Transcription of Rosemount 3144P Temperature Transmitter - …
1 Manual 00809-0100-4021, Rev EBMay 2007 This product is a core component of the PlantWeb digital plant 3144P Temperature TransmitterReference Manual 00809-0100-4021, Rev EBMay 2007 Rosemount 3144P Temperature TransmitterRosemount 3144P Temperature Transmitter may be protected by one or more Patents Pending. Other foreign patents this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this the United States, Rosemount Inc. has two toll-free assistance numbers:Customer CentralTechnical support, quoting, and order-related (7:00 am to 7:00 pm CST)North American Response CenterEquipment service (24 hours)International(952)-906-8888 The products described in this document are NOT designed for nuclear-qualified applications.
2 Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate information on Rosemount nuclear-qualified products, contact your local Emerson Process Management Sales Manual 00809-0100-4021, Rev EBMay 2007 TOC-1 Rosemount 3144 PSECTION 1 IntroductionOverview.. 1-1 Manual.. 1-1 Transmitter .. 1-2 Considerations .. 1-2 General .. 1-2 Electrical .. 1-2 Environmental .. 1-3 Moist or Corrosive Environments .. 1-3 Mounting .. 1-4 Software Compatibility.. 1-4 Return of Materials .. 1-53144P Revisions.. 1-6 SECTION 2 InstallationSafety Messages .. 2-1 Commissioning .. 2-2 Setting the Loop to Manual .. 2-2 Set the Switches .. 2-3 Mounting.. 2-4 Installation.
3 2-6 Typical North American Installation.. 2-6 Typical European Installation .. 2-7In Conjunction with a Rosemount 333 HART Tri-Loop (HART / 4 20 mA only) .. 2-8 LCD Display.. 2-9 Multichannel Installation (HART / 4 20 mA only) .. 2-10 Wiring .. 2-10 HART / 4 20 mA .. 2-10 Foundation Fieldbus .. 2-13 Sensor Connections .. 2-14 Power Supply .. 2-14 Surges/Transients .. 2-15 Grounding .. 2-15 SECTION 3 HART CommissioningOverview.. 3-1 Safety Messages .. 3-1375 Field Communicator.. 3-2 Updating the HART Communication Software .. 3-2 Menu Tree .. 3-3 Fast Key Sequences .. 3-4 AMS .. 3-5 Apply AMS Changes .. 3-5 Review Configuration Data.. 3-5 Check Output .. 3-6 Configuration .. 3-6 Device Output Configuration.
4 3-14 Device Information .. 3-18 Measurement Filtering .. 3-19 Diagnostics and Service .. 3-21 Transmitter Diagnostics Logging .. 3-23 Table of ContentsReference Manual00809-0100-4021, Rev EBMay 2007 Rosemount 3144 PTOC-2 Multidrop Communication .. 3-23 Use with the HART Tri-Loop .. 3-24 Calibration .. 3-27 Trim the Transmitter .. 3-27 Sensor Input Trim.. 3-28 Transmitter -Sensor Matching .. 3-31D/A Output Trim or Scaled Output Trim .. 3-32 Output Trim.. 3-32 Scaled Output Trim.. 3-33 Troubleshooting.. 3-34 HART / 4 20 mA .. 3-34 LCD Display .. 3-37 AMS Software (HART / 4-20 mA only) .. 3-38 AMS Screens .. 3-40 Spare Parts .. 3-41 SECTION 4 Foundation Fieldbus ConfigurationOverview .. 4-1 Safety Messages .. 4-1 General Block Information.
5 4-2 Device Description .. 4-2 Node Address.. 4-2 Modes.. 4-2 Link Active Scheduler.. 4-3 Block Instantiation .. 4-3 Capabilities.. 4-4 Foundation fieldbus function blocks.. 4-4 Resource Block .. 4-5 FEATURES and FEATURES_SEL .. 4-5 PlantWeb Alerts.. 4-7 Recommended Actions for PlantWeb Alerts .. 4-9 Resource Block Diagnostics.. 4-9 Sensor Transducer Block .. 4-10 Damping .. 4-10 Sensor Transducer Block Diagnostics .. 4-10 LCD Transducer Block.. 4-11 Custom Meter Configuration.. 4-12 Self Test Procedure for the LCD .. 4-13 LCD Transducer Block Diagnostics .. 4-13 Analog Input (AI) .. 4-14 Simulation.. 4-14 Configure the AI block .. 4-14 Filtering.. 4-17 Process Alarms .. 4-17 Status .. 4-18 Advanced Features .. 4-18 Analog Input Diagnostics.
