66 The management of surgical complications ofpacemaker and implantable cardioverter -defibrillatorsStephen Pavia, MD, and Bruce Wilkoff, MDThe rate of implantation of pacemakers and implantablecardioverter-defibrillators (ICDs) is ever-increasing. The relativeease of device implantation utilizing a relatively simple, expedi-tious, percutaneous approach, without the requirement forgeneral anesthesia or long recuperation times, has fueledenthusiasm for implantation . However, the complication risk isever-present and forms the subject of this pragmatic review,which is limited to the management of only the surgical compli-cations of device implantation . The management of surgicalcomplications related to the implantation of pacemakers andICDs should include (1) awareness of potential complications ,(2) a meticulous approach to the implantation procedure toavoid complications , (3) approach to diagnosis and (4)specific therapy.

66 The management of surgical complications of pacemaker and implantable cardioverter-defibrillators Stephen Pavia, MD, and Bruce Wilkoff, MD The rate of implantation

Tags:

  Management, Surgical, Complications, Pacemaker, Implantable, Defibrillator, Implantation, Cardioverter, The management of surgical complications of pacemaker, The management of surgical complications of pacemaker and implantable cardioverter defibrillators

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of The management of surgical complications of pacemaker …

1 66 The management of surgical complications ofpacemaker and implantable cardioverter -defibrillatorsStephen Pavia, MD, and Bruce Wilkoff, MDThe rate of implantation of pacemakers and implantablecardioverter-defibrillators (ICDs) is ever-increasing. The relativeease of device implantation utilizing a relatively simple, expedi-tious, percutaneous approach, without the requirement forgeneral anesthesia or long recuperation times, has fueledenthusiasm for implantation . However, the complication risk isever-present and forms the subject of this pragmatic review,which is limited to the management of only the surgical compli-cations of device implantation . The management of surgicalcomplications related to the implantation of pacemakers andICDs should include (1) awareness of potential complications ,(2) a meticulous approach to the implantation procedure toavoid complications , (3) approach to diagnosis and (4)specific therapy.

2 With a clear understanding of the acceptedimplant indications and potential complications , and a meticu-lous approach to the implant and post implant follow up, theincidence of complications can be minimized. Curr Opin Cardiol2000, 16:66 71 2001 Lippincott Williams & Wilkins, Cleveland Clinic Foundation, Cleveland, Ohio, to Bruce Wilkoff, MD, Director of Cardiac Pacing andTachyarrhythmia Devices, The Cleveland Clinic Foundation, 9500 EuclidAvenue, Cleveland, OH 44195, USA;e-mail: Opinion in Cardiology2001, 16:66 71 AbbreviationsICDimplantable cardioverter -defibrillatorsSVCS superior venacava syndromeISSN 0268 4705 2001 Lippincott Williams & Wilkins, rate of implantation of pacemakers and implantablecardioverter-defibrillators (ICDs) is increase parallels the potential for more wide-spread indications pending the results of ongoing trialsof pacing in heart failure and sudden death relative ease of device implantation utilizing a rela-tively simple, expeditious, percutaneous approach,without the requirement for general anesthesia or longrecuperation times, has fueled enthusiasm for implanta-tion.

3 However, the complication risk is ever-present andforms the subject of this may be surgical /hardware, program-ming/software, or normal device function specific complications include failure to pace(Table 1), failure to sense (Table 2), pulse generatorfailure (Table 3), pacemaker syndrome and pacemakermediated tachycardia. complications specific to the ICDinclude ineffective therapy, failure to deliver therapy,multiple shocks (both appropriate and inappropriate), anddevice proarrhythmia. Both pacemakers and ICDs maybe subject to crosstalk, electromagnetic interference, andpatient psychological issues. This review, however, limitsdiscussion to surgical complications . Managing surgicalcomplications related to the implantation of pacemakersand ICDs should include (1) awareness of potentialcomplications, (2) a meticulous approach to the implanta-tion procedure to avoid complications , (3) approach todiagnosis, and (4) specific recent publication of device complication rates showedan acute pacemaker implant complication rate of 4 to 5%[5].

4 implantable cardioverter -defibrillators lead complica-tions occur in 10 to 16% of patients [10], and even higherwhen programming and potential psychological complica-tion data are considered. The nonthoracotomy implanttechnique portends a lower complication rate than thenow superceded open thoracotomy, epicardial patchapproach. Hardware and implantation related complica-tions often require further surgical intervention. The inci-dence of device related implant complications has beenshown to be implanter-experience related [3,5] andsimilar with single or dual chamber implantation [4].An exhaustive list of potential complications is outlinedin Table 4. Each surgical complicationand specificmanagement strategy will be dealt with in will be divided into venous access related,lead related, and pocket related. Infection is anextremely important complication that is discussedseparately.

