The GC Column - Agilent

How to Choose the Correct Type and DimensionSimon JonesApplication EngineerThe GC ColumnThings to TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?

Agilent J&W has over 50 different stationary phase offerings PLOT columns are primarily, but not exclusively, used for the analysis of gases and low boiling point solutes (i.e., boiling point of solute is at or below room temperature).

Tags:

  Agilent, Agilent j

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of The GC Column - Agilent

1 How to Choose the Correct Type and DimensionSimon JonesApplication EngineerThe GC ColumnThings to TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid?

2 How are we getting the Sample Injected? What is the sample Matrix?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What is the sample Matrix? Can we do sample clean up?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What is the sample Matrix? Can we do sample clean up? Is it an established method?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC?

3 Is it a Gas or a Liquid? How are we getting the Sample Injected? What is the sample Matrix? Can we do sample clean up? Is it an established method?--EPA, ASTM, USPG roup/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What is the sample Matrix? Can we do sample clean up? Is it an established method?--EPA, ASTM, USP What do we Know about the analytes?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected?

4 What is the sample Matrix? Can we do sample clean up? Is it an established method?--EPA, ASTM, USP What do we Know about the analytes? What else MAY be present in the sample?Group/Presentation TitleAgilent RestrictedPage 2 Things to Is it Volatile enough to chromatograph by GC? Is it a Gas or a Liquid? How are we getting the Sample Injected? What is the sample Matrix? Can we do sample clean up? Is it an established method?--EPA, ASTM, USP What do we Know about the analytes? What else MAY be present in the sample?Group/Presentation TitleAgilent RestrictedPage 2 Group/Presentation TitleAgilent RestrictedPage 3 CAPILLARY Column TYPESL iquid PhaseCarrier GasPorous Layer Open Tube (PLOT)Wall Coated Open Tube (WCOT)

5 Solid ParticlesCarrier GasPacked ColumnsPacked Column Designs and Materials1950 Introduction with the first gas chromatographsPacked Column AnatomyPacked Columns 1 12 m length Internal Diameter 4mm Tubing Stainless Steel, UltimetalTMSS, Glass, Nickel, PTFEP acked Column AnatomyPacked Columns 1 12 m length Internal Diameter 4mm Tubing Stainless Steel, UltimetalTMSS, Glass, Nickel, PTFE Packing Coated packing Inert, solid support (diatomaceous earth) coated with liquid stationary phase ( OV-1, SE-30, Carbowax 20M, FFAP) Porous packing Porous polymers (PoraPak Q, N, HayeSep Q, R, S, etc.) Porous carbons (Carboxens, Carbosieves, Carbotraps)Packed Column AnatomyPacked Columns 1 12 m length Internal Diameter 4mm Tubing Stainless Steel, UltimetalTMSS, Glass, Nickel, PTFE Packing Coated packing Inert, solid support (diatomaceous earth) coated with liquid stationary phase ( OV-1, SE-30, Carbowax 20M, FFAP) Porous packing Porous polymers (PoraPak Q, N, HayeSep Q, R, S, etc.)

6 Porous carbons (Carboxens, Carbosieves, Carbotraps)WCOT CapillaryPacked Column AnatomyPacked Columns 1 12 m length Internal Diameter 4mm Tubing Stainless Steel, UltimetalTMSS, Glass, Nickel, PTFE Packing Coated packing Inert, solid support (diatomaceous earth) coated with liquid stationary phase ( OV-1, SE-30, Carbowax 20M, FFAP) Porous packing Porous polymers (PoraPak Q, N, HayeSep Q, R, S, etc.) Porous carbons (Carboxens, Carbosieves, Carbotraps)WCOT CapillaryPLOT CapillaryGroup/Presentation TitleAgilent RestrictedPage 6 STATIONARY PHASE POLYMERSSiRROSiRROSiRRSiRROnmOSiRRnSilox aneSiarylene backboneR=methyl, phenyl, cyanopropyl, trifluoropropylOHCH2CH2O nPolyethylene GlycolStationary Phase% Substitution --polysiloxanesGroup/Presentation TitleAgilent RestrictedPage 7% = # of sites on silicon atoms occupiedBalance is methylOSiRRnSiloxaneR=methyl, phenyl, cyanopropyl, trifluoropropylStationary PhasePoly(ethylene) GlycolGroup/Presentation TitleAgilent RestrictedPage 8100% PEG (DB-WAX)

