Example: biology

Rosemount Wireless Pressure Gauge WirelessHART Protocol

Product Data SheetJanuary 201800813-0100-4045, Rev DARosemount Wireless Pressure Gaugewith WirelessHART ProtocolThe Rosemount Wireless Pressure Gauge from Emerson utilizes industry-proven Rosemount Pressure sensor technology to deliver accurate, reliable Pressure information. It features up to 150x overpressure protection and two layers of process isolation providing a safer field environment. Rosemount sensor technology eliminates many Gauge challenges by replacing mechanical parts that inhibit traditional gauges from reporting or displaying the correct Pressure . The Rosemount Wireless Pressure Gauge features a large (114 mm) face for easy field visibility. It has up to a 10-year installed life, reducing costs and time involved with Wireless Pressure GaugeJanuary BenefitsReduce maintenance challenges Get up to 10 years of reliable readings through industry-proven Rosemount Pressure sensor technology Reduce common mechanical Gauge failures caused by vibration, overpressure and other environmental factors Have confidence in Pressure Gauge health with local indicator lightImprove personnel safety Keep people out of hazardous areas by minimizing operator rounds Gain peace of mind with overpressure ratings from to 150x and two layers of process isolationAccess Pressure data continuously Get accurate readings as frequently as once per minute with WirelessHART technology View Pressure reading locally with large (114 mm) dialCo

traditional gauges from reporting or displaying the correct pressure. This gauge features a large 4.5-in. (114 mm) face for easy field visibility. It has up to a 10-year installed life, reducing costs, and time involved with maintenance. ... Q16 Surface finish certificate for sanitary remote seals ★ …

Tags:

  Gauge, Sanitary

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of Rosemount Wireless Pressure Gauge WirelessHART Protocol

1 Product Data SheetJanuary 201800813-0100-4045, Rev DARosemount Wireless Pressure Gaugewith WirelessHART ProtocolThe Rosemount Wireless Pressure Gauge from Emerson utilizes industry-proven Rosemount Pressure sensor technology to deliver accurate, reliable Pressure information. It features up to 150x overpressure protection and two layers of process isolation providing a safer field environment. Rosemount sensor technology eliminates many Gauge challenges by replacing mechanical parts that inhibit traditional gauges from reporting or displaying the correct Pressure . The Rosemount Wireless Pressure Gauge features a large (114 mm) face for easy field visibility. It has up to a 10-year installed life, reducing costs and time involved with Wireless Pressure GaugeJanuary BenefitsReduce maintenance challenges Get up to 10 years of reliable readings through industry-proven Rosemount Pressure sensor technology Reduce common mechanical Gauge failures caused by vibration, overpressure and other environmental factors Have confidence in Pressure Gauge health with local indicator lightImprove personnel safety Keep people out of hazardous areas by minimizing operator rounds Gain peace of mind with overpressure ratings from to 150x and two layers of process isolationAccess Pressure data continuously Get accurate readings as frequently as once per minute with WirelessHART technology View Pressure reading locally with large (114 mm) dialContentsOrdering Information.

2 3 Specifications .. 10 Product Certifications ..13 Dimensional Drawings ..153 Rosemount Wireless Pressure GaugeJanuary Information Table 1. Rosemount Wireless Pressure Gauge Ordering InformationThe starred offerings ( ) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead typeWPGW ireless Pressure Gauge Dial ( mm) Gauge outputXWireless with user-configurable update rate, GHz DSSS, WirelessHART Product certificationsI1 ATEX Intrinsic Safety I2 INMETRO Intrinsic Safety I4 TIIS Intrinsic Safety I5US Intrinsically Safe I6 Canada Intrinsically Safe I7 IECEx Intrinsic Safety IMTechnical Regulations Customs Union (EAC) Intrinsic Safety NANo approval Measurement typeGGage AAbsolute CCompound VVacuum Process connection style(1)Connection styleIsolating diaphragm material 111/2-14 NPT male316L SST 121/2-14 NPT maleAlloy C-276 21G1/2 male (EN 837)316L SST 22G1/2 male (EN 837)Alloy C-276 01 Alternate process connection316 SST 02 Alternate process connectionAlloy C-276 Primary engineering unitApsi BkiloPascals (kPa) 4 Rosemount Wireless Pressure GaugeJanuary EmBar FMegaPascals (MPa) GinH2O Hkg/cm2 IftH2 OJmmH2 OKinHg LcmH2O McmHg NmmHg P(2)(3)

