EasyLine Continuous Gas Analyzers Model EL3020

EasyLine Continuous Gas Analyzers Model EL3020 . Data Sheet 10/24 EN November 2005. Detectors with different measurement principles for Integral gas feed (optional). numerous process and emission monitoring applications Simple menu-driven operator interface Up to five measurement components per gas analyzer Clear-text status messages Suitable for measuring flammable gases Configuration of rarely required functions with included Automatic calibration including pump and valve control configuration program Simplified calibration with air or integral calibration cells Housing version for 19-inch rack mounting (3 height units, eliminating the need for test gas cylinders IP20).

EasyLine Continuous Gas Analyzers Model EL3020 Data Sheet 10/244.10 EN November 2005 Detectors with different measurement principles for numerous …

Tags:

  Model, Continuous, Analyzer, Easyline continuous gas analyzers model el3020, Easyline, El3020

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of EasyLine Continuous Gas Analyzers Model EL3020

1 EasyLine Continuous Gas Analyzers Model EL3020 . Data Sheet 10/24 EN November 2005. Detectors with different measurement principles for Integral gas feed (optional). numerous process and emission monitoring applications Simple menu-driven operator interface Up to five measurement components per gas analyzer Clear-text status messages Suitable for measuring flammable gases Configuration of rarely required functions with included Automatic calibration including pump and valve control configuration program Simplified calibration with air or integral calibration cells Housing version for 19-inch rack mounting (3 height units, eliminating the need for test gas cylinders IP20).

2 Customizable analog outputs, digital inputs and outputs Modular design for ease of service Modbus interface Self-monitoring function indicates when maintenance is required Contents Overview of the Gas Analyzers Page Measuring Technology ( Analyzers ). Overview of the Gas Analyzers 3 The following Analyzers are available for selection: Infrared Photometer Uras26 4 Uras26 infrared photometer for the measurement of infrared- Oxygen analyzer Magnos206 6 active gas components CO, NO, SO2. Thermal Conductivity analyzer Caldos27 8 Magnos206 oxygen analyzer for the measurement of Electrochemical Oxygen Sensor 10 O2 in process gas or in N2. Integral Gas Feed 11 Caldos27 thermal conductivity analyzer for the measurement General Data 12 of binary gas mixtures with different thermal conductivity Ar in O2, H2 in Ar, CH4 in N2.

3 Electrical Connections 14. Gas Connections 16 Electrochemical oxygen sensor for the measurement of O2. Certifications 18 The electrochemical oxygen sensor can only be used in combination with the Uras26 infrared photometer. The Magnos206 oxygen analyzer and the Caldos27 thermal conductivity analyzer can also be used in combination with the Uras26 infrared photometer. Integral Gas Feed The integral gas feed (option) is available in two versions. It either includes the micro-filter and flow sensor modules or the micro-filter, solenoid valve, pump, coarse filter, capillary tube and flow sensor modules. Electrical Interfaces The electrical interfaces for the output of measured values and communication with external systems include The integrated Ethernet-10/100 BASE-T interface (for service and configuration purposes).

4 As well as the integrated I/O modules depending on the functional range and order Analog output module with 2 analog outputs, Digital I/O module with 4 digital inputs and 4 digital outputs and Modbus module with RS485 and RS232 interfaces. Housing Design The housing for the EL3020 gas analyzer Model is designed as a 19-inch housing with 3 height units and degree of protection IP20. Note Regarding the Performance Characteristics of the Analyzers The performance characteristics of the Analyzers have been determined according to the international standard IEC 1207-1: 1994 Expression of performance of gas Analyzers . They are based on N2 as the associated gas. Compliance with these characteristics when measuring other gas mixtures can only be assured if their composition is known.

