IEC 60601 -1 Ed Label-Manual Checklist, Rev. 4 (2015-04-28) Copyright 2015 Page 1 of 16. This checklist covers the IEC 60601 -1, Edition requirements for the labeling and the accompanying documents (IFU) of Medical Electrical Equipment. It also includes information and interpretations for the clause requirements, as applicable. This is a FREE download from and will be updated often, to provide additional guidance. Be sure that you have the latest revision. The following is tool for evaluating medical equipment labeling and user manuals to the requirements of the 60601 -1 standards. This does not replace the standards, and a purchased copy of the IEC and National standards should also be used. Verdict Info Verify Clause Requirement Comment Doc Test - 7 ME Equipment Identification, Marking, and Documents: - General: - Usability of the identification, marking and documents: (Removed). Test Legibility of Markings: See Appended Table Test of markings required in - Observer: visual acuity of 0 on the log Minimum Angle of Resolution (log MAR) scale or 6/6 (20/20) and is able to read N6 of Jaeger test card in normal room lighting condition (~500lx).

Medical Equipment Compliance Associates, LLC Free Download: http://60601-1.com/download Created by Brian Biersach, Anna HerzfeldFree Download

Tags:

  60601

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of FREE download 60601-1.com/download

1 IEC 60601 -1 Ed Label-Manual Checklist, Rev. 4 (2015-04-28) Copyright 2015 Page 1 of 16. This checklist covers the IEC 60601 -1, Edition requirements for the labeling and the accompanying documents (IFU) of Medical Electrical Equipment. It also includes information and interpretations for the clause requirements, as applicable. This is a FREE download from and will be updated often, to provide additional guidance. Be sure that you have the latest revision. The following is tool for evaluating medical equipment labeling and user manuals to the requirements of the 60601 -1 standards. This does not replace the standards, and a purchased copy of the IEC and National standards should also be used. Verdict Info Verify Clause Requirement Comment Doc Test - 7 ME Equipment Identification, Marking, and Documents: - General: - Usability of the identification, marking and documents: (Removed). Test Legibility of Markings: See Appended Table Test of markings required in - Observer: visual acuity of 0 on the log Minimum Angle of Resolution (log MAR) scale or 6/6 (20/20) and is able to read N6 of Jaeger test card in normal room lighting condition (~500lx).

2 Marking read at ambient luminance (100 lx to 1,500 lx), positioned for intended position of the operator, or at any point within the base of a 30 cone (if not defined), at a distance of 1 m. Test Durability of Markings: See appended Tables Required markings can be removed only with a tool or by appreciable force, are durable, and remain clearly legible during expected service life of me equipment in normal use. Marking rubbed by hand with a cloth rag soaked with each of the following, for 15 sec.: Distilled water, ethanol (96%) C2H6O, and Isopropyl alcohol C3H8O. - Marking on The Outside of ME Equipment or Parts: Verify Minimum Requirements For Marking On ME Equipment: See attached copy of Markings If size or the nature of enclosure does not allow affixation of all required markings, at least - (manufacturer, model, serial number, date of manufacture, software rev. identifier), - (external power supply), - (Class II), (Applied Parts), - (Physiological effects), as applicable, shall be affixed.

3 Verify Remaining markings fully recorded in accompanying documents Markings only in IFU: Verify Markings applied to individual packaging when impractical to apply to me equipment Verify - Single use item marked "Single Use Only" /. Do Not Reuse /. Symbol 28 of Table D1. - Identification: See attached copy of Markings ME Equipment marked with the following. Verify - Name or trademark - Contact information of the manufacturer Verify - Model or type reference Verify - Serial number or lot or batch identifier (readable or identification by technology RFID, etc.). Verify - Date of manufacture or use by date (if applicable). - Unless misidentification does not present unacceptable risk (misidentification could lead to a hazardous situation), detachable components marked with: Verify - Name or trademark of the manufacturer Verify - Model or type reference Verify If not marked, the risk management file includes an assessment of the risks relating to RM reference to specific risks (ISO 14971 Clauses).

