Page 1/8 NU-CPJ-FE-1106 SCAIME - BP 501 - F-74105 ANNEMASSE CedexTel. : +33 (0)4 50 87 78 64 - Fax : +33 (0)4 50 87 78 46 Web : - Email : Conditionneur de signal analogique Analog signal conditioner CPJ - CPJ2S Notice d utilisation User manual Page 2/8 NU-CPJ-FE-1106 1. Introduction Introduction Le CPJ est un conditionneur analogique de capteurs. Il est principalement destin aux capteurs pont de jauges tels que capteurs de pesage, force, couple ou pression. La version CPJ2S int gre la gestion de 2 seuils r glables. The CPJ device is a sensor analog conditioner. It is mainly used with strain gauges sensors as load cells, torque sensors or pressure sensors. The CPJ2S version includes management of 2 adjustable set points. 2. Pr sentation presentation CPJ Version Rail DIN CPJ, DIN rail model CPJ Version Carte CPJ, Board model CPJ Version bo tier IP65 CPJ, IP65 Housing model 3.

Page 5/8 NU-CPJ-FE-1106 §4.6. Réglage du zéro – Zero adjustment • Vérifier qu’aucune charge (excepté l’infrastructure dont on désire annuler la tare) n’est appliquée sur le capteur. • Régler le zéro à l’aide du potentiomètre P2, puis affiner le réglage avec P3.

Tags:

  Zero

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of CPJ - CPJ2S - Electronics | Transmitters Controllers

1 Page 1/8 NU-CPJ-FE-1106 SCAIME - BP 501 - F-74105 ANNEMASSE CedexTel. : +33 (0)4 50 87 78 64 - Fax : +33 (0)4 50 87 78 46 Web : - Email : Conditionneur de signal analogique Analog signal conditioner CPJ - CPJ2S Notice d utilisation User manual Page 2/8 NU-CPJ-FE-1106 1. Introduction Introduction Le CPJ est un conditionneur analogique de capteurs. Il est principalement destin aux capteurs pont de jauges tels que capteurs de pesage, force, couple ou pression. La version CPJ2S int gre la gestion de 2 seuils r glables. The CPJ device is a sensor analog conditioner. It is mainly used with strain gauges sensors as load cells, torque sensors or pressure sensors. The CPJ2S version includes management of 2 adjustable set points. 2. Pr sentation presentation CPJ Version Rail DIN CPJ, DIN rail model CPJ Version Carte CPJ, Board model CPJ Version bo tier IP65 CPJ, IP65 Housing model 3.

2 Caract ristiques g n rales General specifications Alimentation Power supply 24+/-4 VDCC lasse de pr cision accuracy class rive thermique de z ro zero temperature drift < C (FS*)D rive thermique de gain Span temperature drift < C (FS*)Plage de temp rature de fonctionnement Operating temperature range +70 CAlimentation capteur (commutable par cavalier) Sensor supply voltage (jumper selection) 3, 5, 10 VImp dance min. capteur - alimentation 3, 5V - alimentation 10V Min. sensor impedance - voltage 3, 5V - voltage 10V 80160 R glage du gain Span adjustment .. 12mV/VConsommation max. CPJ/ CPJ2S Max. supply current CPJ/ CPJ2S 120/170mASortie tension Voltage output +/-10, 0-10 VSortie courant Current output 4-20mAImp dance de charge en sortie tension Load impedance (Voltage output) >2000 Imp dance de charge en sortie courant Load impedance (Current output) <500 Charge capacitive en sortie Output capacitive load <1nFFiltre (commutable par cavalier) Filter (jumper selection) 10 HzBande passante (jusqu' ) Bandwidth (up to) 20 KHz*FS = Pleine Echelle *FS = Full Scale Page 3/8 NU-CPJ-FE-1106 4.

3 Position des raccordements et organes de r glage Wiring and setting devices positionning Raccordements des capteurs Sensors connection Le CPJ peut conditionner jusqu 4 capteurs 350 connect s en parall le l aide d un bo tier de raccordement ALCJB. The CPJ is able to manage up to 4 load cells (350 ) connected in parallel through an ALCJB junction box. Connexion capteur Sensor connection Capteur 4 fils. Les 2 cavaliers restent en place. Capteur 6 fils. Les 2 cavaliers sont enlev s. 4 wire sensor. The two jumpers stay in place. 6 wire sensor. The two jumpers are removed. Tension d alimentation capteur sensor supply voltage Positionner le cavalier l emplacement correspondant la tension choisie : 3 VDC, 5 VDC ou 10 VDC.

4 Les capteurs sont g n ralement aliment s en 5V (par d faut). La tension 3V est utile pour certains capteurs particuliers, la tension 10V est recommand e pour les capteurs ayant une faible sensibilit (ou signal de sortie). Attention : - Pour une alimentation 10 VDC, seulement 2 capteurs 350 peuvent tre raccord s. - Pour des capteurs utilis s en zone ATEX et prot g s par des barri res zener, ne pas utiliser la tension 10V. Set the jumper in position corresponding to the selected voltage : 3 VDC, 5 VDC or 10 VDC. The load cells are generally supplied in 5V (default value). 3V voltage is useful for some particular sensors, 10V is recommended for low sensitivity (or low output signal) sensors. Warning : - With 10 VDC supply voltage, only 2 load cells (350 ) can be connected.