3 Percent of range (% of range) Scale rangesReference tables in Pressure scale ranges section for scale ranges by engineering unit. Options (include with selected model number)Secondary engineering unit (dual scale)DA(4)psi DB(4)kiloPascals (kPa) DD(4)bar DH(4)kg/cm2 DC(2)(3)(5)(6)Custom unitsManifolds assemblies(7)(8)(9)S5 Assemble to Rosemount 306 Integral ManifoldDiaphragm seal assembly(8)(9)(10)S1 Assemble to one Rosemount 1199 Diaphragm SealExtended product warrantyWR33-year limited warranty WR55-year limited warranty Mounting bracketB4 Bracket for 2-in. pipe or panel mounting, all SST Table 1. Rosemount Wireless Pressure Gauge Ordering InformationThe starred offerings ( ) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead Wireless Pressure GaugeJanuary configurationC1 Custom configuration Calibration certificationQ4 Calibration certificate Material traceability certificationQ8 Material traceability certification per EN 10204 NACE certificateQ15 Certificate of compliance to NACE MR0175/ISO 15156 for wetted materials Q25 Certificate of compliance to NACE MR0103 for wetted materials Pressure testingP1 Hydrostatic testing with certificate Surface finish(11)Q16 Surface finish certificate for sanitary Remote Seals Normal range indicationLKGreen, yellow and red sticker kit Alternate process connection(12)P01 Level flange (SST), 1-in.

4 ANSI Class 150P02 Level flange (SST), 1-in. ANSI Class 300P11 Level flange (SST), 2-in. ANSI Class 150P12 Level flange (SST), 2-in. ANSI Class 300P21 Level flange (SST), 3-in. ANS I Class 150P22 Level flange (SST), 3-in. ANSI Class 300P31 Level flange (SST), DIN-DN 50 PN 40P41 Level flange (SST), DIN-DN 80 PN 40W01 Level flange (SST), 1-in. ANSI Class 150, All welded constructionW02 Level flange (SST), 1-in. ANSI Class 300, All welded constructionW11 Level flange (SST), 2-in. ANSI Class 150, All welded constructionW12 Level flange (SST), 2-in. ANSI Class 300, All welded constructionW21 Level flange (SST), 3-in. ANSI Class 150, All welded constructionW22 Level flange (SST), 3-in. ANSI Class 300, All welded constructionW31 Level flange (SST), DIN-DN 50 PN 40, All welded constructionW41 Level flange (SST), DIN-DN 80 PN 40, All welded construction Table 1. Rosemount Wireless Pressure Gauge Ordering InformationThe starred offerings ( ) represent the most common options and should be selected for best delivery.

5 The non-starred offerings are subject to additional delivery lead Wireless Pressure GaugeJanuary 1. Model Number Ordering of Construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining available with Measurement Type available with Measurement Type available with Primary Engineering Unit P (Percent of Range). Primary Engineering Unit of A (psi) or D (bar). Custom Configuration model code C1 . Process Connection Style 11 or 12 . manifold and diaphragm seal assemblies cannot be Assemble-to items are specified separately and require a completed model Process Connection diaphragm seal assembly S1 process connection style 01 or 02 ModelDial sizeGauge outputProduct certificationMeasurement typeProcess connection stylePrimary engineering unitScale rangeOptionsWPG45XI1A11A6-digit numeric valueDA, definedEnd-user defined7 Rosemount Wireless Pressure GaugeJanuary scale rangesAdditional scale ranges available.

