MDR Conformity Assessment Routes - BSI Group

MDR Conformity Assessment Routes Notified Body Assessments Class Is/Im/Ir devices 2 Class IIa devices 4 Class IIb Annex VIII rule 12 devices 8 Class IIb implantable Well-Established Technologies (WET) 10 Class IIb non-implantable non rule 12 devices (non WET) 10 Class IIb implantable devices (excluding WET) 14 Class III non-implantable devices 16 Class III implantable devices 18 Custom-made Class III implantable devices 20 Custom-made devices (excluding custom-made Class III implantable devices) 22 Class I devices (excluding Class Is/Im/Ir devices) 23 ContentsDISCLAIMERS:Information presented in the Conformity Assessment flow charts and tables below is based on our current understanding of the MDR requirements at the time of publishing this document; subject to tables do not cover assessments under the Conformity Routes Annex X (Type Examination) and Annex XI Part B (Product Verification) which may require additional tests or examinations of the tables present a generalization of the requirements based on the classification of devices and

Annex IX* QMS Chapters I, III Annex XI* – Part A Production Quality Assurance Class Is, Im, Ir+ Annex II and III Technical Documentation Declaration of conformity (Annex IV) CE Marking (Annex V) CE 0086 or 2797 + r (Class I re-usable Class I surgical instruments) * Limited to sterility, metrology or re-use aspects as applicable

Tags:

  Aspects

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of MDR Conformity Assessment Routes - BSI Group

1 MDR Conformity Assessment Routes Notified Body Assessments Class Is/Im/Ir devices 2 Class IIa devices 4 Class IIb Annex VIII rule 12 devices 8 Class IIb implantable Well-Established Technologies (WET) 10 Class IIb non-implantable non rule 12 devices (non WET) 10 Class IIb implantable devices (excluding WET) 14 Class III non-implantable devices 16 Class III implantable devices 18 Custom-made Class III implantable devices 20 Custom-made devices (excluding custom-made Class III implantable devices) 22 Class I devices (excluding Class Is/Im/Ir devices) 23 ContentsDISCLAIMERS:Information presented in the Conformity Assessment flow charts and tables below is based on our current understanding of the MDR requirements at the time of publishing this document.

2 Subject to tables do not cover assessments under the Conformity Routes Annex X (Type Examination) and Annex XI Part B (Product Verification) which may require additional tests or examinations of the tables present a generalization of the requirements based on the classification of devices and some exceptions may Is/Im/Ir devicesAnnex IX* QMS Chapters I, IIIA nnex XI* Part A Production Quality AssuranceClass Is, Im, Ir+Annex II and III Technical DocumentationDeclaration of Conformity (Annex IV)CE Marking (Annex V) CE 0086 or 2797+ Class Ir (Class I re-usable surgical instruments)* Limited to sterility, metrology or re-use aspects as applicable3 Class Is/Im/Ir devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecertYesYesMicrobiology AuditsYes*N/AYes*N/AYes*N/ATechnical Documentation AssessmentN/AN/AN/AN/AN/AN/AClinical Evaluation Consultation Procedure (Article 54)N/AN/AN/AN/AN/AN/AConsultations (Rule 14, Rule 18, Rule 21)N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32)N/AN/AN/AN/AN/AN/AClinical Evaluation Report updatesUpdated as per Manufacturer s clinical evaluation planPost Market Clinical Follow-Up Update Report (Article 61)Updated as per Manufacturer s PMS, PMCF plans.