4 Misidentification of all detachable parts : : : 5: : Verify - Software identified with a unique identifier (not required on outside of equipment, can be only Unique identifier: available to designated people). Verify Consult Accompanying Documents: - Table D1, Symbol 11 MAY be used, to advise operator to consult acompanying documents: - Table D2, safety sign 10 MUST be used if risk management uses the acompanying documents to reduce risk to an acceptable level, but only when the manufacturer uses the IFU as a risk control measure for a specific risk Medical Equipment Compliance Associates, LLC Free download : Created by Brian Biersach, Anna Herzfeld MECA IEC 60601 -1 Ed Label-Manual Checklist, Rev. 3 (2015-01-28) Copyright 2015 Page 2 of 16 Verdict Info Verify Clause Requirement Comment Doc Test Verify Accessories: Accessories inspected: - Marked with name or trademark - Contact information of their manufacturer Verify - Model or type reference Verify - Serial number or lot or batch identifier Verify - Date of manufacture or use by date (if applicable).

5 Verify - Markings applied to individual packaging, when not practical to apply to accessories - ME Equipment intended to receive power from other equipment: Provided with one of the following: Verify - Name or trademark of the manufacturer of the other electrical equipment and type reference, marked adjacent to the relevant connection point (or). Verify - Table D2, safety sign 10, adjacent to the relevant connection point (and). - Listing of the required details in the IFU. (or). Verify - Special connector style used that is not commonly available on the market (and). - Listing of the required details in the IFU. - Connection to the Supply Mains: ME Equipment marked with the following information - The following markings are provided on outside of part containing supply mains connection and, adjacent to the connection point. Verify Permanently installed me equipment: - Nominal supply voltage or range marked inside or outside of me equipment Verify All other equipment: Rated Voltage (V-V): - Rated supply voltage(s), or voltage range(s) with a hyphen (-) between min and max voltages Verify - Rated supply voltages, or multiple rated supply voltage ranges separated by slash (/) Rated Voltage (V/V): Verify - Nature of supply (number of phases, except for single phase) Phases AC / DC.

6 - Type of current (AC, DC). (or) Use symbols below Verify Table D1, Symbols 1-5 may be used to identify this: Symbols provided: Verify - Rated supply frequency, frequencies, or range, in hertz Hz: Verify - Table D1, Symbol 9 provided for class II ME Equipment Symbol provided: Verify Electrical Input Power From The Supply Mains: A / VA: - Rated input in amps or volt-amps, when power factor is or less Verify - Rated input in amps, volt-amps, or watts, when power factor exceeds A / VA / W: Verify For equipment with multiple voltage ranges: A / VA / W: If the range(s) are greater than 10 % of the mean value of given range, - The rated input power is given for the upper and lower limits of the range(s). Verify For equipment with multiple voltage ranges: A / VA / W: If the range(s) are NOT greater than 10 % of the mean value of given range, - The mean input power of the input range is given Verify If the ratings include both long-time and momentary current or volt-amp ratings: Long-time VA: Momentary VA: - Markings and IFU provide both long-time and most relevant momentary volt-amp ratings Verify - Marked input of me equipment provided with means for connection of supply conductors A / VA / W: of other electrical equipment, includes rated and marked output of such means.

