iSTAT1: Appendix F - Troubleshooting Guide

PHSA Laboratories CW Site - Point of CareTitle: CWPC_BGE_0160 iSTAT1 Appendix F - Troubleshooting GuideMedical Approval: Dr Benjamin JungMedical Approval Date: 22 Dec 2016 Version: Date: 6/24/2019 12:22:44 AMFolder Name: CW\Point of Care\Blood Gas Electrolytes - iSTATThis is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMSPage 1 of 14iSTAT1: Appendix F - Troubleshooting GuideFor help with Troubleshooting , contact POCT Technologist email local 7521 or after hours contact local 7850 for Loaner PromptExplanationActionNo DisplayDisplay screen remains blankDead batteriesKeypad not respondingInternal start switch brokenReplace BatteriesReinstall current software cartridge Locked Not RemovedAnalyzer will reset & release the cartridge after the testing cycle is complete.

Procedure: i-STAT Ceramic Conditioning Cartridge (CCC) for Analyzer Pin Conditioning REFERENCES. PHSA Laboratories CW Site - Point of Care Title: CWPC_BGE_0160 iSTAT1 Appendix F - Troubleshooting Guide Medical Approval: Dr Benjamin Jung Medical Approval Date: 22 …