6 Contact Emerson for additional to 0000000-15/0 Gage/absolute0 to psiCompound(1)-30 inHg to 0 to scale will be in inHg and positive Pressure in psi. Only applies to CodeVacuum-bar to 0 or -kg/cm2 to 0000000-1/0 Gage/absolute0 to bar or 0 to kg/cm2 Compound-1 bar to 0 to bar or -1 kg/cm2 to 0 to Wireless Pressure GaugeJanuary (kPa)CodeVacuum-kPa to 0000000-100/0 Gage/absolute0 to kPaCompound-100 kPa to 0 to kPa0000404040000060606000010010010000015 0150150000160160160000200200200000250250 2500003003003000004004004000005005005000 0060060060000090090090000100010001000001 5001500150000160016001600002000200020000 024002400N/A0025002500N/A0040004000N/A00 50005000N/A0060006000N/A01000010000N/A02 500025000N/AmbarCodeVacuum-mbar to 0000000-1000/0 Gage/absolute0 to mbarCompound-1000 mbar to 0 to mbar000400400400000600600600001000100010 0000150015001500002000200020000030003000 3000004000400040000050005000500000600060 00600000900090009000 MegaPascals (MPa)CodeVacuum-MPa to to MPa to 0 to to 0000000-400/0 Gage/absolute0 to inH2 OCompound-400 inH2O to 0 to inH2O00020020020000030030030000080080080 09 Rosemount Wireless Pressure GaugeJanuary to 0000000-30/0 Gage/absolute0 to ftH2 OCompound-30 ftH2O to 0 to ftH2O00003535350000606060000070707000010 0100100000140140140000240240240000400400 400000500500500000700700700000900900N/Am mH2 OCodeVacuum-mmH2O to 0000000-10000/0 Gage/absolute0 to mmH2 OCompound-10000 mmH2O to 0 to mmH2O00750075007500040000400004000020000 0200000200000inHgCodeVacuum-inHg to 0000000-30/0 Gage/absolute0 to inHgCompound-30 inHg to 0 to inHg000012121200001515150000161616000020 202000003030300000606060000300300300cmH2 OCodeVacuum-cmH2O to 0000000-1000/0 Gage/absolute0 to cmH2 OCompound-1000 cmH2O to 0 to cmH2O000500500500000900900900cmHgCodeVac uum-cmHg to 0000000-75/0 Gage/absolute0 to cmHgCompound-75 cmHg to 0 to cmHg0001501501500007507507500040004000N/ A02000020000N/AmmHgCodeVacuum-mmHg to

7 0000000-750/0 Gage/absolute0 to mmHgCompound-750 mmHg to 0 to mmHg001500150015000075007500750004000040 000N/A200000200000N/APercent of range(1) will read 0-100 percent. Code selected is representative of the desired working Pressure range in Wireless Pressure GaugeJanuary specificationsMaterial selectionEmerson provides a variety of Rosemount products with various product options and configurations including materials of construction that can be expected to perform well in a wide range of applications. The Rosemount product information presented is intended as a guide for the purchaser to make an appropriate selection for the application. It is the purchaser s sole responsibility to make a careful analysis of all process parameters (such as all chemical components, temperature, Pressure , flow rate, abrasives, contaminants, etc.), when specifying product materials, options, and components for the particular application. Emerson is not in a position to evaluate or guarantee the compatibility of the process fluid or other process parameters with the product options, configuration, or materials of construction ( millimeter) Scale rangesFrom vacuum up to 4,000 psi (275 bar)Single scale considerationsThe number of major graduations is a direct result of the specified combination of primary engineering unit and scale range.

