Prepared by Jonathan FitzsimmonsAnalytical method Validation: ICP-OESP urpose: The following is a summary of tests/experiments performed to validate the ICP-OES instrument and establish an analysis method for samples. Instrument description: The instrument Inductively Coupled Plasma-Optical Emission Spectroscopy ( ICP-OES ) is used in atomic spectroscopy, and during analysis the sample is decomposed by intense heat into a cloud of hot gases containing free atoms and ions of the element(s) of interest. The high temperatures cause significant amounts of collisional excitation and ionization of the sample atoms.
analyzed in 2% nitric acid. The QC and desired production samples contain elements with concentrations on the range of the calibration curve. o Calibration curve acceptance or rejection criteria From OPM 19.2.24 ICP-OES Operation Model Optima 7300 DV: After analysis of the