Blanketrol II Operation Manual - Frank's Hospital Workshop

I0Do not overfill. Overfilling may result in overflow when the water in theblanket drains back into the system when the system is turned not operate without water to avoid damage to internal not use alcohol. Alcohol may cause blanket of 7. If the de-ionized water is acidic, it will cause a batteryeffect and the copper refrigeration line will begin to deteriorate and causea leak in the refrigeration not use De-ionized Water. The majority of de-ionizers do not maintaina neutral 0 Use distilled/sterile water to use distilled/sterile water mayresult in poor performance and damage to the Blanketrol resulting in no therapy to the patient.

operatinq manual all precautions and warninss prior to use. It is recommended this manual be reviewed at least semi-annually as a refresher to safe operation and application.

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Blanketrol II Operation Manual - Frank's Hospital Workshop

1 I0Do not overfill. Overfilling may result in overflow when the water in theblanket drains back into the system when the system is turned not operate without water to avoid damage to internal not use alcohol. Alcohol may cause blanket of 7. If the de-ionized water is acidic, it will cause a batteryeffect and the copper refrigeration line will begin to deteriorate and causea leak in the refrigeration not use De-ionized Water. The majority of de-ionizers do not maintaina neutral 0 Use distilled/sterile water to use distilled/sterile water mayresult in poor performance and damage to the Blanketrol resulting in no therapy to the patient.

2 Follow instructions fordesired mode to resume to resume therapy could result in serious injury or interruption will cause the Blanketrol II to revert to CHECK0Do not use the Blanketrol II system in the presence of of explosion can the physician if the patient s temperature is not responding properlyor does not reach the temperature prescribed in the prescribed time or ifthere is a change in the prescribed temperature range. Failure to informthe physician of the deviation may result in injury to the @ IIOPERATION MANUALThe operator must read and understand the Operation Manual in its entirety prior tooperating the Sub-Zero Products, inc.

3 , reserves the right to make equipment changes andimprovements which may not be reflected in this law restricts this device to use only on the order of a @ Copyright 1982, Cincinnati Sub-Zero Products, Inc. All rights series, or equivalent, probes on CSZ equipment. Failureto do this will cause incorrect temperature is a Registered Trademark of Cincinnati Sub-Zero Products, physician s order is required for setting blanket temperature and use ofequipment. At least every 20 minutes, or as directed by physician, checkpatient s temperature and skin condition of areas in contact with blanket;also, check blanket water temperature.

4 Pediatric, temperature-sensitive,patients with vascular disease, and operating room patients should bechecked more frequently. Notify the physician promptly of any change inorder to avoid serious injury or method of temperature control provided by all hyper-hypothermiaunits presents the danger of heating or cooling body tissues, particularlythe skin, to a point where they are injured, , burns or frostbite,respectively. Depending on the extent and severity of a burn, very seriousand even fatal complications may excessive and/or prolonged tissue pressure and shearing forces,especially over boney prominences, to prevent skin damage that not place additional heat sources between the patient and damage may area between the patient and the blanket should be kept dry to avoidinjury to only YSI.

5 III.. 43 Status Display Messages .. 41 Concluding Hyper-Hypothermia Treatment .. 40 Operating the Blanketrol II Unit in Monitor Only Mode.. the Blanketrol II Unit in Manual Control Modewith the Addition of the Patient Probe .. the Blanketrol II Unit in Manual Control Mode .. the Blanketrol II Unit in Automatic Control Mode .. the System Components .. - . 33 ..31 SECTION 3 OPERATING THE Blanketrol II UNIT ..30 Patient Preparation and Bedside and Patient Related for Protecting the Reservoir from Bacteria andfor Decontamination Time Set-up/System Test and Arranging the a System Test the of the Blanketrol II 2 GENERAL PREPARATION OF THE Blanketrol II SYSTEM.

6 19 . Safety Control Description of the Blanketrol II Control Description of the Blanketrol II System- Rear Features - Right Side Features - Front Features .. Description of the Blanketrol II Unit- Safety Description of this Description of the Blanketrol II ApplicationsPace 1. INTRODUCTION.. 1 TABLE OF 3-lBLANKETROL II System Equipment - Control l-5 Temperature Safety Limits- Rear l-4 Blanketrol II Unit ..7 Figure l-3 Blanketrol II Unit - Right Side l-2 Blanketrol II Unit - Front ViewPaqe l-lBLANKETROL II Unit .. ILLUSTRATIONS operatinq Manual allprecautions and warninss prior to is recommended this Manual be reviewedat least semi-annually as a refresher to safe Operation and DESCRIPTION OF THIS MANUALThis Manual describes the Operation , maintenance, and service of the CSZBLANKETROL II hyper-hypothermia system.

7 Section One describes the physical andfunctional characteristics of the Blanketrol II System. Section Two describes howto prepare the Blanketrol II unit for general use. Section Three describes how tooperate the unit in the Manual Control Mode, Automatic Control Mode, and MonitorOnly Manual is prepared for professional personnel who use the equipment for patientcare. It is designed to be stored with the Blanketrol II unit and to be readilyavailable for reference when operating the Blanketrol II DESCRIPTION OF THE Blanketrol II SYSTEMThe CSZ Blanketrol II, Model 222R Hyper-Hypothermia System is used to eitherlower or to raise a patient s temperature and/or maintain a desired patient temperaturethrough conductive heat CSZ Blanketrol II unit is composed of aheater, a compressor, a circulating pump, and a microprocessor board.

8 This unitrequires no field adiustments or calibrations in order to maintain the precisemeasurement of temperature and temperature safetv ,nurses, technicians and operators must read and understand this svstem, such as hyper-hypothermia system, a thorough knowledge, understanding of the system, the correctapplication and operating use of the system is person who isresponsible for the use or the direction of the use of the MANDALSECTION SAFETY PRECAUTIONSTo provide the patient maximum safety during the use of the Blanketrol II 22ZR Operation HDDEL BLANJ5 ZTFtOL II, INTRODUCTION* The recommended blanket(s) for use is described in Section (I-5.)

9 APPLICATIONSThe Blanketrol II unit is used primarily in Hospital Intensive and Coronary CareUnits, in Operating,Recovery and Emergency Rooms, in Burn Units, and onMedical/Surgical floors. This hyper-hypothermia system can be used with adult andpediatric patients to produce normothermia by lowering a patient s elevatedtemperature or raising a patient s sub-normal temperature. It can also be utilized tomaintain normal body temperature (normothermia) during surgical , this system can be used to produce moderate to profound hypothermia forsuch procedures as amputations, cardiopulmonary by-pass surgery, vascular surgery,and intracranial surgery.

10 Medically, this system can be used to decrease the rate ofcirculation, to reduce intracranial pressure, to control cerebral edema, and to reduceoxygen requirements. This system is also used in the treatment of burns, shock,cardiac arrest, and gastrointestinal hemorrhage.(cont d)Water is heated or cooled and pumped from the unit to a blanket. The blanket* restsunder and/or on top of the patient and is designed so that the water circulates throughthe blanket and returns back to the cooled water is circulated through the blanket, the desired effect is to reduce thepatient s temperature. If warmed water is circulated through the blanket, the desiredeffect is to elevate the patient s Blanketrol II unit can be set so that it operates based on the temperature ofthe circulating water ( Manual Control) or it can be set so that it operates based on thetemperature of the patient (Automatic Control).