Example: air traffic controller

Micro Motion LF-Series Low Flow Sensor and …

Product Data SheetPS-00566, Rev. GApril 2013 Micro Motion LF-Series meters are the smallest Coriolis meters on the market, with a form factor the size of a handheld organizer. Highest precision from a miniaturized flow meter Patented optical sensing system delivers unparalleled sensitivity and stability Palm-sized form factor is ideal for tight spacesScalable platform that solves challenges in the most demanding low-flow applications Three sizes are available, with the smallest capable of measuring flow rates down to 1 g/h High-speed DSP for accuracy under the toughest conditionsSuperior reliability No moving parts to wear or replace minimizes maintenance for long-term reliability Rugged construction for durability in the lab or in the fieldMicro Motion LF

Micro Motion® LF-Series Low Flow Sensor and Transmitter 3 Liquid flow rates Gas flow rates Liquid and gas performance Mass Volume(1) (1) Volume measurement is based on a process-fluid density of 1 g/cm 3 (1000 kg/m3).For fluids with density other than 1 g/cm 3, the volumetric flow rate equals the maximum mass flow rate divided by the fluid’s …

Tags:

  Sensor

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of Micro Motion LF-Series Low Flow Sensor and …

1 Product Data SheetPS-00566, Rev. GApril 2013 Micro Motion LF-Series meters are the smallest Coriolis meters on the market, with a form factor the size of a handheld organizer. Highest precision from a miniaturized flow meter Patented optical sensing system delivers unparalleled sensitivity and stability Palm-sized form factor is ideal for tight spacesScalable platform that solves challenges in the most demanding low-flow applications Three sizes are available, with the smallest capable of measuring flow rates down to 1 g/h High-speed DSP for accuracy under the toughest conditionsSuperior reliability No moving parts to wear or replace minimizes maintenance for long-term reliability Rugged construction for durability in the lab or in the fieldMicro Motion LF-Series Low Flow Sensor and Transmitter High performance compact drainable Coriolis meterExtreme low-flow Coriolis meterHygienic compact drainable Coriolis meterPeak performance Coriolis meterGeneral purpose

2 Flow-only Coriolis meterF-SeriesLF-SeriesH-SeriesELITE R-SeriesStraight tubefull-bore Coriolis meterT-Series2 Micro Motion LF-Series Low Flow Sensor and TransmitterMicro Motion LF-Series meters are the smallest Coriolis meters available on the market. With a footprint the size of a handheld organizer, you can fit this meter into any tight space for your low-flow only is the footprint small so are the flow rates. With a range of to lb/hr ( to 27 kg/h), you can measure mass or volume flow for drops of liquid or other Micro Motion meters, the LF-Series offers all the benefits of Coriolis.

3 High accuracy, for a quicker ROI No moving parts, for reduced maintenance costs and downtime Direct measurement of mass, volume, temperature, and density for reduced instrumentation requirements Measurements unaffected by changing fluid properties, for no special installation requirements or manual calculationsApplicationsMicro Motion LF-Series meters are ideal for many low-flow applications, including flavoring, fragrances, and catalysts. LF-Series meters are designed to meet the demanding low-flow needs of a wide variety of industries such as chemical, petrochemical, life sciences, food and beverage, and oil and transmitters with MVD LF-Series transmitters are specially designed to work with LF-Series sensors on your low-flow applications.

4 LF-Series transmitters incorporate MVD technology an innovative, multivariable, digital signal processing capability. MVD technology: Improves ease of use with an easy-to-use display Reduces downtime with enhanced diagnostics Lowers your flowmetering costs with a 4-wire features include: Milliampere and frequency/pulse outputs Transmitter and process control functions in a single device Field-mount or DIN rail mount transmitter Optional NEMA enclosures (field-mount transmitters only) Optional filling and dosing application softwareMake your process more profitableMicro Motion LF-Series sensors and transmitters can be installed as part of a HART Bell 202 multidrop network, a Modbus RS-485 digital communications network, or a FOUNDATION fieldbus system.

