SPECIFICATIONS - Medtronic

Physical characteristicsPhysical characteristicsVolumea33 cm3 Mass77 gH x W x D64 mm x 51 mm x 13 mmSurface area of device can57 cm2 Radiopaque IDbPFZM aterials in contact with human tissuec Titanium, polyurethane, silicone rubberBatteryHybrid CFx lithium/silver vanadium oxidea Volume with connector ports The radiopaque ID, which includes a Medtronic -identif ier symbol, can be viewed in a f luoroscopic image of the These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue during normal indicatorsRecommended Replacement Time (RRT)< V on 3 consecutive daily automatic measurementsEnd of Service (EOS)3 months after RRTM aximum energy levels and typical full energy charge timesMaximum programmed energy35 JMaximum delivered energya,b36 JMaximum stored energyc42 JTypical charge time at Beginning of Service (BOS) sTypical charge time at Recommended Replacement Time (RRT) sa Energy delivered at connector block into a 50 For

Device parameters Tachyarrhythmia detection parameters Parameter Programmable values VF Detection On ; Off VF Interval (Rate)a …

Tags:

  Specification, Rates, Medtronic

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of SPECIFICATIONS - Medtronic

1 Physical characteristicsPhysical characteristicsVolumea33 cm3 Mass77 gH x W x D64 mm x 51 mm x 13 mmSurface area of device can57 cm2 Radiopaque IDbPFZM aterials in contact with human tissuec Titanium, polyurethane, silicone rubberBatteryHybrid CFx lithium/silver vanadium oxidea Volume with connector ports The radiopaque ID, which includes a Medtronic -identif ier symbol, can be viewed in a f luoroscopic image of the These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue during normal indicatorsRecommended Replacement Time (RRT)< V on 3 consecutive daily automatic measurementsEnd of Service (EOS)3 months after RRTM aximum energy levels and typical full energy charge timesMaximum programmed energy35 JMaximum delivered energya,b36 JMaximum stored energyc42 JTypical charge time at Beginning of Service (BOS) sTypical charge time at Recommended Replacement Time (RRT) sa Energy delivered at connector block into a 50 For 35 J programmed energy, delivered energy exceeds 35 Energy stored at charge end on capacitor.

2 D Charge time during a nonwireless telemetry session may be slightly MRI XT VR SureScan DVMB1D4 MR Conditional with PhysioCurve Design SmartShock Technology OptiVol Fluid Status Monitoring Complete Capture Management DF4 Device parametersTachyarrhythmia detection parametersParameterProgrammable valuesVF DetectionOn ; OffVF Interval (Rate)a240; 250 .. 320 .. 400 msVF Initial Beats to Detect12/16; 18/24; 24/32; 30/40 ; 45/60; 60/80; 75/100; 90/120; 105/140; 120/160VF Beats to Redetect6/8; 9/12; 12/16 ; 18/24; 21/28; 24/32; 27/36; 30/40 FVT DetectionOff ; via VF; via VTFVT Interval (Rate)a200; 210 .. 240 .. 600 msVT DetectionOn; Off VT Interval (Rate)a280; 290 .. 360 .. 650 ms VT Initial Beats to Detect12; 16.

3 52; 76; 100VT Beats to Redetect8; 12 .. 52VT MonitorMonitor ; OffVT Monitor Interval (Rate)a280; 290 .. 450 .. 650 msMonitored VT Beats to Detect16; 20; 24; 28; 32 .. 56; 80; 110 ; 13 0 Wavelet WaveletbOn ; Off; MonitorTemplate[date]Match Threshold40; 43; 46 .. 70 .. 97%Auto CollectionOn ; Off SVT V. Limita240; 250; 260 .. 650 msOther enhancementsStabilityaOff ; 30; 40 .. 100 ms OnsetOff ; On; MonitorOnset Percent72; 75; 78; 81 ; 84; 88; 91; 94; 97%High Rate TimeoutVF Zone OnlyOff; ; ; ; 1; ; ; ; 2; ; 3; ; 4; ; 5 minAll ZonesOff ; ; 1; .. 5; 6; 7 .. 20; 22; 24; 26; 28; 30 min T- W a v eOn ; OffRV Lead NoiseOn; On+Timeout ; ; ; .. 2 minSensitivityRV Sensitivityc, ; ; ; ; ; mVa The measured intervals are truncated to a 10 ms multiple (for example, 457 ms becomes 450 ms).

