Technical Information Condumax W CLS21 and CLS21D

TI085C/24/ InformationCondumax W CLS21 and CLS21 DConductivity sensors, analog or digital with Memosens technologyCell constant k = 1 cm-1 ApplicationMeasurements in media of medium and high conductivities: Medium separation in medium conductivities (milk/water) Medium separation in high conductivities (alkaline solution/water) Drinking water treatment Wastewater treatmentThe cell constant of the sensor is k = 1 cm-1. The measuring range reaches from 10 S/cm to 20 with a temperature sensor are used together with conductivity transmitters equipped with automatic temperature compensation: Liquiline M CM42 Mycom S CLM153 Liquisys M CLM223/253 For measurement of resistivity, M cm measuring ranges are available in the menus of these benefits Various designs guarantee optimum adaptation to the process conditions and methods of installation Installation in pipes or flow chambers Compact design Available with plug-in head or fixed cable High chemical, thermal and mechanical stability IP 65/NEMA 4 (with four-pole plug-in head), IP 67 / NEMA 4X (with fixed cable), IP 68 / NEMA 6P (with Memosens plug-in head) Quality certificate with statement of the individual cell constantFurther benefits offered by Memosens technology Maximum process safety through contactless inductive signal transmission Data safety through digital data transmission Easy handling thanks to storage of sensor-specific data Predictive maintenance possible thanks to registration o

TI085C/24/ae/10.07 Technical Information Condumax W CLS21 and CLS21D Conductivity sensors, analog or digital with Memosens technology Cell constant k = 1 cm-1 Application

Tags:

  Information, Applications, Technical, Technical information condumax w cls21, Condumax, Cls21

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Technical Information Condumax W CLS21 and CLS21D

1 TI085C/24/ InformationCondumax W CLS21 and CLS21 DConductivity sensors, analog or digital with Memosens technologyCell constant k = 1 cm-1 ApplicationMeasurements in media of medium and high conductivities: Medium separation in medium conductivities (milk/water) Medium separation in high conductivities (alkaline solution/water) Drinking water treatment Wastewater treatmentThe cell constant of the sensor is k = 1 cm-1. The measuring range reaches from 10 S/cm to 20 with a temperature sensor are used together with conductivity transmitters equipped with automatic temperature compensation: Liquiline M CM42 Mycom S CLM153 Liquisys M CLM223/253 For measurement of resistivity, M cm measuring ranges are available in the menus of these benefits Various designs guarantee optimum adaptation to the process conditions and methods of installation Installation in pipes or flow chambers Compact design Available with plug-in head or fixed cable High chemical, thermal and mechanical stability IP 65/NEMA 4 (with four-pole plug-in head), IP 67 / NEMA 4X (with fixed cable), IP 68 / NEMA 6P (with Memosens plug-in head)

2 Quality certificate with statement of the individual cell constantFurther benefits offered by Memosens technology Maximum process safety through contactless inductive signal transmission Data safety through digital data transmission Easy handling thanks to storage of sensor-specific data Predictive maintenance possible thanks to registration of sensor load dataCondumax W CLS21 / CLS21D2 Endress+HauserFunction and system designMeasuring principleConductive measurement of conductivityGeneral properties ElectrodesThe sensor has two coaxial electrodes made of graphite for a large measuring graphite guarantees high chemical stability and low polarization effects. Temperature compensationA temperature sensor is integrated to measure the medium temperature. Durability The sensor is pressure-proof up to 16 bar at 20 C (232 psi at 68 F). It can be applied with temperatures of up to 135 C at bar (275 F at psi).Important properties of CLS21 DMaximum process safetyThe inductive and non-contacting measured value transmission of Memosens guarantees maximum process safety and offers the following benefits: All problems caused by moisture are eliminated.

3 The plug-in connection is free from corrosion. Measured value distortion from moisture is not possible. The plug-in system can even be connected under water. The transmitter is galvanically decoupled from the medium. EMC safety is guaranteed by screening measures for the digital measured value safety through digital data transferThe Memosens technology digitalizes the measured values in the sensor and transfers them to the transmitter contactlessly and free from interference potential. The result: An automatic error message is generated if the sensor fails or the connection between sensor and transmitter is interrupted. The availability of the measuring point is dramatically increased by immediate error detection. Application in hazardous areas is unproblematic; the integrated electronics are intrinsically measurement of conductivityACPower supplyICurrent meterUVoltage meterThe conductivity of liquids is measured with the following measurement setup: Two electrodes are immersed in the medium.