5 All transmitters support Emerson Process Management's PlantWeb field-based architecture, which uses the power of intelligent, interoperable field devices to improve plant Motion is known worldwide for increasing plant efficiency, production, and profitability. More than 500,000 Micro Motion meters are installed and working in processes just like yours. Micro Motion LF-Series flowmetersMicro Motion LF-Series Low Flow Sensor and Transmitter3 Liquid flow rates Gas flow rates Liquid and gas performanceMassVolume(1)(1) Volume measurement is based on a process-fluid density of 1 g/cm3 (1000 kg/m3).

6 For fluids with density other than 1 g/cm3, the volumetric flow rate equals the maximum mass flow rate divided by the fluid s flow rates that produce approximately psid ( bar) pressure drop on air at 70 F ( C) and 500 psi (35 bar)MassVolumelb/hrkg/hSCFHSCCMT ypical flow ,350 Mass flow accuracy(1)(1) Stated flow accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on reference conditions of water at 70 F ( C).Standard of rate or % of rate, whichever is greaterOptional of rate or % of rate, whichever is greaterMass flow repeatability of rate or % of rate.

7 Whichever is greater zero stabilityflow rate--------------------------------- 100 zero stabilityflow rate--------------------------------- 100 12--- zero stabilityflow rate--------------------------------- 100 Sensor specifications4 Micro Motion LF-Series Low Flow Sensor and TransmitterLiquid and gas performance continuedHazardous area classifications Mass flow reproducibility of rate or % of rate, whichever is greaterZero stabilitylb/ (1)(1) Contact factory for applications with fluid density between and to g/cm3 (gas) and to g/cm3 (liquid)Accuracy(2) liquid only(2) At temperatures other than 70 F ( C), you can expect an additional density error of approximately g/cm3 per C.

8 G/cm3 Repeatability liquid only g/cm3 TemperatureAmbient and process0 to 65 C (32 to 149 F)Accuracy C ( F)Maximum operating pressureStandard35 bar (500 psi)Optional100 bar (1500 psi)CSAC lass I Division 2 Groups A, B, C, DClass II Division 2 Groups F and GClass III Division 2AT E XII 3 G EEx nA II T4II 3 D IP65 T135 C12--- zero stabilityflow rate--------------------------------- 100 Sensor specifications continuedMicro Motion LF-Series Low Flow Sensor and Transmitter5 Materials of construction Physical specificationsDimensionsWetted parts316/316L stainless steelOptional filter components (wetted)302 and 316 stainless steelProcess seals Viton fluoroelastomer(1)

9 , Buna, Kalrez, or EPDM(1) Viton is a registered trademark of DuPont Performance aluminumWeight4 lbs (2 kg)Shipping weight5 lbs (2 kg)ModelFitting optionsFitting codeDim. A with filterinches (mm)Dim. A without filterinches (mm)Dim. B inches (mm)All 1/4 tube compression fittings3246 1/4 (158)5 (126)2 1/2 (63)1/8 tube compression fittings3285 15/16 (151)4 11/16 (119)2 5/16 (59)1/8 VCR3275 11/16 (144)4 3/8 (112)2 3/16 (56)1/4 VCR3226 3/8 (162)5 1/8 (129)2 9/16 (65)1/4 VCO3235 7/8 (149)4 5/8 (117)2 5/16 (59)6 mm tube compression fittings3256 1/4 (158)5 (126)2 1/2 (63) Sensor specifications continued6 Micro Motion LF-Series Low Flow Sensor and TransmitterDimensions continuedDimensions ininches(mm)7/16(12)6 5/16(161)

10 1/4(6)3 1/16(77)Dim. B4 7/8(124)5 1/2(140)3/4(19)2 X 1/4(7)2 7/16(62)2 9/16(65)5 1/8(130)1(25)1 1/2(38)Dim. A 1/8( 3)CLCLM ounting holesSensor specifications continuedMicro Motion LF-Series Low Flow Sensor and Transmitter7 Transmitter outputsTransmitters with configurable I/OThe field-mount and DIN rail transmitters (mounting and outputs codes 4 and 5, with configurable I/O), are designed to increase transmitter flexibility and reduce the number of transmitter variations required in inventory. The table below shows the various configuration options that can be produced with the configurable I/O output option.