INSTALLATION GUIDEG C GC Columns2 TROUBLESHOOTING TOOLS Flow meter Leak detector New syringe COLUMN performance test sample Methane or other non-retained compound Reference COLUMN New septa, ferrules and injector liners Instrument manualsPRE-INSTALLATION CHECKLIST Replace oxygen, moisture and hydrocarbon traps as necessary. Check gas cylinder pressures to ensure that an adequate supply of carrier, make-up and fuel gases are available. Carrier gases should be of the highest purity. Note: It is critical that oxygen and water be removed from the carrier gas by the appropriate use of filters and COLUMN INSTALLATION STEPS Ensure that the injection port is clean and free of sample residues, septum, or capillary debris. Check and replace as necessary critical injector components such as seals, liners, and septa.
of the GC column, allowing a section of column to protrude. Trim one to two centimeters from the protruding end to remove ferrule contamination that may have entered the column. Inspect the cut with a ... COLUMN PHASE SELECTION GUIDE. 7 Polarity Scale ZebronTM HighlightPhase
Domain:
Source:
Link to this page:
Please notify us if you found a problem with this document:
{{id}} {{{paragraph}}}
GC Column, Selection Guide, GC Stationary Phase Selection guide, Column, GC Column Selection GC Columns and Accessories, GC Column Selection, Syringe Selection, Care and Maintenance Guide, GC Column Care Guide, Gas Chromatography, Selection, Experience the Zebron Difference, Gas Chromatograph Gas Chromatography, Column selection GC, Gas Chromatograph, Solvent Retention Data, Solvent Retention Data Solvent Retention Data Column