Example: stock market

Mark V ProVis - Edomed

Mark V ProVis SPECIFICATIONSM edrad reserves the right to change product designs and specifications at any time in the continuing effort to improvetheir products. The following specifications refer to new units operating at room temperature (approximately 70 degrees For 21 degrees C).INJECTOR HEADLoad SpeedForward or reverse 5-10 Heater37 C (98 F) nominalSyringesDisposable: 60 ml, 150 ml or 200 mlAutomaticAutomatically sets higher than theMechanical Stop selected volume before each ml with a 60 ml ml with a 150 ml ml with a 200 ml syringeWEIGHTP edestal System (total) 213 lbs (97 kg)Injector Head lbs (9,8 kg)Pedestal System (without head) lbs (87,2 kg)Control Room Console (excluding head) 45 lbs (20,4 kg)Remote Control Panel (without cable) lbs (4,8 kg)POWER REQUIREMENTSA vailable:95-105 VAC, 50 or 60 Hz, 8 A typical 10 amp max105-120 VAC, 50 or 60 Hz, 8 A typical 10 amp max120-125 VAC, 50 or 60 Hz, 8 A typical 10 amp max190-210 VAC, 50 or 60 Hz, 4 A typical 5 amp max210-240 VAC, 50 or 60 Hz, 4 A typical 5 amp max240-250 VAC, 50 or 60 Hz, 4 A typical 5 amp max(Typically defined at 15 ml/s with pressure of 600 psi)INJECTORFlow Rate Syringe SizeFlow Rate60 ml to 10 ml/s, ml/min, ml/hr increments10 to 20 ml/s, in 1 ml increments10

Mark V ProVis ™ SPECIFICATIONS Medrad® reserves the right to change product designs and specifications at any time in the continuing effort to improve their products. The following specifications refer to new units operating at room temperature (approximately 70 degrees F

Tags:

  Marks, Mark v provis, Provis

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of Mark V ProVis - Edomed

1 Mark V ProVis SPECIFICATIONSM edrad reserves the right to change product designs and specifications at any time in the continuing effort to improvetheir products. The following specifications refer to new units operating at room temperature (approximately 70 degrees For 21 degrees C).INJECTOR HEADLoad SpeedForward or reverse 5-10 Heater37 C (98 F) nominalSyringesDisposable: 60 ml, 150 ml or 200 mlAutomaticAutomatically sets higher than theMechanical Stop selected volume before each ml with a 60 ml ml with a 150 ml ml with a 200 ml syringeWEIGHTP edestal System (total) 213 lbs (97 kg)Injector Head lbs (9,8 kg)Pedestal System (without head) lbs (87,2 kg)Control Room Console (excluding head) 45 lbs (20,4 kg)Remote Control Panel (without cable) lbs (4,8 kg)POWER REQUIREMENTSA vailable:95-105 VAC, 50 or 60 Hz, 8 A typical 10 amp max105-120 VAC, 50 or 60 Hz, 8 A typical 10 amp max120-125 VAC, 50 or 60 Hz, 8 A typical 10 amp max190-210 VAC, 50 or 60 Hz, 4 A typical 5 amp max210-240 VAC, 50 or 60 Hz, 4 A typical 5 amp max240-250 VAC, 50 or 60 Hz, 4 A typical 5 amp max(Typically defined at 15 ml/s with pressure of 600 psi)INJECTORFlow Rate Syringe SizeFlow Rate60 ml to 10 ml/s, ml/min, ml/hr increments10 to 20 ml/s, in 1 ml increments10 to 59 ml/min, ml/hr in 1 mlincrements150 ml to 10 ml/s, ml/min, ml/hr in200 ml increments10 to 50 ml/s in 1 ml increments10 to 59 ml/min, ml/hr in 1 mlincrementsAccuracy 5% and.

2 25 ml/s and ml/min 10% and .25 ml/hrRise/Fall TimeFlow Rate to secs in s incrementsVolume Maximum programmable Volume is limited only bysyringe capacity. Minimum programmable Volume:60 ml syringe ml minimum, 150 ml and 200 ml syringe ml ml increments with either syringe Limit In ml/s Flow scale, Pressure can be set to thefollowing limits:60 ml syringe150 ml syringe 200 ml syringe100-1100 psi100-1200 psi100-1000 psi6-75 ATU (bar) 6-82 ATU (bar) 6-68 ATU (bar)7-77 kg/sq cm7-84 kg/sq cm7-70 kg/sq cm689-7584 kPa689-8274 kPa689-6895 kPaPressure can be selected in increments of 1 forany scale (1 psi, 1 ATU (bar), etc.)In ml/min or ml/hr Flow scales, the maximumselectable Pressure is:700 psi48 atu (bar)49 kg/sq. cm4826 KpaDelay Timer seconds in increments of secondX-Ray Control Pair of normally-open isolated contacts rated toswitch 10 A maximum at 28 VDCP rogram Up to 49 Single-Level injections, or 33 Multiple-Control Level injectionsLevel Control Up to 4 levels (Flow Rate and Volume changes)per injection (single Flow scale only).