5 - For load cells used in ATEX area and protected by Zener barriers, do not use 10V supply voltage. 2 Cavaliers en place2 jumpers set in placeGnd Sens - Sens + Ex + Sig - Sig + Ex - Capteur 4 fils 4 wire sensor A+/Ex+ M-/Sig- M+/Sig+ A-/Ex- 7 6 5 4 3 2 1 Gnd Sens - Sens + Ex + Sig - Sig + Ex - Capteur 6 fils 6 wire sensorR-/Ref- R+/Ref+ A+/Ex+ M-/Sig- M+/Sig+ A-/Ex- 2 Cavaliers enlev s2 jumpers removed7 6 5 4 3 2 1 P1 P2 P3 R glage z ro zero adjustment S lection sensibilit capteur Sensor sensitivity selection R glage gain Span adjustment S lection filtrage Filtering selection Voyant d alimentation Power supply led Bouton poussoir Cal Cal push button S lection capteur 4 / 6 fils4/6 wire sensor selection et 5. Connexion capteurSensor connection et 5.

6 S lection tension d alimentation capteur Sensor supply voltage selection et 5. Connexion alimentation et signal de sortie Power supply and output signal connection 8 9 10 11 12 1 2 3 4 5 6 7 3V 5V 10V Page 4/8 NU-CPJ-FE-1106 Alimentation et sorties analogiques Power supply and analog outputs Les sorties analogiques courant 4/20mA et tension 0/10V peuvent tre utilis es simultan ment. Les bornes Gnd sont interconnect es sur le circuit. 4/20mA current analog output and 0/10V voltage analog output can be used simultaneously. The Gnd points are internally connected in the circuit board. Filtrage du signal capteur Sensor signal filtering Le filtre est du type Passe-bas du 2 me ordre avec une fr quence de coupure 10Hz, cavalier en position Filtr (Par d faut).

7 Le filtre permet g n ralement d liminer les perturbations g n r es par des vibrations de la Si un temps de r ponse rapide est n cessaire passer le cavalier en position Non filtr . The filter is a second order low-pass type with 10Hz cutting frequency. The filter is generally used to eliminate troubles caused by vibrations of the installation. If a quick response time is necessary, it s recommended to disable the filter. S lectionner le filtrage Enable filtering : S lection sensibilit capteur Sensor sensitivity selection Pour obtenir un signal correct sur la sortie analogique (0/10V ou 4/20mA), positionner le cavalier de configuration selon le tableau suivant : Nota : La sensibilit du capteur est indiqu e dans la fiche de contr le jointe au capteur. To get the correct signal on the analog output (0/10V or 4/20mA), set the sensor sensitivity jumper according to the following table: Nota : The load cell sensitivity is indicated on the quality control sheet attached with the load cell.

8 Position du cavalier de configuration Positionning of configuration jumper Pour la position X, mettre le cavalier entre 8 et 9. For X position, put the jumper between 8 and 9. Sensibilit capteur (mV/V) Sensor sensitivity (mV/V) Alimentation capteur 3V Sensor power supply 3V Alimentation capteur 10 VSensor power supply 10V Alimentation capteur 5V Sensor power supply 5V Position cavalier Jumper position X 1 2 3 4 5 6 7 8 9 Alimentation 24 VDC 4 Power supply 24 VDC 4 Sortie analogique tension 0/10V ou +/-10 VVoltage analog output 0/10V or +/-10 VSortie analogique courant 4/20mACurrent analog output 4/20mA+ 24V Gnd Gnd Iout Vout 1211109 8 Exemple - Example Capteur 500kg avec sensibilit 2mV/V.

9 Alimentation capteur : 5V Load cell with 500kg capacity and 2mV/V cell supply voltage : 5V Charge maximale mesurer : 200kg pour 10V Maximum load to be measured : 200kg for 10V Sensibilit pour une charge de 200kg = (200kg/500kg) x 2mV/V = Sensitivity for a 200kg load = (200kg/500kg) x 2mV/V = On placera le cavalier en position 3. The jumper will be set in position number 3 123456789123456789 Position XX position Position 1 Position 1 Cavalier de validation du filtrageJumper to set filterNon filtr without filter Filtr - filter Page 5/8 NU-CPJ-FE-1106 R glage du z ro zero adjustment V rifier qu aucune charge (except l infrastructure dont on d sire annuler la tare) n est appliqu e sur le capteur. R gler le z ro l aide du potentiom tre P2, puis affiner le r glage avec P3.

10 - Sortie analogique 0/10V : Le voltm tre doit afficher 0V. - Sortie analogique 4/20mA : l amp rem tre doit afficher 4mA. Verify that no load (except the load receptor) is applied on the load cell. Approach the zero with P2 potentiometer and make a fine adjustment with P3 potentiometer. - Analog output 0/10V : voltmeter displays 0V. - Analog output 4/20mA : ampere meter displays 4mA. R glage du gain Span adjustment Apr s avoir fait le r glage du z ro, appliquer une charge connue sur le capteur. R gler le gain avec le potentiom tre P1 jusqu obtenir un signal correspondant la charge appliqu e. Retirer la charge du capteur et v rifier le r glage du z ro. Si n cessaire, refaire un r glage du z ro suivi d un nouveau r glage de gain. After zero adjustment, apply a known load on the load cell.