4 An AC voltage is applied to these electrodes which generates a current in the electric resistance or its reciprocal value, the conductance G, is calculated according to Ohm s law. The specific conductivity is determined using the cell constant k that is dependent on the sensor W CLS21 / CLS21 DEndress+Hauser3 Easy handlingSensors with Memosens technology have integrated electronics that allow for saving calibration data and further Information such as total hours of operation and operating hours under extreme measuring conditions. When the sensor is connected, the calibration data are automatically transferred to the transmitter and used to calculate the current measured value. Storing the calibration data in the sensor allows for calibration and adjustment away from the measuring point. The result: Sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather do neither affect the calibration quality nor the operator. The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated sensors.

5 Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive maintenance is possible. The sensor history can be documented on external data carriers and evaluation programs at any time. Thus, the current application of the sensors can be made to depend on their previous with the transmitterAlways connect digital sensors to a transmitter with Memosens technology. Data transmission to a transmitter for analog sensors is not storage of CLS21 DDigital sensors are able to store the following system data in the sensor. Manufacturing data Serial number Order code Date of manufacture Calibration data Calibration date Cell constant Change in cell constant Number of calibrations Serial number of the transmitter used for the last calibration Application data Temperature application range Conductivity application range Date of first commissioning Maximum temperature value Operating hours at temperatures above 80 C / 120 C (176 F / 248 F) Number of sterilizationsMeasuring systemA complete measuring system comprises.

6 A CLS21 or CLS21D conductivity sensor a transmitter, Liquiline M CM42 a measuring cable, CYK71 or CYK10 Memosens data cablea0003417 Measuring system example1 Liquiline M CM42 transmitter2 CYK10 Memosens data cable3 Condumax W CLS21D sensor123 Condumax W CLS21 / CLS21D4 Endress+HauserInputMeasured valuesConductivityTemperatureCell constant kk = 1 cm 1 nominalMeasuring rangesTemperature sensorCable specificationCLS21 CLS21 is connected to the measuring transmitter using the special measuring cable CYK71 or CYK71-Ex or the fixed or fixed cableCLS21 DCLS21D is connected to the transmitter using the Memosens data cable characteristicsMeasured errorEach individual sensor is factory measured in a solution of approx. 5 mS/cm on a reference measuring system referred to NIST or DKD. The accurate cell constant is entered in the supplied quality certificate. The maximum measured error in cell constant determination is %.Conductivity measurement (referred to water at 25 C / 77 F)10 S/cm to 20 mS/cmin the following temperature range: CLS21 : -20 to 135 C (-4 to 275 F) CLS21D : -20 to 100 C (-4 to 212 F), spec.

7 Measuring accuracy up to 100 C (212 F)Temperature measurementCLS21: CLS21D :- 20 to 135 C (-4 to 275 F)- 20 to 100 C (-4 to 212 F), spec. measuring accuracy up to 100 C (212 F) CLS21 :Pt 100 CLS21D :NTCGN/YEOuter shield(see wiring diagramof transmitter)YEGNWHPt 100 Coax BKCoax innerMeasuring electrodeBN electrode}CableSensor pin231231})GYa0003350 Special measuring cable CYK10GN/YEYEGNBNWHGND+ Com ACom BGYC ondumax W CLS21 / CLS21 DEndress+Hauser5 InstallationInstallation instructionsThe sensors are mounted directly via the process , they can be installed in flow installation of sensors with G1 thread in tanks, the CLA111 immersion and process assembly is available (see Accessories).a0003419 Dipfit W CLA111, mounting versions A, B and DA = Standard flange, B = Adjustable flange, D = Suspended bracket!Note! The measuring surfaces must be completely immersed in the medium during protectiona0003418 Installation in the CLA751 flow assemblya0003422 Installation in the CLA752 flow assemblyABDCLS21 Fixed cable:IP 67 (i NEMA 4X)Plug-in head:IP 65 (i NEMA 4) CLS21D .

