Example: dental hygienist

Oper ating instructions K- Control 4960, 4965, 49 70.

Operating instructionsK- Control 4960, 4965, 4970. Always on the safe :Kaltenbach & Voigt GmbH Bismarckring 39D-88400 BiberachVertrieb/distribution:KaVo Dental GmbHBismarckring 39 D-88400 BiberachTel.: 0 73 51 / 56-0 Fax: 0 73 51 / 56 -14 88A1 User information .. Meaning of the pictograms .. Important information .. Precautions .. Purpose and potential uses .. Technical data .. Combination Controller - handpiece ..4A2 Scope of delivery/Accessories ..5A3 Electrical connection ..5A4 Location ..6A5 Mounting and connection .. Mounting 4960 .. Mounting 4965 / 4970 .. Connection ..8A6 Controls and functional elements ..9A7 Preparations for commencing operation ..10A8 Operation .. Starting K- Control knee Control unit Starting K- Control Benchtop Control unit 4965.

Oper ating instructions K- Control 4960, 4965, 49 70. Alw ays on the safe side.

Tags:

  Control, Instructions, Rope, Giant, 9460, Oper ating instructions k control 4960

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Oper ating instructions K- Control 4960, 4965, 49 70.

1 Operating instructionsK- Control 4960, 4965, 4970. Always on the safe :Kaltenbach & Voigt GmbH Bismarckring 39D-88400 BiberachVertrieb/distribution:KaVo Dental GmbHBismarckring 39 D-88400 BiberachTel.: 0 73 51 / 56-0 Fax: 0 73 51 / 56 -14 88A1 User information .. Meaning of the pictograms .. Important information .. Precautions .. Purpose and potential uses .. Technical data .. Combination Controller - handpiece ..4A2 Scope of delivery/Accessories ..5A3 Electrical connection ..5A4 Location ..6A5 Mounting and connection .. Mounting 4960 .. Mounting 4965 / 4970 .. Connection ..8A6 Controls and functional elements ..9A7 Preparations for commencing operation ..10A8 Operation .. Starting K- Control knee Control unit Starting K- Control Benchtop Control unit 4965.

2 Starting a K- Control installation with the type 4970 Foot Control unit..11A9 Operating faults ..12 Guarantee conditions ..14 Spare Declaration of Control 4960, 4965, Control 4960, 4965, Important informationThe instructions for use should beread by the user before starting up theunit for the first time, in order to avoidincorrect operation and other damage. Ifother language versions are required, pleaserequest these from your responsible KaVoagent. Duplication and distribution of theinstructions for use (IU) require KaVo'sprior technical data, information and proper-ties of the product described in the IU cor-respond to the state on going to and improvements to theproduct as a result of new technical devel-opments are does not imply any right to retrofittingof existing assumes no responsibility for damagearising through.

3 External influences (poor quality of themedia or inadequate installation) use of incorrect information improper use improperly performed and maintenance work - apart fromthe activities described in these instructionsfor use - may be performed only by quali-fied technical the event of modifications by third par-ties, the approvals become null and recommends using only originalspare parts for operation and for note that the EC Directive onwaste electrical and electronic equip-ment applies to this product. Within Europetherefore, this product must undergo specialdisposal. For more detailed informationabout this, please contact KaVo or yourspecialist dental PrecautionsSafe operation and protection of the unit areensured only through proper use in accor-dance with the instructions for use andusing the tools approved for the following should also be observed: the work safety regulations, the accident prevention regulations.

4 !Each time before switching on, check theset speed.! Observe the permissible maximumspeed and maximum pressure of the tools(according to tool manufacturer'sinstructions). Purpose and potential usesK- Control -installation is versatile and verysuitable for working on crowns and Technical dataDimensions and weightK- Control Knee-Operated Control Unittype 4960 Width:95 mmDepth:280 mmHeight:235 kgK- Control Benchtop Control Unittype 4965 Width:95 mmDepth:280 mmHeight: 235 mmWeight:approx. 3 kgK- Control Foot Control Unittype 4960 Width:245 mmDepth:275 mmHeight: 125 kgRated voltageLine voltage fluctuations < 10 % Voltage range:100 / 120 / 230 V 50 / 60 HzRated powerK- Control WA1 User Meaning of the pictogramsSituations where failure to follow theinstructions may lead to danger,damage to material or operating information for operatorand modeAutomatic sequenceClose, screw in, fasten, , release, loosen+ more, higher- less, lower Continuous operationTime, time sequenceDisconnect mains plugIDevice ON, turned onODevice OFF, unplug from the powersupplyK- Control 4960, 4965, rangeRight-handed rotation.

5 1 000 - 35 000 min-1(K 5 Handpiece)1 000 - 25 000 min-1(K 9 Handpiece)1 000 - 40 000 min-1(K11 Handpiece)1 000 - 40 000 min-1(K10 Handpiece)1 000 - 50 000 min-1(K12 Handpiece)5 000 - 60 000 min-1(SF-Handpiece)1000 - 50 000 min-1(K-POWER grip)1000 - 50 000 min-1(K-ERGO grip)Left-handed rotation, limited to about 5,000min-1 Intermittent service2 min / on8 min / offPollution level2 Overvoltage categoryllAmblent conditions:Max. elevation 2000 m above sea level Permissible in inner rooms Permissible ambient temperature range 5 C - 40 C Permissible up to a max. relative humidityof 80%We reserve the right to make any and storage conditionsTemperature range: -20 C to +70 CRelative humidity: 5% to 95% (non-condensing)Air pressure: 700 hPa to 1060 hPaPrior to start-up, very cold productsmust be heated to a temperature of20 to 25 C.