8 Dual scale considerationsThe number of major graduations on the inner scale is the direct result of the combination of primary engineering unit and secondary engineering unit. Process connections1/2-14 NPT male, G1/2 male (EN 837) and alternate process connectionsField Communicator connectionsCommunication terminals are accessible by removing of constructionHousingEngineered Polymer, NEMA 4X and IP66/67 Cover O-ringSilicone rubberProcess-wetted parts316L SST, Alloy C-276 Shipping lb ( kg)OptionsMounting bracket (Code B4) lb ( kg)Normal Range Indication (Code LK) lb (11 g) Rosemount 1199 Seal SystemsReference the Rosemount DP Level Product Data Sheet for shipping 306 Integrated ManifoldsReference the Rosemount Manifolds Product Data Sheet for shipping specificationsConformance to specification ( 3 [Sigma])Technology leadership, advanced manufacturing techniques, and statistical process control ensure specification conformance to at least 3 .AccuracyASME Grade 2A ( of span)Temperature limitsAmbient 40 to 185 F ( 40 to 85 C)Storage 40 to 185 F ( 40 to 85 C)Process 40 to 250 F ( 40 to 121 C)(1)(2)OptionsNormal Range Indication (Code LK)Ambient: 40 to 185 F ( 40 to 85 C)Storage: 70 F (21 C)Aplication: Minimum of 50 F (10 C) temperatures above 185 F (85 C) require de-rating the ambient limits by a :1 F (104 C) limit in vacuum service; 130 F (54 C) for pressures below Wireless Pressure GaugeJanuary connections/batteryReplaceable, non-rechargeable, V primary cell, lithium-thionyl chloride battery10 year battery life at reference conditions(1)Burst Pressure limit11,000 psi (758 bar)Minimum span limits for percent of range engineering unitThe maximum rangedown is 10:1.

9 The device maintains reference accuracy specification up to 6:1 rangedown. After 6:1 rangedown the reference accuracy decreases to 1% of zero trimAn offset adjustment to compensate for mounting position effects, up to 5 percent of SpanHumidity limits0-95% relative humidityElectromagnetic compatibility (EMC)Meets all industrial environment requirements of EN61326. Maximum deviation < 1 percent span during EMC indicationDevice status is indicated by local LED. Reference the Rosemount Wireless Pressure Gauge Quick Start Guide for further 62591 ( WirelessHART ), GHz DSSSW ireless radio (internal antenna) Frequency: GHz Channels: 15 Modulation: IEEE compliant DSSS Transmission: Maximum of 10 dBm EIRP Integrated omni-directional antennaWireless update rateWireless update rate is user-selectable from one minute to 60 minutes and is separate from local display. When Wireless is activated, the update rate defaults to once a effectNo significant effect when tested per IEC60770-1 or ASME requirementsIEC60770-1 high vibration level - field or pipeline: 10 60 Hz mm displacement peak amplitude/60 2000 Hz 3 conditions are 70 F (21 C), Stable operating Pressure with peri-odic changes, transmit rate of once per minute, and routing data for three additional network limitScale rangeMaximum working pressureMaximum overpressure limit5 30 psi ( 2 bar)30 psi (2 bar)750 psi ( bar)31 150 psi ( bar)150 psi ( bar)1,500 psi ( bar)151 800 psi ( bar)800 psi ( bar)1,600 psi ( bar)801 4,000 psi ( 275 bar)4,000 psi (275 bar)6,000 psi ( bar)Scale rangeSpan (6:1 ratio)Minimum span (10.)

10 1 ratio)5 30 psi ( 2 bar)5 psi ( bar)3 psi ( bar)31 150 psi ( bar)25 psi ( bar)15 psi ( bar)151 800 psi ( bar)134 psi ( bar)80 psi ( bar)801 4,000 psi ( 275 bar)667 psi ( bar)400 psi ( bar)Ambient temperature effect per 18 F (10 C)Scale rangeAmbient temperature effectWireless Pressure gaugeUp to 4,000 psi (275 bar) of spanWireless Pressure Gauge with remote sealUp to 4,000 psi (275 bar)See Instrument Toolkit Wireless Pressure GaugeJanuary Table 2. Wireless Connectivity Out of the BoxProducts in orderNetwork ID and Join Key are generatedWireless connectivity out of the boxRosemount Wireless Pressure GaugeAutomatically Manual activation requiredRosemount Wireless Pressure Gauge Customer specifiedActivatedRosemount Wireless Pressure Gauge and Emerson Wireless GatewayAutomatically (matching)Manual activation required13 Rosemount Wireless Pressure GaugeJanuary CertificationsRev: Union Directive InformationA copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide.


Related search queries