Example: biology

August 2017 Micro Motion Compact Density MetersPS …

Product Data Sheet PS-001482, Rev G. August 2017. Micro Motion Compact Density Meters Peak performance precision Density meter Unparalleled real-world performance Superior application performance via traceable calibrations, performed at combined pressure and temperature conditions OIML R117-1 approved for MID conformance ISO/IEC accredited calibration Superior multi-variable I/O, meter health, and application capabilities Flow rate indication (velocity/volume flow) ensures sample integrity Internal diagnostics for fast verification of meter health and installation Application-specific factory configurations ensure fit-for-purpose operation Installation flexibility and compatibility Fluid, process and environmental effects are minimized to ensure superb measurement confidence Supports multiple protocols for connection to DCS, PLC, and flow computers Retrofit option available for Micro Motion 7835 and 7845 liquid Density meters Optional stainless steel transmitter housing for corrosion resistance in harsh environments Compact Density Meter August 2017.

Micro Motion Compact Density Meters Compact Density Meters use the Micro Motion dual curved-tube meter technology to measure density. These meters use a multi-

Tags:

  Technology

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of August 2017 Micro Motion Compact Density MetersPS …

1 Product Data Sheet PS-001482, Rev G. August 2017. Micro Motion Compact Density Meters Peak performance precision Density meter Unparalleled real-world performance Superior application performance via traceable calibrations, performed at combined pressure and temperature conditions OIML R117-1 approved for MID conformance ISO/IEC accredited calibration Superior multi-variable I/O, meter health, and application capabilities Flow rate indication (velocity/volume flow) ensures sample integrity Internal diagnostics for fast verification of meter health and installation Application-specific factory configurations ensure fit-for-purpose operation Installation flexibility and compatibility Fluid, process and environmental effects are minimized to ensure superb measurement confidence Supports multiple protocols for connection to DCS, PLC, and flow computers Retrofit option available for Micro Motion 7835 and 7845 liquid Density meters Optional stainless steel transmitter housing for corrosion resistance in harsh environments Compact Density Meter August 2017.

2 Micro Motion Compact Density Meters Compact Density Meters use the Micro Motion dual curved-tube meter technology to measure Density . These meters use a multi- variable measurement system, designed for fiscal metering of high-value products such as crude oil, refined hydrocarbons, alcohol, and many aggressive process liquids. Application configurations You can preselect an application-specific configuration for your meter from a wide range of options. Density Referred Density Specific Gravity Flow rate indication (velocity). API. %Concentration %Alcohol by Volume (ABV). Brix, Baume, Plato Integral transmitter . Supports Time Period Signal (TPS), 2-wire TPS, Analog (4 20 mA), HART, WirelessHART , and Modbus RS-485 communications. STATUS. SCRO. LL. SELECT. Meter diagnostics Ensure measurement health through known Density verification (KDV) and other meter and installation diagnostic capabilities.

3 0(7(5 +($/7+. 2 August 2017 Compact Density Meter Retrofit capabilities Retrofit option has the same face-to-face dimensions as the Micro Motion 7835 and 7845 Density meters. Thermal insulation The CDM is available with a soft, weather-proof insulating jacket that is easily fitted to all CDM versions. Accreditation and standards Accredited calibrations are in compliance with domestic and international standards. 3. Compact Density Meter August 2017.. ProLink III software: a configuration and service tool ProLink III software is an easy-to-use interface that allows you to view key process variables and diagnostics data for your meter. For more information on ordering the software, contact your local sales representative or email customer support at Operating principle Tube vibration Dual, parallel tubes vibrate at their natural frequency.)))

4 The natural frequency changes with the Density of the liquid inside the tubes. Density calibration Micro Motion transmitters accurately measure time period. Measured time periods are converted into Density readings using meter calibration coefficients. Minimum of 12 calibration points ensures optimum meter performance. A. Density (kg/m3). B. Time period = 1 / frequency C. [Time period]2 ( s2). 4 August 2017 Compact Density Meter Flow rate indication (velocity/volume flow). Measuring the twist in the vibrating tubes gives an indication of the liquid flow rate (velocity/volume flow). Performance specifications Density measurement CDM100P (Peak performance precision CDM100M (General purpose precision Specification Density meter) Density meter). Accuracy (liquid) kg/m3 ( g/cm3) kg/m3 ( g/cm3). Repeatability kg/m3 ( g/cm3) kg/m3 ( g/cm3).