3 ELECTRICAL LEAKAGEE arth<100 MA at 120 VAC<300 MA at 240 VACC hassis <100 MAPatient <10 MAProtection Per EN 60601-1, designated as Class 1 Against Electrical Shock equipment with CF applied partsEMI/RFI Classified as Group 1, Class B equipmentper EN 60601-1-2 CORD LENGTHSP ower Cord Rack Mount 11 ft (3,3 m)Pedestals 15 ft (4,6 m)Film Changer Extension 25 ft (7,6 m)Cable RMC 100 IAuxillary Handswitch and Cable 25 ft (7,6 m)KMA 550 IRHandswitch with Coiled 25 ft (7,6 m)Cable RPA 550 IRInjector Head Cable 7 ft (2,13 m) pedestal ft (3,05 m) Rack Mountstd. in Americas/Asia15 ft (4,6 m) std. inEurope and Japan90 21 12 0010"(25,4 cm) "(128,8 cm)58"(147,3cm)55"(139,7cm) "(128cm) " (62,2 cm) " (67,6 cm)60"(152,4 cm)180 205 "(66,8 cm) "( cm) "( cm) "( cm) "( cm) "( cm) "( cm)DIMENSIONS12"( cm) "( cm)To " (43 cm) "(32 cm) "(27 cm)19" (48 cm)2 in.

4 (5,08 cm) in.(123,19 cm) in.(67,95 cm) in.(90,42 cm) Injector Head PedestalKMA 320 RT Table BracketKMA 350 Accomodates rails 7/8 1-1/4 wide ( cm cm)HEAD MOUNTING SYSTEMS * 50 ft. ( m) and 75 ft. ( m) Remote Panel Cables are also available** 60 ft. ( m) and 75 ft. ( m) Injector Head Extension Cables are also availableRemote Control PanelPower CordElectronics BoxRemote Panel Cable (100 ft. [ m])*RPC 100P-100 Injector Head Extension Cable (100 ft. [ m])**KMA 950-S100 ITable BracketKMA 350 Height Range: 12"-17" ( cm - cm)TV26" (66 cm)USE 12-3/8" BOLTSTO CEILING SUPPORTSTRUCTURE(BY ENGINEER OF RECORD)TV26" (66 cm)TV26" (66 cm) " (63 cm)TV26" (66 cm) WT. =138 LBS(63 Kg) "(110 cm)TV26" (66 cm)TVHEAD MOUNTING " (81 cm)or 19" (48 cm)*46" (117cm)or 25" (64 cm)* "6,4 cm18 cm15 cm7"6" "80 " 4 cm360 360 360 " - 9 cm high pinwith cover45 50 "6,4 "11,4 cm28"71 "80 cm Ceiling Mounted CounterpoiseOCS 115A Weight: 27 kg (60 lbs) "179 cmRequired for fullyextended "80 cm45 50 "3,8 "80 cm22"56 cmWall57"145 "6,4 "11,4 cm28"71 cm80"203 "9 cm Wall Mounted CounterpoiseOCS 125A Weight: 18 kg (40 lbs)All weights are without injector head.

5 Support Arm OnlyOCS 135 AWeight: 16 kg (35 lbs)Used when an existing ceilingcolumn is in place. Mobile Ceiling Mounted CounterpoiseOCS 155 AWeight: 44 kg (97 lbs)NOTES: 1. Forces are determined per California Building Code, Section 1630 Aand Table acceleration (VH) = acceleration (VV)= Forces are maximums, and occur when equipment is moved to themost eccentric Provide ceiling structure designed to support weights and forcesshown. (By engineer of record for building).*with optional short ceiling mountMedrad, Inc. Corporate OfficeOne Medrad DriveIndianola, PA 15051-0780 : (412) 767-2400 FAX: (412) 767-4128 Web Site SPEC USA 200112 Medrad reserves the right to modify the specifications and features described herein, ordiscontinue manufacture of the product described at any time without prior notice orobligation. Please contact your authorized Medrad representative for the most only source for Medrad Engineering, MedradQuality, Medrad Service and Medrad is dedicated to continually improving the quality of all ourproducts and services such that our customers satisfaction, loyalty, and respect areunsurpassed.

6 It is our policy to clearly understand and agree upon the valid require-ments of the work we perform for our customers, both internal and external, and topursue 100% conformance to those time, every Life Medrad, Inc. is EN-ISO 9001/EN 46001 CertifiedMedrad, Quality for Life, Q and design are federally registered trademarks and Mark VProVis is a trademark of Medrad, Inc. in 1999 Medrad, Inc. All Rights Reserv