6 Avoid plate Provides information on:@Manufacturer2 Order number (REF number)3GS-Mark4 For information about proper disposal,see of manufacture - Serial number7CE-Mark according to 89/336 EWG8 VDE-Mark9 Mode: intermittent service: 2 min, 8 min pause0 Unit type 4620 496503 nmaxx1000/minKcontrol03 nmaxx1000/minKcontrol 4960control30000/minmax 4970@426057839K- Control 4960, 4965, Combination Controller - hand-pieceIn the combination with K- Control controlunits of types 4960, 4965 and 4970, thefollowing handpieces can be operated:K 5 Handpiece4910K 9 Handpiece950/955K 9 Handpiece4930K 9 Compact motor970K 9 Cutter spindle960K 10 Handpiece4950 (with Adapter )K 11 Handpiece4990K 12 Handpiece4940SF Handpiece4005 (with supply cable )K-POWER grip4941K-ERGO grip4944 The safty of the K- Control controlunits with handpieces can beguaranteed only with the handpiece/controlunit combinations approved by KaVo.

7 496503 nmaxx1000/minKcontrolcontrol30000/minmax 497003 nmaxx1000/minKcontrol 4960K-POWER grip K12 K5 K11 K9 SF K-ERGO grip K- Control 4960, 4965, Scope of delivery/AccessoriesK- Control Knee-operated Control unit 4960orK- Control Benchtop, Control unit 4965optionalFoot switch 4620 Foot Control unit 4970A3 Electrical connectionElectrical connectionCheck that the local mains voltage andmains frequency conincide with the specifi-cations on the rating plate and the fuseholder Adapt if adaptation to 100 V/120 V or230 V. By removing the fuse holder @, the sleeve2with contacts (including fuses) can beremoved. The adaptation is achieved byappropriately rotating the sleeve untilthe required voltage, 100 V/120 V or 230 V,appears in the inspection window of fuseholder power max.

8 220 :Use mains plug as disconnection.@230120100 496003 nmaxx1000/minKcontrol 4965 462003 nmaxx1000/minKcontrol 4970control30000/minmax @6K- Control 4960, 4965, LocationThe K- Control Knee Control unit, type4960, is made to be attached to the side ofa storage cabinet ( ).The K- Control Benchtop Control unit, type4965, is made to be set on a workbenchand connected underbench with Footswitch type 4620 ( ).The K- Control Foot Control unit, type 4970, can be set unsecured on the floor ( ). 496003 nmaxx1000/minKcontrol 496503 nmaxx1000/minKcontrol 4620 4970control30000/minmax7K- Control 4960, 4965, Mounting and Mounting 4960 The K- Control Knee Control unit, type4960, is made to be fastened to the side ofa storage cabinet on a suspenslonrail 1(drilling template see page 20).

9 Fix in place at the front with screwsand insert mains plug into a ground-ed mains socket which is easy to reach.(Similarly if using an extension line)Demounting 2,312mm03 nmaxx1000/minKcontrol 3,503 nmaxx1000/minKcontrol1page 20K- Control 4960, 4965, Mounting 4965 / 4970 Set K- Control Benchtop Control unit, type4965 on a workbenchSet K- Control Foot Control unit, type 4970,on the floor under a in place at the front with screwand insert mains plug into agrounded mains socket which is easy toreach. (Similarly if using an extension line) ConnectionPlug in Handpiece and foot switch andscrew finger in Handpiece and screw finger nmaxx1000/minKcontrolcontrol30000/minmax 4965 4970 4960 4965 4970K- Control 4960, 4965, Controls and functional elements1 Digital speed and malfunction indicator2 Speed limitation preselector3 Direction switch (counterclockwise running limited to max.)

10 5000 min-1)4 Mains switch "0"/"1"5 Knee switch6 Speed preselector (continuous)7 Suspension rail 8 Fuse holder with voltage adjustment9 Motor sockets (4-pole) Starter socket (5-pole) (Foot switch) Foot switch Type 4620wFoot lever Control cable jack (connection to thedust extractor)03 nmaxx1000/minKcontrol03 nmaxx1000/minKcontrolcontrol30000/minmax # @ # fi \| @ # @ | 10A7 Preparations for commencingoperationOn K- Control Control units 4960 and 4965, push mains switch to the "0" OFF the K- Control Foot Control unit,type 4970, to "0" OFF by letting back thefoot lever. Set sliding switch 2or button2for speed limitation to max. 30,000min-1 .Set direction switch # to clockwise (yellow LED does not light up) or R .For counterclockwise , the handpiecespeed is limited to about 5000 not operate or lay down yourMotorhandpiece, unless an instru-ment or the test pin is clamped in thechuck.


Related search queries