operating or disconnected. The s mark is not operating or when the patient is on Do not use when suction is allows quick and easy confir mation of vacuum gravity drainage.

4 Not operating. operation over a wide range of continuously Bellows must be expanded to s Placement of unit Copyright 2015 Atrium Medical Corp. or its affiliates. All rights not expressly granted are reserved. 03/15 Part No. 010396 Rev AA. Refer to Instructions for Use for current indications, warnings, contraindications, and precautions. adjustable suction control settings. mark or beyond for a 20 cmH2O. Always place chest drain below the patient's chest in an upright or higher regulator setting. Changing suction pressures position. To avoid accidental knockover hang the system bedside with Suction regulator is preset to -20 cmH2O and can the hangers provided. be adjusted from -10 cmH2O to -40 cmH2O. To Vacuum indicator change suction setting, adjust rotary suction When vacuum is present in the chest drain, a symbol will remain visible in regulator dial located on the side of the drain.

5 The vacuum indicator window. When vacuum Dial down to lower suction pressure and dial up to is not present (atmospheric pressure) no sym- increase suction pressure. To lower regulator bol will appear. All patient tube connections setting from a higher level (-40 cmH2O) to a lower and the vacuum indicator window should be level (-20 cmH2O), adjust regulator down to lower setting and then checked regularly for vacuum confirmation. Vacuum is Present. Vacuum is Not temporarily depress the manual high negativity vent located on top of the Present. drain to reduce excess vacuum. Positive pressure relief protection The positive pressure relief valve is located on top of the drain and opens to Increase vacuum source when bellows is not expanded to s mark release accu mulated positive pressure.

6 Do not obstruct the positive pres- If the bellows is observed sure relief valve. to be expanded, but less than the s mark, the Graduated air leak monitor vacuum source pressure Fill the graduated air leak monitor with 30 ml sterile water or sterile saline to Not enough Normal suction the fill line via the needleless Luer port located on the back of the drain. must be increased to vacu um for operation for 80 mmHg or higher. -20 cmH2O or -20 cmH O or 2. When air bubbles are observed going from right to left, this will confirm a higher suction higher. Increase suction source patient air leak. Air leak bubbling can range from 1 (low) to 5 (high). Air control setting. to -80 mmHg or higher. bubbles create an easy-to-follow air leak pattern for monitoring patient air Automatic high negativity release leak trends.

7 As required, additional water may be added via the needleless The EXPRESS incorporates an automatic high negativity relief valve. This Luer port located on the back of the drain filtered valve activates automatically to limit system pressure to Recording drainage volume approximately -50 cmH2O. The collection chamber incorporates a writing surface with easy-to-read fluid level graduations. Please refer to individual product inserts for specific model graduations. Frequently asked questions What should I do if the chest drain gets knocked over? What does it mean when the vacuum indicator is visible? We recommend that the drain be placed back into the upright position, however The indicates that there is vacuum present in the system (negative pressure, it will continue operating if knocked over.)

8 -20 cmH2O). Is it required to fill the air leak monitor? When will I see a rise in the air leak monitor column? The air leak monitor should be filled for confirmation and detection of air leaks. A rise in the air leak monitor column will only be seen if there is an increase in negative pressure on the patient side. When changing suction pressure from a Have a question or need help in a hurry? higher to lower level, depress the manual high negativity vent to reduce excess Call Maquet toll free at 1-800-528-7486. vacuum to the lower prescribed level. Maquet Medical Systems USA | 45 Barbour Pond Drive, Wayne, NJ 07470 | MCV00033024 REVA.