Example: bankruptcy

DINAMAP - Frank's Hospital Workshop

Revision AProCare Patient Monitor Service Manual2009381-001 DINAMAP ProCare MonitorService ManualThis manual is for DINAMAP ProCare Monitors models 100, 200, 300, and 400, with or without printers. ProCare 100: BP, Pulse ProCare 200: BP, Pulse, and Temp ProCare 300: BP, Pulse, and SpO2 ProCare 400: BP, Pulse, Temp, and SpO2 The model of the Monitor determines which parameters are in your monitor. Please refer to applicable sections. Reissues and UpdatesChanges occurring between issues are addressed through Change Information Sheets, Addendums, and replacement pages.

T-2 ProCare Monitor Revision A 2009381-001 NOTE: The information in this manual only applies to ProCare Monitor.It does not apply to earlier Monitors. Due to continuing product innovation, specifications in this manual are subject to change without notice.

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of DINAMAP - Frank's Hospital Workshop

1 Revision AProCare Patient Monitor Service Manual2009381-001 DINAMAP ProCare MonitorService ManualThis manual is for DINAMAP ProCare Monitors models 100, 200, 300, and 400, with or without printers. ProCare 100: BP, Pulse ProCare 200: BP, Pulse, and Temp ProCare 300: BP, Pulse, and SpO2 ProCare 400: BP, Pulse, Temp, and SpO2 The model of the Monitor determines which parameters are in your monitor. Please refer to applicable sections. Reissues and UpdatesChanges occurring between issues are addressed through Change Information Sheets, Addendums, and replacement pages.

2 If a Change Information Sheet does not accompany this manual, it is correct as and OmissionsIf errors or omissions are found in this manual, please notify:GE Medical Systems Information TechnologiesTechnical Publications4502 Woodland Corporate BoulevardTampa, FL 336141-877-274-8456 Part No. 2009381-001 AThe content of this document including all figures and drawings is proprietary information of GE Medical Systems Information Technologies, provided solely for purposes of operation, maintenance or repair, and dissemination for other purposes or copying thereof is prohibited without prior written consent by GE Medical Systems Information Technologies, Tampa, may show design models.

3 Production units may incorporate of Warnings and CautionsA general warning is a statement that alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the misuse of the device. A warning relates to steps in a general caution is a statement that alerts the user to the possibility of a problem with the device associated with its use or misuse. Such problems include device malfunction, device failure, damage to the device or damage to other property. A caution relates to steps in a procedure. GE Medical Systems Information Technologies 2002, TAMPA, FL 33614 Printed in the All rights Patient Monitor Service ManualRevision A2009381-001 United StatesGE Medical Systems Information Technologies4502 Woodland Corporate BoulevardTampa, FL 33614 GermanyGE Medical Systems Information Technologies GmbHMunzinger Strasse 3, - 7911 Frieburg, GermanyDINAMAP ProCare MonitorService Manual2009381-001 Revision AT-2 ProCare MonitorRevision A2009381-001 NOTE: The information in this manual only applies to ProCare Monitor.

4 It does not apply to earlier Monitors. Due to continuing product innovation, specifications in this manual are subject to change without notice. GE Medical Systems Information Technologies, 2002. All rights AProCare Patient Monitor Service Manual2009381-001 Table of Contents1 Introduction .. Scope of Manual .. Manual Changes .. Service Policy .. Service Contracts .. Assistance .. Return to Factory Repair Service .. Service Loaners .. Repair Parts .. Replacement Accessories .. Product Description.

5 General Description .. 82 Product Description .. Introduction .. Product Configurations .. Controls, Indicators, and Connectors .. ProCare Monitor Rear Panel Connections .. Front Panel Controls and Indicators .. Host Communications Connector .. DB15 Connector Pin Assignments .. Compatible Parts .. Specifications .. 123 Principles of Operation .. Introduction .. Overall Principles of Operation .. SpO2 (Model 300 and 400) .. 4 ProCare Patient Monitor Service ManualRevision A2009381-001 Table of Cuff Blood Pressure (BP) and Pulse.

6 Temperature (Model 200 and 400) .. Host Communication Port .. Functional Description .. Main Board PWA .. User Interface (UI) Board PWA .. SPO2 PWA .. Printer .. Pneumatic Valve/Manifold (PVM) .. Optical Switch .. 74 Calibration & Maintenance .. Introduction .. Configuring Your ProCare Monitor .. Unpacking and Preparation for Installation .. Set the Date and the Clock .. Parameter Level Functional Testing .. Periodic Maintenance .. As Required .. Annual Procedures .. Care of the Storage Battery.

7 Battery Charging .. Safety Testing .. 9 Notes on Electrical Safety Testing of the ProCare Monitor: .. Temp Circuit Leakage Test .. SpO2 Circuit Leakage Test .. ProCare Patient Monitor Parameter Tests .. SETUP .. Leakage Testing .. Pressure Transducer Verification .. Pressure Transducer Calibration .. Overpressure Verification .. Button Testing .. LED Tests .. External DC Verification .. NIBP Determination .. NIBP Overpressure Verification .. 16 Revsion AProCare Patient Monitor Service Manual2009381-001 Table of Temperature (Perform if equipped with Temp module).

8 SpO2 (Perform only if equipped with SpO2 module) .. Printer Output Test .. Communication Port Test .. Alarm Code Interpretation .. System Failures .. 19 Appendix A - Test Results Form .. 21 Appendix B - Connectivity .. 23 Appendix C - Display Cover: Removal, Installation .. 25 Appendix D: Replacement Parts and Assemblies .. 275 Assembly Drawings & ProCare Schematics ProCare Patient Monitor Service ManualRevision A2009381-001 Table of ContentsRevision AProCare Patient Monitor Service Manual1-12009381-0011 Introduction1-2 ProCare Patient Monitor Service ManualRevision A2009381-0011-3 ProCare Patient Monitor Service ManualRevision Scope of ManualThis Service Manual provides service and parts repair information about the DINAMAP ProCare Monitor.

9 This manual is intended for use by trained service technicians who are familiar with electromechanical devices and analog circuit qualified technicians should perform repairs to this information about operating the Monitor in a clinical environment, refer to the separate Operations Service Manual consists of the following five sections:Section 1 describes this volume and tells you how to use it. Information is also provided about the physical and functional characteristics of the Monitor, and how to get assistance in the event the unit fails to function 2 provides a general overview of the ProCare Monitor including the user controls, external connections and product/parameter 3 presents principles of operation for the Monitor, including an overall system description and principles of operation at the component 4 provides information about periodic and corrective maintenance of the Monitor and part list.

10 Replacement part lists. Procedures include module performance testing, and calibration procedures. Information is also provided to facilitate isolating faults to the subassembly 5 provides component information about the Monitor, including assembly drawings and electrical reduce the risk of electric shock, do not remove cover or back of any component. Refer servicing to qualified service Patient Monitor Service ManualRevision A2009381-001 Introduction: Manual Manual ChangesIf, in the normal course of using this manual, you notice errors, omissions, incorrect data, or if you can suggest comments that may help improve this manual, please send suggestion to.