5 Unless specifically stated otherwise, discus-sion incorporates complications applicable to both pace-maker and ICD access related complicationsPneumothoraxThis complication occurs uncommonly and is directlyrelated to operator experience, the difficulty of thesubclavian puncture, and is almost eliminated using thecephalic cut-down technique. However, these traditionalcomparisons may become obsolete as the axillary veincannulation technique [16] threatens to eliminate thiscontroversy. Often the pneumothorax is asymptomaticand noted on routine follow-up plain chest complication results from trauma to the greatvessels rather than the lung. The risk can be minimizedby direct inward and outward passes of the punctureneedle rather than a side-to-side, potentially lacerat-ing movement. If an arterial puncture is performed,recognition, withdrawal, and digital pressure are impor-tant.

6 Never cannulate the artery with the introducer, asituation that should indicate vascular surgical this situation by always checking the fluoroscopicpath of the guidewire into the inferior venacava beforeintroducer embolismDeep inspiration at the time of central venous access maycause significant air to be drawn into the venous systemdue to the physiological negative pressure developed. Itcan be prevented through operator care and using intro-ducers with hemostatic valves. The diagnosis is obviousbecause it is heralded by a hissing sound as the air issucked in and with the fluoroscopic confirmation thatfollows. Patients are surprisingly tolerant of this , respiratory distress, hypotension, and arterialoxygen desaturation may occur depending on the size ofthe embolus and 100% oxygen should be administeredalong with ionotropic support in some cases [7].

7 Aspirationof the embolus from the right heart has also been success-ful. However, usually no therapy is required, as the air isfiltered and consequently absorbed in the complicationsPerforationThe lead may perforate any of the great veins, the atria,or ventricle during the implant procedure. This compli-cation almost always occurs in the cardiac chamber onlead manipulation or fixing a screw in lead, and conse-quently bleeds into the pericardial space. Perforationusually occurs without serious sequelae. However, amost devastating manifestation is cardiac tamponade,which requires prompt diagnosis and percutaneouspericardiocentesis, possibly followed by surgical inter-vention if the bleeding persists. An initial suspicionshould be evaluated by fluoroscopy of the heart border,an immediately available diagnostic of the diagnosis is obtained by emergentbedside transthoracic echocardiography.

8 This proce-dure is usually performed after emergency placementof the pericardiocentesis catheter because, on suspicionof this diagnosis, definitive therapy should not bedelayed. Less dramatic manifestation may be poorManaging surgical complications Pavia and Wilkoff67 Adapted from [18].Table 1. Causes of loss of captureLead dislodgment or malpositionMicro/macro-dislodgmentPerfor ationCapture threshold elevationLead maturationMetabolic/electrolyte abnormalityDrugsCardiac disease progression (infarction/cardiomyopathy)Local electrode myocardial interface damageElectrocautery/defibrillationBatte ry depletionCircuitry failureInappropriate programmingToo low a safety marginPseudo malfunctionRecording artifact mimicking pacing stimulusIsoelectric evoked potentialFunctional non-capture output delivered during refractory period (true or functional undersensing)Adapted from [18].Table 2. Causes of undersensingPoor intrinsic signal from implantSignal amplitude and/or slew rateDeterioration in signal over timeLead maturationDisease progression (cardiomyopathy, myocardial infarction, newleft bundle branch block)Respiratory or motion variationEctopic complexes not present at implantTransient decrease in signal complexPost cardioversion/defibrillationMetabolic derangement (eg,hyperkalemia)Component malfunctionSensing circuit abnormalityStuck reed switchBattery depletionMechanical lead dysfunctionInsulation failurePartial open circuitPseudomalfunctionRecording artifactNormal device function misintrepretationTriggered modeFusion and pseudofusion beatsFunctional undersensingLong refractory periodsBlanking periodSafety pacingOversensingOversensing intiated functional undersensingimplant pacing threshold and comparative differingunipolar versus bipolar sensing.

9 Rarely, trauma to thegreat veins above the pericardial reflection may causebleeding directly into the mediastinum. This is much moreof a concern when extracting leads than implantingthem and is an emergent indication for open chestsurgery, which may still be inadequate to save thepatient s life. These complications are most commonlyseen with lead extraction a recent series, this complication occurred in ofpacemaker implants [5,6]. With regard to ICDs,dislodgment may be associated with inadequate sensingor therapy. Dislodgment usually occurs very early post-implant (usually 24 48 hours) but may occur up to 3months after initial implant [6]. Intermittent under-sensing or loss of capture on post-implant telemetryshould prompt consideration of this is obtained by device interrogationrevealing adverse changes in sensing and pacing thresh-olds compared to implant values, and often CXRevidence of macro-dislodgment.

10 Lead repositioning orreplacement is lead malposition is diagnosed by unacceptableimplant pacing, sensing, and/or defibrillation thresholds,and remedied with repositioning during the initialimplant. Often, particularly with atrial lead site selectionin a patient post cardiac surgery and/or with significantstructural heart disease, extensive intracardiac electro-gram mapping may be required to determine the bestsite. In other cases, malposition may not be so immedi-ately obvious. The presence of an intracardiac abnormal-ity such as an atrial or a ventricular septal defect mayallow passage of the lead from the right to the left occurrence is more likely with the ventricular lead,and because fluoroscopy during simple lead implantationis almost limited to the AP projection. Certainly, onemust be alert to the resultant-paced ECG-QRS complexmorphology. If right bundle, a left-sided ventricular leadposition should be excluded.