7 Less stable than polysiloxanesUnique separation characteristicsHOCH2CH2 OHnPoly(Ethylene) GlycolModifiedGroup/Presentation TitleAgilent RestrictedPage 9 Base deactivated (CAM) Acid Modified (DB-FFAP) Extended Temperature RangeWCOT Column Types Mexico 2012 Agilent J&W has over 50 different stationary phase offeringsPLOT columns are primarily, but not exclusively, used for the analysis of gases and low boiling point solutes ( , boiling point of solute is at or below room temperature).PLOT Column Types Agilent J&W PLOT columns begin with the designation of GS (Gas Solid) or HP-PLOT followed by a specific name CP (ChromPack) followed by name 10 stationary phases GS-OxyPLOT/ CP-Lowox GS-Alumina HP-PLOT Al2O3 M HP-PLOT Al2O3 S HP-PLOT Al2O3 KCl / CP-AL2O3/KCl HP-PLOT MoleSieve/ CP-Molsieve5A GS-CarbonPLOT/ CP-CarboBOND HP-PLOT Q / CP PoraBONDQ HP-PLOT U / CP-PoraBONDU GS-GasPro/ CP-SilicaPLOT GS-OxyPLOT: oxygenates HP-PLOT Molesieve: O2, N2, CO, Methane HP-PLOT Alumina and GS-Alumina: complex hydrocarbon gas matrices, ethylene and propylene purity, 1,4-butadiene HP-PLOT Q.

8 Freons, sulfides HP-PLOT U: C1 to C7 hydrocarbons, CO2, Polar Hydrocarbons GS-GasPro: freons, sulfurs, inorganic gases GS-CarbonPLOT: inorganic and organic gasesIntegrated Particle Trap PLOT ColumnsGroup/Presentation TitleAgilent RestrictedPage 12 Particle trap is on both ends of the columnOn the front end to help facilitate backflushingwithout blowing particles back into the inlet / valveSpecialty PhasesGroup/Presentation TitleAgilent RestrictedPage 13 Columns developed for particular applicationsExamples: DB-UI 8270D, DB-624UI <467>,DB-VRX, DB-MTBE, DB-TPH, DB-ALC1, DB-ALC2, DB-HTSimDis, DB-Dioxin, Select Low Sulfur, CP-Volamine, Select PAH, DB-EUPAH, DB-CLP1 & 2, DB-Select 624 UI 467, CP-LowOx, Select Permanent Inert PhasesGroup/Presentation TitleAgilent RestrictedPage 14DB-1msUIHP-1msUIDB-5msUIHP-5msUIDB-17m sUIDB-624 UIDB-Select 624UI 467DB-WAXUISame Selectivity, more Inertness!

9 Three Types Of Low Bleed PhasesGroup/Presentation TitleAgilent RestrictedPage 15 Phases tailored to mimic currently existing polymersExamples: DB-5ms, DB-35ms, DB-17ms, VF-1701ms New phases unrelated to any previously existing polymersExamples: DB-XLB Optimized manufacturing processesDB-1ms, HP-1ms, HP-5ms, VF-5msSiRROSiRROSiRRSiRROnmSiarylene backboneWhat is Column Bleed??? Back Biting Mechanism of Product FormationGroup/Presentation TitleAgilent RestrictedPage 16+RepeatSiSiSiSiSiSiSiOOOOOOOHCH3CH3CH3 CH3CH3CH3CH3CH3CH3CH3CH3CH3CH3CH3 OHSiSiOSiOSiOCH3CH3CH3CH3CH3CH3CH3CH3 SiOOOCH3H3CH3CH3 CSiSiCH3CH3 OOOOOOSiHOH3 CCH3CH3CH3CH3 SiSiSiSiSiSiCH3CH3CH3CH3CH3CH3CH3CH3CH3 Cyclic products are thermodynamically more stable!

10 DB-5ms StructureGroup/Presentation TitleAgilent RestrictedPage 17CH3 SiSiSiSiOOOODB-5 StructureDB-55% PhenylSiSiSiSiOODB-5ms to match DB-5CH3CH3CH3CH3CH3CH3CH3CH3CH3CH3CH3CH3 DB-35MS VS STANDARD 35% PHENYLB enzo[g,h,i]perylene, 1ngGroup/Presentation TitleAgilent RestrictedMonth ##, 35% PhenylDB-35msGroup/Presentation TitleAgilent RestrictedPage 19 Solid line:DB-5ms 30 m x .25 mm x .25 mmDashed line: DB-5 30 m x .25 mm x .25 mmOven:60oC isothermalCarrier gas: H2at 40 cm/sec1: Ethylbenzene2: m-Xylene3: p-Xylene4: o-XyleneChromatograms are offset on x-axisfor easier viewingWhy is stationary phase type important?