Technical Information Proline t-mass 65F, 65I

TI069D/06/ InformationProline t-mass 65F, 65 IThermal Mass Flow Measuring SystemDirect Mass Flow Measurement of GasesApplicationFor measuring the mass flowrate of a wide range of gas types Compressed air Natural gas flow to boilers/dryers Carbon Dioxide flow in breweries Biogas and aeration air in waste water plants Gas production (e. g. Ar, N2, CO2, He, O2) Leakage detectionApprovals for hazardous area: ATEX, FM, CSA, IECEx, NEPSIC onnection to all common process control systems: HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, MODBUS RS485 Your benefitsDirect measurement of gas mass flow.

Proline t-mass 65F, 65I 4 Endress+Hauser Input Measured variable •Ma fssolw • Gas temperature • Gas heat flow Measuring range (air at ambient conditions) The measuring range is dependant on the gas selection, line size and use of flow conditioner.

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Technical Information Proline t-mass 65F, 65I

1 TI069D/06/ InformationProline t-mass 65F, 65 IThermal Mass Flow Measuring SystemDirect Mass Flow Measurement of GasesApplicationFor measuring the mass flowrate of a wide range of gas types Compressed air Natural gas flow to boilers/dryers Carbon Dioxide flow in breweries Biogas and aeration air in waste water plants Gas production (e. g. Ar, N2, CO2, He, O2) Leakage detectionApprovals for hazardous area: ATEX, FM, CSA, IECEx, NEPSIC onnection to all common process control systems: HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, MODBUS RS485 Your benefitsDirect measurement of gas mass flow.

2 Provides temperature as an Proline transmitter concept comprises: Modular device and operating concept resulting in a higher degree of efficiency Quick setup operating menus for ease of commissioning On board software freely allows the selection of up to 20 pure gases and creation of mixed gases with a maximum of 8 components (e. g. Biogas)The t-mass sensors offer: Negligible pressure drop or loss Wide turndown of up to 100:1 Insertion version can be programmed for circular pipe or rectangular ducting installation Each device individually calibrated and delivered with a traceable certificate Can be calibrated with flow conditioner on request.

3 Optional hot tap device for insertion allowing ease of removal/replacement for range of process pressure up to 16 Barg (230 psig) and non-toxic gas t-mass 65F, 65I2 Endress+HauserTable of contentsFunction and system design.. 3 Measuring principle .. 3 Measuring system .. 3 Input .. 4 Measured variable .. 4 Measuring range (air at ambient conditions) .. 4 Input signal .. 5 Output .. 5 Output signal .. 5 Signal on alarm .. 7 Load .. 7 Low flow cut off .. 7 Galvanic isolation .. 7 Switching output .. 7 Power supply .. 8 Electrical connection Measuring unit.

4 8 Terminal assignment .. 9 Electrical connection Remote version .. 10 Supply voltage .. 10 Cable entries .. 10 Remote version cable specifications .. 10 Power consumption .. 10 Power supply failure .. 11 Potential equalization .. 11 Performance characteristics.. 11 Calibration reference conditions .. 11 Maximum measured error .. 11 Repeatability .. 11 Influence of medium pressure(Pressure co-efficient) .. 11 Response time .. 11 Operating conditions: Installation .. 12 Installation instructions .. 12 Inlet and outlet runs .. 14 Mounting conditions for the insertion version.

5 17 Length of connecting cable .. 18 Operating conditions: Environment .. 18 Ambient temperature .. 18 Storage temperature .. 18 Degree of protection .. 18 Shock resistance .. 18 Vibration resistance .. 18 Electromagnetic compatibility (EMC) .. 18 Operating conditions: Process .. 19 Medium temperature range .. 19 Pressure loss .. 19 Medium pressure range (nominal pressure) .. 19 Flow limit .. 19 Process conditions for Hot tap .. 19 Mechanical construction .. 20 Design, dimensions .. 20 Weight .. 36 Materials .. 36 Material load curves .. 37 Process connections.

6 38 Human interface .. 39 Display elements .. 39 Operating elements .. 39 Languages .. 39 Remote operation .. 39 Certificates and approvals .. 39CE mark .. 39C-Tick mark .. 39Ex approval .. 39 FOUNDATION Fieldbus certification .. 39 PROFIBUS DP/PA certification .. 40 MODBUS certification .. 40 Pressure measuring device approval .. 40 Oxygen service .. 40 Other standards and guidelines .. 40 Ordering Information .. 41 Accessories .. 41 Device-specific accessories .. 41 Measuring principle-specific accessories .. 41 Service-specific accessories.

7 42 Communication-specific accessories .. 42 Documentation .. 43 Registered trademarks .. 43 Proline t-mass 65F, 65 IEndress+Hauser3 Function and system designMeasuring principleThermal dispersion principleThe thermal principle operates by monitoring the cooling effect of a gas stream as it passes over a heated transducer (PT100).Gas flowing through the sensing section passes over two PT 100 RTD transducers one of which is used conventionally as a temperature sensing device, whilst the other is used as a heater. The temperature sensor monitors the actual process values whilst the heater is maintained at a constant differential temperature above this by varying the power consumed by the sensor.

8 The greater the mass flow, the greater the cooling effect and power required to maintain the differential temperature. The measured heater power is therefore a measure of the gas mass systemThe measuring system consists of a transmitter and a sensor. Two versions are available: Compact version: transmitter and sensor form a single mechanical unit. Remote version: transmitter and sensor are mounted physically separate from one 65a0003671 Two-line liquid-crystal display Configuration using push button operationSensorFa0005137 Flanged version Nominal diameters DN 15 to 100 ( " to 4") Sensor body material:- , 316L- CF3M Transducer body material.

9 - , 316L, Alloy C22Ia0005138 Insertion version Sensor length 235/335/435/608 for DN 80 to 1500( " " " " for line size 3" to 60") Sensor body , 316/316L Transducer body , Alloy C22 EscE-+ Proline t-mass 65F, 65I4 Endress+HauserInputMeasured variable Mass flow Gas temperature Gas heat flowMeasuring range (air at ambient conditions)The measuring range is dependant on the gas selection, line size and use of flow conditioner. Each meter is individually calibrated on air and mathematically converted to suit the customer specific gas, where table below defines the ranges available for air without flowconditioner.

10 Please refer to your Endress+Hauser representative or to Applicator, the selection tool, for other gases and process range for the flanged version, metric units:Measuring range for the flanged version, US units:Measuring range for the insertion version, metric units:In order to achieve optimum performance it is recommended that under operating conditions the maximum velocity is limited to a value below 70 at 0 C, bar ascf/min. at 15 C, bar at 59 F, psi ascf/min. at 59 F, psi aminimummaximumminimummaximumminimummaxi mum " " " " "444466161656109744"848250303060181800 DNkg/hNm3/h at 0 C, bar ascf/min.