2.2.24. Absorption spectrophotometry, infrared - …

EUROPEAN PHARMACOPOEIA Absorption spectrophotometry , infrared sensitivity to obtain a maximum absorbance reading. Rinse the residuals of each calibration level are randomly in order to avoid contamination and memory effects. After distributed around the calibration curve. completing the analysis, rinse with water R or acidified water. Calculate the mean and relative standard deviation for the If a solid sampling technique is applied, full details of the lowest and highest calibration level. procedure are provided in the monograph. When the ratio of the estimated standard deviation of the Ensure that the concentrations to be determined fall lowest and the highest calibration level is less than or preferably within the linear part of the calibration curve. If greater than , a more precise estimation of the calibration this is not possible, the calibration plots may also be curved curve may be obtained using weighted linear regression.

2.2.24. Absorption spectrophotometry, infrared EUROPEAN PHARMACOPOEIA 6.0 PREPARATION OF THE SAMPLE FOR RECORDING BY TRANSMISSION OR ABSORPTION Prepare the substance by one of the following methods.

Tags:

  Infrared, Absorption, Absorption spectrophotometry, Spectrophotometry