4 The device uses this truncated interval value when applying the programmed criteria and calculating interval The Wavelet feature is automatically set to On when VF Detection is set to This setting applies to all sensing in this chamber for both tachyarrhythmia detection and bradycardia pacing Carefully evaluate the possibility of increased susceptibility to EMI and oversensing before changing the sensitivity threshold to its minimum (most sensitive) setting of mV. Ventricular tachyarrhythmia therapy parametersParameterProgrammable valuesVF Therapy parametersVF Therapy StatusOn ; OffEnergyRx1-Rx2: ; .. ; 2; 3 .. 16; 18; 20; 22; 24; 25; 26; 28; 30; 32; 35 JRx3-Rx6: 10; 11 .. 16; 18; 20; 22; 24; 25; 26; 28; 30; 32; 35 JPathwayaAX>B; B>AXRx1-Rx4: B>AX ; Rx5-Rx6: A X>B AT P During Charging ; Before Charging; Off Deliver ATP if last 8 R-R 200; 210.

5 240 .. 300 msTherapy TypeBurst ; Ramp; Ramp+ChargeSaver On ; Off Switch when number of consecutive ATP successes equals1 ; 2; 3; 4; 6; 8; 10 Smart ModeOn ; OffVT/FVT Therapy parametersVT Therapy StatusOn; Off FVT Therapy StatusOn; Off Therapy TypeCV; Burst; Ramp; Ramp+ Rx1: Burst ; Rx2-Rx6: CV ; .. ; 2; 3 .. 16; 18; 20; 22; 24; 25; 26; 28; 30; 32; 35 J VT Rx1-Rx2: 20 J VT Rx3-Rx6: 35 J FVT Rx1-Rx6: 35 JPathwayaAX>B; B>AX Rx1-Rx4: B>AX ; Rx5-Rx6: A X>B Burst therapy parametersInitial # Pulses1; 2 .. 8 .. 15R-S1 Interval = (%RR)50; 53; 56; 59; 63; 66 .. 84; 88 ; 91; 94; 97%Interval Dec0; 10 .. 40 ms# Sequences1; 2 .. 10VT Therapies: 3 ; FVT Therapies: 1 Smart ModebOn; Off Ramp therapy parametersInitial # Pulses1; 2.

6 8 .. 15R-S1 Interval = (%RR)50; 53; 56; 59; 63; 66 .. 84; 88; 91 ; 94; 97%Interval Dec0; 10 .. 40 ms# Sequences1; 2 .. 10VT Therapies: 3 ; FVT Therapies: 1 Smart ModebOn; Off Ventricular tachyarrhythmia detection parameters, continuedParameterProgrammable valuesRamp+ therapy parametersInitial # Pulses1; 2; 3 .. 15R-S1 Interval = (%RR) 50; 53; 56; 59; 63; 66 .. 75 .. 84; 88; 91; 94; 97%S1S2 (Ramp+) = (%RR)50; 53; 56; 59; 63; 66; 69 .. 84; 88; 91; 94; 97%S2SN (Ramp+) = (%RR)50; 53; 56; 59; 63; 66 .. 84; 88; 91; 94; 97%# Sequences1; 2 .. 10VT Therapies: 3 ; FVT Therapies: 1 Smart ModebOn; Off Shared Settings V-V Minimum ATP Interval150; 160 .. 200 .. 400 msV. Amplitude1; 2 .. 6; 8 VV. Pulse ; .. msV.