5 Operating Density range 0 3000 kg/m3 (0 3 g/cm3) TPS transmitter version:0-1000 kg/m3. (0 1 g/cm3). Analog/Discrete transmitter versions: 0 3000 kg/m3 (0 3 g/cm3). Calibration range 300 1300 kg/m3 ( g/cm3) 300 1300 kg/m3 ( g/cm3). Process temperature effect (corrected)(1) kg/m3 per C kg/m3 per C. kg/m3 per 100 F kg/m3 per 100 F. Sensor maximum working pressure 248 bar (3600 psi) or flange limit 100 bar (1450 psi) or flange limit Process pressure effect (corrected)(2) kg/m3 per bar kg/m3 per bar kg/m3 per 100 psi kg/m3 per 100 psi (1) Process temperature effect is the maximum measurement offset due to process fluid temperature changing away from the reference calibration temperature of 20 C. (2) Process pressure effect is the maximum measurement offset due to process fluid pressure changing away from the reference calibration pressure of 1 bar.

6 Temperature measurement Specification Value Operating temperature range 58 F to +400 F ( 50 C to +204 C). Integral temperature sensor Traceable calibration technology : 100 RTD. Accuracy: BS1904 Class, DIN 43760 Class A ( + x Temp C). 5. Compact Density Meter August 2017. Specification Value Case temperature sensors(1) technology : 3 x 100 RTD. Accuracy: BS1904 Class, DIN 43760 Class B ( + x Temp C). (1) Case temperature sensors are used for environmental temperature effect correction in applications where the case temperature measurement does not need to be traceable and/or accredited. Where accreditation and measurement traceability are required, these sensors are used for diagnostics purposes only and do not perform any correction on the Density measurement. Case pressure Specification Value Maximum case working pres- 27 bar (389 psig).

7 Sure Typical burst pressure (case) 195 bar (2824 psig). Diagnostic flow rate indication (velocity/volume flow). Expected accuracy is within 5% of reading. Typical flow recommendations Flow rate Velocity Minimum 3 gpm (700 L/hr) ft/sec ( m/sec). Normal 11 gpm (2,500 L/hr) 5 ft/sec ( m/sec). Maximum 75 gpm (17,000 L/hr) 30 ft/sec (9 m/sec). Note For fluids that contain abrasive particles, velocity should be below 10 ft/s (3 m/s). Transmitter specifications Available transmitter versions For more information on the transmitter outputs and ordering codes, see the product ordering information. Analog Output channels Typical application A B C. General purpose measurement 4 20 mA + HART 4 20 mA (passive) Modbus/RS-485. DCS/PLC connection (passive). 6 August 2017 Compact Density Meter Processor for remote-mount Model 2700 FOUNDATION fieldbus transmitter Output channels Typical application A B C.

8 General purpose measurement Disabled Disabled Modbus/RS-485. DCS/PLC connection Discrete Output channels Typical application A B C. General purpose measurement with output 4 20 mA + HART Discrete output Modbus/RS-485. switch (passive). DCS/PLC connection Time Period Signal (TPS). Output channels Typical application A B C. Fiscal/Custody Transfer 4 20 mA + HART Time Period Signal Modbus/RS-485. Flow computer connection (passive) (TPS). The 4-20 mA output cannot be configured to output line Density . 2-wire TPS. Output channels Typical application A B C. Fiscal/Custody Transfer Disabled 4-wire 100 , RTD. Flow computer connection For the 2-wire transmitter version, TPS is superimposed on power lines. Local display Design Features Physical Segmented two-line LCD screen. Can be rotated on transmitter, in 90-degree increments, for ease of viewing.

9 Suitable for hazardous area operation. Optical switch controls for hazardous area configuration and display. Glass lens. Three-color LED indicates meter and alert status. 7. Compact Density Meter August 2017. Design Features Functions View process variables. View and acknowledge alerts. Configure mA and RS-485 outputs. Supports Known Density Verification (KDV). Supports multiple languages. Process measurement variables Variables Value Standard Density Time period Temperature Drive gain External temperature input External pressure input Flow rate (velocity/volume flow). Derived The derived output variables vary, depending on the application configuration of the meter. Referred Density (API Tables 53A, 53B). Referred Density (Concentration). Specific gravity (Concentration). %Alcohol by Volume (ABV). Alcohol proof API. Balling Baume Brix Plato %Mass %Solids Twaddle User-defined calculation output Note For the TPS transmitter version, process measurement variables are limited to Time Period, Temperature, and Flow.

10 For the 2-wire transmitter version, process measurement variables are limited to Time Period and Temperature. 8 August 2017 Compact Density Meter Additional communication options The following communications accessories are purchased separately from the meter. Type Description WirelessHART WirelessHART is available via the THUM adapter FOUNDATION fieldbus Remote-mount Model 2700 transmitter with FOUNDATION fieldbus One FOUNDATION fieldbus H1 connection provided . HART Tri-Loop Three additional 4-20 mA outputs are available via connection to a HART Tri- Loop Hazardous area approvals Ambient and process temperature limits are defined by temperature graphs for each meter and electronics interface option. Refer to the detailed approval specifications, including temperature graphs for all meter configurations, and safety instructions.