3 NB QMS audits to verify implementation of the plan by sampling complaints, vigilance information Safety Update Report (Article 86)N/AN/AN/AN/AN/AUnannounced Audits (BSI policy as of Feb 2019)At least once every 5 years*if sterile or re-usable surgical instrumentsCLASS Is/Im/Ir DEVICES4 Annex XI Part B Production VerificationAnnex IX QMS Chapters I, IIIA nnex IX Chapter II Technical Documentation Assessed per device categoryAnnex II and Annex III Technical Documentation Assessed per device categoryClass IIa devicesCE Marking (Annex V) CE 0086 or 2797 Declaration of Conformity (Annex IV)Class IIaAnnex XI Part A Production Quality Assurance5 Class IIa non-implantable devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecertYesYesMicrobiology AuditsYes (if sterile)N/AYes (if sterile)N/AYes (if sterile)N/ATechnical Documentation AssessmentSample per category of devicesAs per the Technical Documentation Sampling PlanClinical Evaluation Consultation Procedure (Article 54)N/AN/AN/AN/AN/AN/AConsultations (Rule 14, Rule 18, Rule 21)N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32)N/AN/AN/AN/AN/AN/AClinical Evaluation Report updatesUpdated as per Manufacturer s clinical evaluation plan.

4 NB to review as per Technical Documentation Sampling PlanPost Market Clinical Follow-Up Update Report (Article 61)Updated as per Manufacturer s PMS, PMCF plans; NB to review as per Technical Documentation Sampling PlanPeriodic Safety Update Report (Article 86)PSUR update required at least once every 2 years; NB to review as per Technical Documentation Sampling PlanUnannounced Audits (BSI policy as of Feb 2019)At least once every 5 yearsContinued on page 6 CLASS IIa NON-IMPLANTABLE DEVICESC lass IIa implantable devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecertYesYesMicrobiology AuditsYes (if sterile)N/AYes (if sterile)N/AYes (if sterile)N/ATechnical Documentation AssessmentSample per category of devicesAs per the Technical Documentation Sampling PlanClinical Evaluation Consultation Procedure (Article 54)N/AN/AN/AN/AN/AN/AConsultations (Rule 14, Rule 18, Rule 21)N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32)YesUpdated at least annually if indicated.

5 NB to review as per Technical Documentation Sampling Plan or at the time of PSUR assessmentsClinical Evaluation Report updatesUpdated as per Manufacturer s clinical evaluation plan; NB to review updates as per Technical Documentation Sampling Plan or at the time of PSUR assessmentsPost Market Clinical Follow-Up Update Report (Article 61)Updated at least annually; NB to review as per Technical Documentation Sampling Plan or at the time of PSUR assessmentsPeriodic Safety Update Report (Article 86)Updated when necessary and at least every two years; submitted to NB via EUDAMED for NB review Unannounced Audits (BSI policy as of Feb 2019)At least once every 3 yearsClass IIa devices continued6 CLASS IIa IMPLANTABLE DEVICES7 Our website offers a wealth of useful resources including white papers, guidance documents and find out more, Class IIb Annex VIII Rule 12 devices Annex XI Part B Production VerificationAnnex IX QMSC hapters I, IIIA nnex IX Chapter II Technical Documentation Assessed per generic device groupClinical Evaluation Consultation Procedure Annex IX Sec 5 / Annex X Sec 6 Declaration of Conformity (Annex IV)Class IIb(Annex VIII Rule 12 devices)Annex XI Part A Production Quality AssuranceAnnex XType Examination8CE Marking (Annex V)

6 CE 0086 or 27979 Annex VIII Rule 12 devices All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body. Class IIb Annex VIII Rule 12 devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecert YesYesMicrobiology AuditsYes (if sterile)N/AYes (if sterile)N/AYes (if sterile)N/ATechnical Documentation AssessmentSample per Generic Device Group As per the Technical Documentation Sampling Plan Clinical Evaluation Consultation Procedure (Article 54) Yes, but exemptions may apply as per Article Maybe required if any modifications to the device adversely affect the risk-benefit ratio Consultations (Rule 14, Rule 18, Rule 21) N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32)

