Biomet Compression Hip Screw System This brochure is presented to demonstrate the surgical technique utilized by Frank R. Ebert, , Baltimore, Moryland. Biomet, as the manufacturer of this device, does not practice medicine and does not recommend this or any other surgical technique for use on o specific patient. The surgeon who performs any implant proce-dure is responsible for determining and utilizing the appropriate techniques for implanting the prosthesis in each individual patient. Biomet is not responsible for selection of the appropriate surgical technique to be util-ized for an individual patient. Step 1 Perform a closed reduction of the fracture, on the selected fracture table, check the fracture under the image intensifier in the A/P and lateral views to ensure adequate reduction and to identify areas of comminution, particularly seen posteriorly. Step 1a Operatively expose the fracture and make any appropriate modifications to the fracture reduction.