The Difference Between Chemical Integrators and Biological IndicatorsIMPORTANT RESEARCH FINDINGS INSIDE product type% of Failures Detectedchemical integratorsbiological indicatorsself-contained enzymaticself-contained growthself-contained growthspore strip% positive / rejectintegrator Bintegrator Cintegrator A1009080706050403020100 Under acceptable conditions for steam sterilization, the Integrators compared reasonably well with the biological indicators. However, the Integrators were not able to consistently detect the failure conditions of superheated steam and incomplete air removal as did the biological indicators in all Methodology:250 F Superheated steam using a prevacuum cycle with an 8-minute exposure in a BIER Vessel (% positive for biological indicators and % reject for Chemical Integrators )KEY FINDING: BIs and CIs are not study, published in the american Journal of Infection Control, demonstrates that Chemical Integrators (CIs) and biological indicators (BIs)

ABSTRACT The study, published in the American Journal of Infection Control, demonstrates that chemical integrators (CIs) and biological indicators (BIs) are not equivalent in the role of identifying sterilization failures. CIs provide immediate valuable information about specific parameters of the

Tags:

  American, Chemical, Between, Differences, Integrator, The difference between chemical integrators and

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of The Difference Between Chemical Integrators and …

1 The Difference Between Chemical Integrators and Biological IndicatorsIMPORTANT RESEARCH FINDINGS INSIDE product type% of Failures Detectedchemical integratorsbiological indicatorsself-contained enzymaticself-contained growthself-contained growthspore strip% positive / rejectintegrator Bintegrator Cintegrator A1009080706050403020100 Under acceptable conditions for steam sterilization, the Integrators compared reasonably well with the biological indicators. However, the Integrators were not able to consistently detect the failure conditions of superheated steam and incomplete air removal as did the biological indicators in all Methodology:250 F Superheated steam using a prevacuum cycle with an 8-minute exposure in a BIER Vessel (% positive for biological indicators and % reject for Chemical Integrators )KEY FINDING.

2 BIs and CIs are not study, published in the american Journal of Infection Control, demonstrates that Chemical Integrators (CIs) and biological indicators (BIs) are not equivalent in the role of identifying sterilization provide immediate valuable information about specific parameters of the sterilization process such as time, temperature and steam quality. CIs are designed to test certain conditions within a specified range, but do not reflect the complete range of processing conditions. While Class 5 Chemical indicators, termed integrating indicators, react to multiple parameters, the study shows that they can fail to detect sterilization Integrators are not equivalent to biological indicators.

3 Only biological indicators demonstrate the ability of a steam process to kill microbes. Performance of Various Steam Sterilization Indicators Under Optimum and Sub-optimum Exposure ConditionsPhilip M. Schneider, Robert R. Reich, Steven S. Kirckof, William G. Foltz american Journal of Infection Control, Vol. 33, No. 5, Sup. 2, June 2005, pp. S55 S673 Health Care3M Center, Building 275-4W-02St. Paul, MN 55144-1000 USA1 800 Canada Post Office Box 5757 London, Ontario N6A 4T1 Canada1 800 563-2921 Minimum 10% Post-Consumer Fiber Printed in 3M 2005 All Rights Reserved70-2009-7086-4 ( ) DPIR ecycle, NegFor additional information contact your 3M sales representative or call the 3M Health Care Customer Helpline at 1-800-228-3957.

4 Published study findings are available online through the AJIC Journal website at Chemical Integrators can never be equivalent because a biological indicator shows spores were actually killed. Chemical indicators do not. Biological indicators are the only monitoring devices that automatically measure every factor that could possibly affect spore destruction. Dr. Donald Vesley, University of Minnesota Chemical indicators could be used in conjunction with biological indicators, but should not replace them, because of inadequacies at marginal sterilization times and because only a biological indicator consisting of resistant spores can measure the microbial killing power of the sterilization process.

5 Dr. William Rutalla, University of North Carolina. Infection Control and Hospital Epidemiology, Vol. 14, 1993 The pass response of a CI does not prove that the item monitored by the indicator is sterile. The use of CIs is part of an effective quality assurance program; they should be used in conjunction with physical monitors and BIs to demonstrate the efficacy of the sterilization process. ANSI/AAMI ST46, 2002 The only currently accepted system capable of integrating all physical parameters responsible for lethality is a The BI is the only tool that will accurately integrate the combined lethal parameters within the More weight must be given to biological results because all critical parameters cannot be measured by physical means.

6 FDA Journal of Pharmaceutical Sciences and Technology, Vol. 58, No. 2, March/April 2004