7 Pace Blanking150; 160 .. 240 .. 450 msActive Can /SVC CoilcCan+SVC On ; Can Off; SVC OffProgressive Episode TherapiesOn; Off Confirmation+On ; Offa I f the Active Can/SVC Coil parameter is set to Can Off, the Active Can electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the SVC Coil electrode is not used as part of the high-voltage delivery Smart Mode is available only for The Active Can/SVC Coil parameter applies to all automatic, manual, and emergency high-voltage therapies. It also applies to T-Shock parametersModes, rates , and intervalsParameterProgrammable valuesModeVVI ; VVIR; VOO; OVOL ower Ratea30; 35 .. 40 ; 45 .. 150 bpma The corresponding Lower Rate Interval can be calculated as follows: Lower Rate Interval (ms) = 60,000/Lower parametersParameterProgrammable valuesRV.

8 5; ; 6; 8 VRV Pulse ; ; ; ; ; .. msRV ; ; ; ; ; mV RV Pace PolarityBipolar; Tip to CoilRV Sense PolarityBipolar; Tip to Coila This setting applies to all sensing in this chamber for both tachyarrhythmia detection and bradycardia pacing Capture Management parametersParameterProgrammable valuesRV Capture ManagementAdaptive ; Monitor; OffRV Amplitude Safety ; ; ; Minimum Adapted ; ; ; ; ; VRV Acute Phase RemainingOff; 30; 60; 90; 120 ; 150 daysBlanking periodsParameterProgrammable valuesV. Blank Post VP150; 160 .. 200 .. 450 msV. Blank Post VS120 ; 130 .. 170 msRate response pacing parametersParameterProgrammable valuesUpper Sensor Rate80; 85 .. 120 .. 175 bpmADL Rate60; 65.

9 95 .. 170 bpmRate Profile OptimizationOn ; OffADL Response1; 2; 3 ; 4; 5 Exertion Response1; 2; 3 ; 4; 5 Activity ThresholdLow; Medium Low ; Medium High; HighActivity Acceleration15; 30 ; 60 sActivity DecelerationExercise ; ; 5; 10 minADL Setpoint5; 6 .. 40; 42 .. 80UR Setpoint15; 16 .. 40; 42 .. 80; 85 .. 180 Conducted AF response parametersParameterProgrammable valuesConducted AF ResponseOn; Off Response LevelLow; Medium ; HighMaximum Rate80; 85 .. 110 .. 130 bpmVentricular rate stabilization parametersParameterProgrammable valuesV. Rate StabilizationOn; Off Maximum Rate80; 85 .. 100 .. 120 bpmInterval Increment100; 110 .. 150 .. 400 msPost VT/VF shock pacing parametersParameterProgrammable valuesPost VT/VF Shock PacingOn; Off Overdrive Rate70; 75; 80.

10 120 bpmOverdrive ; 1; 2; 3; 5; 10; 20; 30; 60; 90; 120 minPost shock pacing parametersParameterProgrammable valuesPost Shock V. Amplitude1; 2 .. 6 ; 8 VPost Shock V. Pulse ; .. msSleep parametersParameterProgrammable valuesSleepOn; Off Sleep Rate30; 35 .. 50 ; 55; 60; 70; 75 .. 100 bpmBed Time00:00; 00:10 .. 22:00 .. 23:50 Wake Time00:00; 00:10 .. 07:00 .. 23:50 MRI SureScan parametersParameterProgrammable valuesMRI SureScanOn; OffMRI Pacing ModeVOO (Asynchronous); OVO (Of f )MRI Pacing Rate60; 70; 120 bpmAdditional pacing featuresParameterProgrammable valuesRate HysteresisOff ; 30; 40 .. 80 bpmMedtronic CareAlert parametersClinical management alertsParameterProgrammable valuesOptiVol Fluid Settings Device ToneOptiVol Alert EnableOff (Observation only)OptiVol Thresholdc30; 40; 50; 60.