8 IP 68 (i NEMA 6P) Condumax W CLS21 / CLS21D6 Endress+HauserProcessProcess temperature-20 to +135 C @ bar (-4 to +275 F @ psi)Process pressure16 bar @ 20 C (232 psi @ 68 F)Pressure/temperature load curvea0003425 Mechanical pressure-temperature stability of the sensorMechanical constructionDesign, dimensions CLS214812163006090120150T [ C] 4010060140 180 22030050100150232-200T [ F]2026068275200250p [psi]p [bar] cable version with NPT 1" thread1 Fixed cable2 NPT 1" thread3 Sensor shafta0003434 Connector version with 2" clamp1 Four-pole connector2Pg 9 cable gland32" clamp4 Sensor shafta0003427 Dimensions of fixed-cable version with NPT 1" threada0003428 Dimensions of fixed-cable version with 2" clamp1231423AF 3516( )61( )113( )157( )14( ) 24( )NPT 1"NPT 1"mm (inch)16( ) 24( ) 64( ) 34( )15( )61( )113( )144( )mm (inch) Condumax W CLS21 / CLS21 DEndress+Hauser7a0007560 Dimensions of fixed-cable version with SMS threada0003429 Dimensions of plug-in head version with G 1 threada0003431 Dimensions of plug-in head version with 2" clampa0003432 Dimensions of plug-in head version with dairy fittinga0003450 Dimensions of plug-in head version with SMS thread 34( ) 47( ) 55( )16( )61( )54( )144( )5( )10( ) 24( )mm (inch)151( )16( )79( )36( )18( )21( )36( ) 34( ) 24( )G1mm (inch)151( )61( )54( )36( )15( )16( ) 24( ) 64( ) 34( )mm (inch)151( )61( )54( )36( )10( )16( ) 34( )D 25/DN 40N 24( )mm (inch) 34( ) 47( ) 55( )16( )61( )54( )36( )151( )5( )10( ) 24( )mm (inch) Condumax W CLS21 / CLS21D8 Endress+HauserDesign, dimensions CLS21 DWeightDepending on version, approx.

9 Kg / with G 1 thread1 Memosens plug-in head2G 1 thread3 Sensor shafta0007506 Dimensions of version with G 1 threada0007503 Dimensions of version with 2" clampa0007505 Dimensions of version with dairy fitting123 44( )186( )16( )79( )18( ) 24( )G1mm (inch)186( )61( )15( )16( ) 24( ) 64( ) 44( )mm (inch)186( )61( )10( )16( ) 44( )DN 25 24( )mm (inch)Electrodes:graphiteSensor shaft:polyethersulfone (PES-GF20)Thermal conductivity socket for temperature sensor: titanium W CLS21 / CLS21 DEndress+Hauser9 Process connections!Note! Clamp connection:Sensors with clamp connections can be fixed using sheet metal brackets or solid metal brackets have a lower dimensional stability, uneven bearing surfaces causing point loads and sometimes sharp edges that can damage the strongly recommend to always use solid brackets because of their higher dimensional stability. Solid brackets may be applied over the total pressure-temperature range (see temperature-pressure load curve).

10 Certificates and approvalsEx approval1) ATEX II 1G EEx ia IIC T3 / T4 / T6 FM in combination with the Liquiline M CM42 and Mycom S CLM153 transmittersfor all product versions listed in the product structure (see Ordering Information )Quality certificateWith statement of the individual cell constantCLS21:Fixed-cable version:Thread NPT 1"Clamp 2" acc. to ISO 2852 SMS DN 38 (1 ")Plug-in head version:Thread G 1 Clamp 2" acc. to ISO 2852 SMS DN 38 (1 ")Dairy fitting DN 25 or DN 40 acc. to DIN 11851 CLS21D :Thread G 1 Thread NPT 1"Clamp 2" acc. to ISO 2852 Dairy fitting DN 25 acc. to DIN 118511) Approval for CLS21D pendingCondumax W CLS21 / CLS21D10 Endress+HauserOrdering informationProduct structure Condumax W CLS21 Product structure Condumax W CLS21 DAccessoriesAssemblies Dipfit W CLA111 immersion and process assemblyFor open and closed tanks with DN 100 flange,for ordering Information , see Technical Information Dipfit W CLA111 (TI135C/07/en)a0003435 Dipfit W CLA111, DN 100 flange or suspension bracket, mounting versions A, B und DMeasuring range and cell constantCMeasuring range: S.