5 10 EN November 2005 EasyLine EL3000 Series Data Sheet 3. Infrared Photometer Uras26. Measurement Principle Stability Non-dispersive infrared absorption in the = 8 m The following data apply only if all influence factors ( flow wavelength range rate, temperature, atmospheric pressure) are constant. Photometer to measure up to 4 components with 1 or 2 beam Linearity Deviation paths and 1 or 2 receivers per beam path in one gas path or two 1 % of span separate gas paths. Repeatability % of span Sample Components and Measurement Ranges Zero Drift 1 % of span per week The analyzer has one physical measurement range per sample component. The smallest measurement ranges are shown in the Sensitivity Drift following table.

6 1 % of measured value per week Sample Component Smallest Measurement Range Output Fluctuation (2 ). % of span at electronic T90 time (static / dynamic). CO 0 100 ppm = 5 / 0 sec CO2 0 100 ppm Detection Limit (4 ). NO 0 150 ppm % of span at electronic T90 time (static / dynamic). SO2 0 100 ppm = 5 / 0 sec N2O 0 100 ppm CH4 0 100 ppm Influence Effects Measurement Range Limits Flow Effect 0 500 (NO: 750) / 1000 / 3000 ppm, 0 1/ 3 / 10 / 30 / 100 Flow rate in the 20 100 l/h range: Within detection limits The measurement ranges are freely adjustable within a range ratio of max. 1: 5. An individual measurement range can be Associated Gas Effect / Cross Sensitivity factory-set on request.

7 Measurement ranges should not be The knowledge of the sample gas composition is necessary set within ignition limits. for the analyzer configuration. Selectivity measures to reduce associated gas effect (optional): Incorporation of interference filters or filter cells, internal electronic cross-sensitivity correction for one sample component by other sample components measured with the gas analyzer . Temperature Effect Ambient temperature in permissible range At zero-point: 2 % of span per 10 C. On sensitivity without thermostat: 3 % of measured value per 10 C. On sensitivity with thermostat (optional): 2 % of measured value per 10 C. Thermostat temperature = 55 C.

8 Air Pressure Effect At zero-point: No effect On sensitivity with pressure correction by means of integral pressure sensor: % of measured value per 1 %. barometric pressure change The pressure sensor is located in the sample gas path if hoses are used as the internal gas lines. If tubing is used for internal gas lines the pressure sensor is routed to the outside via a hose. Pressure sensor working range: pabs = 600 1250 hPa Power Supply Effect Voltage and frequency in the permissible range: No effect 4 EasyLine EL3000 Series Data Sheet 10 EN November 2005. Infrared Photometer Uras26. Dynamic Response Gas Inlet Conditions Warm-Up Time The analyzer must not be used for measurement of ignitable Approx.

9 30 minutes without thermostat; approx. 2 hours with gas/air or gas/oxygen mixtures. thermostat Temperature 90% Response Time The sample gas dew point should be at least 5 C below the T90 = sec for measurement cell length = 175 mm, sample temperature throughout the sample gas path. Otherwise a gas flow = 60 l/h and electronic T90 time (static / dynamic) sample gas cooler or condensate trap is required. = 5 / 0 sec. Inlet Pressure pe = 2 500 hPa Lower pressures require a sample gas pump and higher Calibration pressures require a pressure reducer. Zero-Point Calibration Outlet Pressure With inert gas, N2, or with ambient air that is free of the Atmospheric pressure sample component.

10 Flow Rate End-Point Calibration 20 100 l/h With gas-filled calibration cells (optional) or with test gas mixtures. It is recommended to verify the calibration cell set Corrosive Gases values once a year. Highly corrosive associated gas components, chlorine (Cl2) and hydrogen chloride (HCl), as well as gases or aerosols containing chlorine must be cooled or undergo prior Materials in Contact with the Sample Medium absorption. analyzer (Sample Cells) Flammable Gases Tubing: Aluminum; Window: CaF2 or BaF2; Connectors: In the version with gas lines and connectors made of stainless Stainless steel steel the analyzer is suitable for measuring flammable gases in general purpose environment.