7 N/AN/AN/AN/AN/AN/AClinical Evaluation Report Updates Updated as per Manufacturer s clinical evaluation plan; NB to review updates as per Technical Documentation Sampling PlanPost Market Clinical Follow-Up Update Report (Article 61)Updated as per Manufacturer s PMCF plan; NB to review updates as per Technical Documentation Sampling PlanPeriodic Safety Update Report (Article 86)Updated at least annually; NB to review updates as per Technical Documentation Sampling Plan Unannounced Audits (BSI policy as of Feb 2019)At least once every 5 years CLASS IIb ANNEX VIII RULE 12 DEVICESC lass IIb implantable WET Class IIb non-implantable, non rule 12, non WETA nnex XI Part B Production VerificationAnnex IX QMS Chapters I, IIIA nnex IX Chapter II Technical Documentation Assessed per generic device groupCE Marking (Annex V) CE 0086 or 2797 Declaration of Conformity (Annex IV)Class IIbimplantable WET + Class IIbnon-implantable, non rule 12 non-WETA nnex XI Part A Production Quality AssuranceAnnex XType Examination10 Well-Established Technologies (WET)

8 - sutures, staples, dental fillings and braces, tooth crowns, screws, wedges, plates, wires, pins, clips & connectors as per Article 52 of IIb implantable WET devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecertYesYesMicrobiology AuditsYes (if sterile)N/AYes (if sterile)N/AYes (if sterile)N/ATechnical Documentation Assessment Sample per Generic Device Group As per the Technical Documentation Sampling Plan Clinical Evaluation Consultation Procedure (Article 54) N/AN/AN/AN/AN/AN/AConsultations (Rule 14, Rule 18, Rule 21) N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32) YesUpdated at least annually if indicated ; NB to review updates as per Technical Documentation Sampling Plan or at the time of PSUR assessmentsClinical Evaluation Report updatesUpdated as per Manufacturer s clinical evaluation plan; NB to review as per Technical Documentation Sampling PlanPost Market Clinical Follow-Up Update Report (Article 61)Updated at least annually; NB to review updates as per Technical Documentation Sampling Plan or at the time of PSUR assessmentsPeriodic Safety Update Report (Article 86)Updated at least annually.

9 Submitted to NB via EUDAMED for NB review (assuming WET devices are implantable devices)Unannounced Audits (BSI policy as of Feb 2019)At least once every 3 years 11 Continued on page 12 CLASS IIb IMPLANTABLE WET12 Class IIb non-implantable, non-WET, non-Rule 12 devicesInitial Conformity AssessmentSURVEILLANCEY1Y2Y3Y4Y5 QMS AuditsYesYesYesRecertYesYesMicrobiology AuditsYes (if sterile)N/AYes (if sterile)N/AYes (if sterile)N/ATechnical Documentation Assessment Sample per Generic Device Group As per the Technical Documentation Sampling Plan Clinical Evaluation Consultation Procedure (Article 54) N/AN/AN/AN/AN/AN/AConsultations (Rule 14, Rule 18, Rule 21) N/AN/AN/AN/AN/AN/ASummary of Safety and Clinical Performance (Article 32)N/AN/AN/AN/AN/AN/AClinical Evaluation Report updatesUpdated as per Manufacturer s clinical evaluation plan; NB to review updates as per Technical Documentation Sampling PlanPost Market Clinical Follow-Up Update Report (Article 61)Updated as per Manufacturer s PMCF plan; NB to review updates as per Technical Documentation Sampling PlanPeriodic Safety Update Report (Article 86)Updated at least annually.

10 NB to review updates as per Technical Documentation Sampling PlanUnannounced Audits (BSI policy as of Feb 2019)At least once every 5 yearsCLASS IIb NON-IMPLANTABLE, NON WET, NON RULE 12 BSI CE-Excellence Programmes are designed for medical device manufacturers wanting to get their products to European markets efficiently and safely. Technical Documentation Reviews for CE Marking Standard DedicatedDon t more information on our CE-Excellence services call BSI on+44 345 080 9000 or visit The CE-Standard review service allows you to work closely with your assigned BSI Product Expert on your product certification. These reviews are conducted remotely, with communication between you and your BSI Product Expert via phone and email, as CE-Dedicated review service allows you to book your technical documentation review in advance.