6 4-19 Operation.. 4-20 Overview.. 4-20 Trim the Transmitter .. 4-20 Statistical Process Monitoring (SPM) .. 4-22 SPM Configuration .. 4-23 Troubleshooting Guides.. 4-25 Foundation Fieldbus.. 4-28 LCD Display .. 4-29 Reference Manual 00809-0100-4021, Rev EBMay 2007 TOC-3 Rosemount 3144 PSECTION 5 MaintenanceSafety Messages .. 5-1 Maintenance .. 5-2 Test Terminal (HART / 4 20 mA only) .. 5-2 Sensor Checkout .. 5-2 Electronics Housing .. 5-2 SECTION 6 Certified Safety Instrumented System (Safety-Certified)Safety Messages .. 6-1 Certification .. 6-13144P Safety-Certified Identification .. 6-1 Installation .. 6-1 Commissioning .. 6-2 Configuration .. 6-2 Operation and Maintenance .. 6-3 Proof Test.. 6-3 Inspection .. 6-4 Specifications.
7 6-4 Failure Rate Data .. 6-4 Product Life .. 6-4 Spare Parts .. 6-4 SECTION 7 Prior Use (PU) Safety Instrumented SystemOverview .. 7-1 Safe Failure Fraction .. 7-2 Installation .. 7-2 Switches .. 7-2 Changing Switch Position.. 7-2 Proof Test.. 7-4 APPENDIX AReference DataHART and Foundation Fieldbus Specifications ..A-1 Functional Specifications ..A-1 Physical Specifications..A-1 Performance Specifications ..A-2 HART / 4 20 mA Specifications ..A-6 Foundation Fieldbus Specifications..A-8 Dimensional Drawings..A-10 Ordering Information ..A-14 APPENDIX BProduct CertificationsRosemount 3144P With HART / 4 20 mA..B-1 Approved Manufacturing Locations ..B-1 European Union Directive Information ..B-1 Hazardous Locations Installations.
8 B-1 Rosemount 3144P With Foundation fieldbus ..B-6 Approved Manufacturing Locations ..B-6 European Union Directive Information ..B-6 Hazardous Locations Installations ..B-6 Installation Drawings ..B-9 Reference Manual00809-0100-4021, Rev EBMay 2007 Rosemount 3144 PTOC-4 Reference Manual 00809-0100-4021, Rev EBMay 2007 Rosemount 1 IntroductionOverview .. page 1-1 Considerations .. page 1-2 Return of Materials .. page 1-53144P Revisions .. page 1-6 OVERVIEWM anualThis manual is designed to assist in the installation, operation, and maintenance of the Rosemount 1: Introduction Transmitter and Manual Overview Considerations Return of MaterialSection 2: Installation Mounting Installation Wiring Power SupplySection 3: HART Configuration 375 Field Communicator Configuration Multidrop Communication Calibration Trim the transmitterSection 4: FOUNDATION Fieldbus Configuration Calibration Hardware Maintenance Diagnostic Messaging Trim the TransmitterSection 5: Operation and Maintenance Maintenance TroubleshootingReference Manual00809-0100-4021, Rev EBMay 2007 Rosemount 3144P1-2 Appendix A.
9 Specifications and Reference Data Specifications Dimensional drawings Ordering InformationAppendix B: Approvals Product Certifications Installation DrawingsSafety Instrumented System (SIS) HART only Information regarding Safety Certified transmitters Transmitter Features of the Rosemount 3144P include: Accepts inputs from a wide variety of sensors Configuration using HART protocol or FOUNDATION fieldbus Electronics that are completely encapsulated in epoxy and enclosed in a metal housing, making the Transmitter extremely durable and ensuring long-term reliability A compact size and two housing options allowing mounting flexibility for the control room or the field Special dual-sensor features include Hot Backup , sensor drift alarm, first good, differential and average Temperature measurements, and four simultaneous measurement variable outputs in addition to the analog output signalRefer to the following literature for a full range of compatible connection heads, sensors.
10 And thermowells provided by Emerson Process Management. Temperature Sensors and Assemblies Product Data Sheet, Volume 1 (document number 00813-0100-2654) Temperature Sensors and Assemblies Product Data Sheet, Metric (document number 00813-0200-2654) CONSIDERATIONSG eneralElectrical Temperature sensors, such as resistance Temperature detectors (RTDs) and thermocouples (T/Cs), produce low-level signals proportional to Temperature . The 3144P Transmitter converts low-level signals to HART or FOUNDATION fieldbus signals. This signal is then transmitted to the control room via two power/signal Proper electrical installation is essential to prevent errors due to sensor lead resistance and electrical noise. The current loop must have between 250 and 1100 ohms resistance for HART communications.