7 Medical Equipment Compliance Associates, LLC Free download : Created by Brian Biersach, Anna Herzfeld MECA IEC 60601 -1 Ed Label-Manual Checklist, Rev. 3 (2015-01-28) Copyright 2015 Page 3 of 16 Verdict Info Verify Clause Requirement Comment Doc Test - Output Connectors Info Mains power Output: For integrated MSOs (Multiple Socket-Outlets = power strips), see b). - Other Power Sources: Power output connectors marked with the following. (except MSOs or connectors specified for specific parts or accessories). Verify - Rated voltage V, A / VA / W: - Rated current or power (when applicable). Verify - Output frequency (when applicable) Hz: Verify IP Classification: IPXX: - ME Equipment or its parts marked with the IP code, per IEC 60529. (marking optional for me equipment or parts rated IPX0). - Applied Parts: Markings provided: Degrees of protection against electric shock marked with relevant symbols for all applied parts Verify - Type B applied parts with Table D1, symbol 19: Applied Part: Verify - Type BF applied parts with Table D1, symbol 20: Applied Part: Verify - Type CF applied parts with Table D1, symbol 21: Applied Part: Verify - Defibrillation-proof applied parts marked with Table D1, symbols 25-27: Applied Part: Verify Proper symbol marked adjacent to or on connector for applied part Marking location.

8 -If no connector, then marked on applied part -If connector used for multiple applied parts with different ratings, marked on applied part -If isolation for BF or CF is not provided in the equipment, but in the applied part, marked on the applied part Verify - Table D2, Safety sign 2 placed near connector if part of defib-proof protection is in patient cable Relevant connector: Verify - IFU indicates that the protection of ME Equipment against effects of a cardiac defibrillator discharge Explanation in IFU: depends on use of proper cables, as applicable. Verify Mode of operation: ME Equipment suitable for continuous operation Verify If NOT continuous use, duty cycle appropriately marked to provide maximum on and off time. Duty Cycle: Verify - X-Ray systems marked as long time operation or momentary operation (NFPA 70) long time operation / momentary (US) operation Verify Fuses: Markings provided adjacent to accessible fuse-holder Doc - Fuse type Type: Doc - Voltage rating V, A: - Current rating Doc - Operating speed (letter or color code) Fast / Slow, A breaking capacity: - Breaking capacity Medical Equipment Compliance Associates, LLC Free download : Created by Brian Biersach, Anna Herzfeld MECA IEC 60601 -1 Ed Label-Manual Checklist, Rev.

9 3 (2015-01-28) Copyright 2015 Page 4 of 16 Verdict Info Verify Clause Requirement Comment Doc Test Verify Physiological Effects (Safety Sign And Warning Statements): Physiological effects: - ME Equipment producing physiological effects, not obvious to the operator, and can cause harm to the patient or operator provides suitable safety sign in a prominent location. Verify Nature of hazard and precautions for avoiding or minimizing the associated risk described in IFU. RM reference to specific risks (ISO 14971 Clauses). (Risk management to address risk of harm) : : : 5: : Verify High Voltage Terminal Devices: When provided on the outside of ME Equipment, accessible without the use of a tool, - Marked with Table D1, symbol 24. Verify Cooling Conditions: Cooling requirements: - Requirements for cooling provisions marked, if applicable Verify Protective Packaging: Special handling instructions: - Packaging marked with special handling instructions for transport and/or storage, if applicable Verify - Permissible environmental conditions marked on outside of packaging Environmental conditions: Verify When premature unpacking of me equipment could result in an unacceptable risk, Safety sign provided: - Packaging marked with a suitable safety sign Verify Risk management file includes the assessment to determine risk of premature unpacking of the ME RM reference to specific risks (ISO 14971 Clauses).

10 Equipment or its parts , that could result in an unacceptable risk. : : : 5: : : Verify Packaging of sterile ME Equipment or accessories, Sterile, Method: - Marked sterile and indication of the method of sterilization Verify External Pressure Source: Max supply pressure: Marked on me equipment adjacent to each input connector, - Rated maximum supply pressure from an external source Verify - Rated flow rate required to meet basic safety and essential performance Verify Functional Earth Terminals: Symbol provided: - Marked with Table D1, Symbol 7. Verify Removable Protective Means: Mark provided: - Marked to indicate the necessity for replacement when the function is no longer needed Verify Mass of Mobile Equipment: Equipment mass in kg: - Marked with its mass, including its safe working load in kilograms (Marked in a way that's obvious that it applies to the entire mobile ME Equipment, with maximum safe working load, and separate from part load ratings).