LEAD by Flame AAS 7082
NIOSH Manual of Analytical Methods (NMAM), Fourth EditionLEAD by Flame AAS 7082Pb (Pb)CAS:7439-92-1 (Pb)RTECS:OF7525000 (Pb) (PbO)1317-36-8 (PbO)OG1750000 (PbO)METHOD: 7082, Issue 2 EVALUATION: FULLIssue 1: 15 February 1984 Issue 2: 15 August 1994OSHA mg/m3NIOSH:< mg/m3; blood Pb 60 g/100 mg/m3PROPERTIES:soft metal;d g/cm3; MP Cvalences +2, +4 in salts SYNONYMS:elemental lead and lead compounds except alkyl leadSAMPLINGMEASUREMENTSAMPLER:FILTER( m cellulose ester membrane)FLOW RATE:1 to 4 L/minVOL-MIN: 200 L @ mg/m3 -MAX:1500 LSHIPMENT:routineSAMPLE STABILITY:stableBLANKS:2 to 10 field blanks per setTECHNIQUE:ATOMIC ABSORPTIONSPECTROPHOTOMETER, FLAMEANALYTE:leadASHING:conc. HNO3, 6 mL + 30% H2O2,1 mL; 140 CFINAL SOLUTION:10% HNO3, 10 mLFLAME:air-acetylene, nmBACKGROUNDCORRECTION:D2 or H2 lamp, or ZeemanCALIBRATION:Pb2+ in 10% HNO3RANGE10 to 200 g per sample [2,3]ESTIMATED g per sample [4]PRECISION (r) [1]ACCURACYRANGE STUDIED: to mg/m3 [1]; to mg/m3 (fume) [2]BIAS: PRECISION (rT): [1]; (fume) [2]ACCURACY: : The working range is to >1 mg/m3 for a 200-L air sample.
LEAD by Flame AAS: Method 7082, Issue 2, dated 15 August 1994 - Page 3 of 7 NIOSH Manual of Analytical Methods (NMAM), Fourth Edition C C sV s C bV b V, mg/m3. NOTE: If the concentration (M) of any of the following is expected to exceed the lead
Download LEAD by Flame AAS 7082
Information
Domain:
Source:
Link to this page:
Please notify us if you found a problem with this document:
Related search queries
LEAD in Surface Wipe Samples, PREPARATION FOR FLAME ATOMIC, LEAD ATOMIC ABSORPTION,, LEAD ATOMIC ABSORPTION, DIRECT ASPIRATION, Flame Photometer, Atomic absorption spectrometry, Atomic absorption spectrometry Atomic absorption spectrometry, Atomic, Atomic Emission Spectroscopy, 7000B: Flame Atomic Absorption, 7000B FLAME ATOMIC ABSORPTION SPECTROPHOTOMETRY, Flame atomic absorption spectroscopy, Flame atomic