Transcription of NI Switches Help About These Specifications
1 NI PXI/PXIe-2575 Specifications196 1 Relay MultiplexerThis document lists Specifications for the NI PXI/PXIe-2575 (NI 2575) 196 1 multiplexer relay module. All Specifications are subject to change without notice. Visit for the most current 1-wire 196 1 multiplexer 2-wire 95 1 multiplexer 2-wire 98 1 multiplexerRefer to the NI Switches Help for detailed topology These SpecificationsSpecifications characterize the warranted performance of the instrument under the stated operating conditions. Typical Specifications are Specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 C. Typical Specifications are not warranted. All voltages are specified in DC, ACpk, or a combination unless otherwise specified. Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a soft, lint-free, dampened cloth. Do not use detergent or chemical solvents.
2 The unit must be completely dry and free from contaminants before returning to to the Read Me First: Safety and Electromagnetic Compatibility document for important safety and electromagnetic compatibility information. To obtain a copy of this document online, visit , and search for the document ensure the specified EMC performance, operate this product only with shielded cables and protection provided by the NI 2575 can be impaired if it is used in a manner not described in this document. 2| |NI PXI/PXIe-2575 SpecificationsInput CharacteristicsMaximum switching voltageChannel-to-channel .. 100 VChannel-to-ground .. 100 V, CAT ICautionThis module is rated for Measurement Category I and is intended to carry signal voltages no greater than 100 V. This module can withstand up to 500 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III, or IV. Do not connect to MAINs supply circuits (for example, wall outlets) of 115 or 230 VAC.
3 Refer to the Read Me First: Safety and Electromagnetic Compatibility document for more information About measurement hazardous voltages (> Vpk/60 VDC) are present on any relay terminal, safety low-voltage ( Vpk/60 VDC) cannot be connected to any other maximum switching power is limited by the maximum switching current and the maximum voltage, and must not exceed 60 W, switching power(per channel) .. 60 W, VA (DC to 60 Hz)DC isolation resistance .. >1 G , typical at 25 C(between channel and COM terminals)Maximum total current .. 1 A(switching or carry)Minimum switch load .. 20 mV/1 mANoteThe NI 2575 is not recommended for 2-wire resistance inductive loads (for example, motors and solenoids) can produce high voltage transients in excess of the module's rated voltage. Without additional protection, These transients can interfere with module operation and impact relay life. For more information About transient suppression, visit and enter the Info Code path resistanceInitial.
4 < 1 DC path resistance typically remains low for the life of the relay. At the end of relay life, the path resistance rapidly rises above 1 . Load ratings apply to relays used within the specification before the end of relay PXI/PXIe-2575 Specifications | National Instruments|3 Differential thermal 3 <12 VBandwidth (-3 dB, 50 termination) >20 >8 MHzChannel-to-channel isolation (50 termination)Each relay in the NI 2575 is shared by two channels. Refer to Table 4 for a list of channel channels in different relays10 >90 dB100 >70 dB1 MHz .. >50 dB1-wire channels in the same relay10 >68 dB100 >48 dB1 MHz .. >28 dB2-wire channels10 >95 dB100 >75 dB1 MHz .. >55 dBOpen channel isolation (50 termination)10 >90 dB100 >70 dB1 MHz .. > 50 dBDynamic CharacteristicsRelay operate timeTypical .. 1 msNoteCertain applications may require additional time for proper settling. Refer to the NI Switches Help for information About including additional settling To ensure the typical thermal EMF, power down all relays and avoid pulsing high currents near the channels you are measuring.
5 For more information About powering down latching relays, refer to the Power Down Latching Relays After Debounce property in NI-SWITCH or the Power Down Latching Relays After Settling property in | |NI PXI/PXIe-2575 SpecificationsExpected relay lifeMechanical .. 1 108 cyclesElectrical10 VDC, 100 mADC 106 cycles10 VDC, 1 ADC resistive .. 1 106 cycles30 VDC, 1 ADC resistive .. 5 105 cycles60 VDC, 1 ADC resistive .. 1 105 cyclesNoteThe relays used in the NI 2575 are field replaceable. Refer to the NI Switches Help for information About replacing a failed CharacteristicsInput PXI trigger lines 0 to 7 Minimum pulse width .. 150 nsNoteThe NI 2575 can recognize trigger pulse widths less than 150 ns by disabling digital filtering. For information About disabling digital filtering, refer to the NI Switches triggerDestinations .. PXI trigger lines 0 to 7 Pulse Programmable (1 sto62 s)Physical CharacteristicsRelay type .. Electromechanical, latchingRelay contact material.
6 Silver, gold coveredI/O connector .. 200 POS LFH Matrix 50, receptaclePower requirementPXI .. 6 W at 5 V, W at VPXI Express .. W at 12V, W at VDimensions (L W H) .. 3U, one slot, PXI/cPCI module, PXIe compatible cm ( )Weight .. 231 g ( oz)NI PXI/PXIe-2575 Specifications | National Instruments|5 EnvironmentOperating temperature .. 0 C to 55 CStorage temperature .. -20 C to 70 CRelative humidity .. 5% to 85%, noncondensingPollution Degree .. 2 Maximum 2,000 mIndoor use and VibrationOperational Shock .. 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)Random VibrationOperating .. 5 to 500 Hz, grmsNonoperating .. 5 to 500 Hz, grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)6| |NI PXI/PXIe-2575 SpecificationsDiagramsFigure 1 shows the NI 2575 hardware 1.
7 NI 2575 Hardware Diagram COM+COM CH0CH95CH93CH188CH94CH189CH190CH193CH191 CH194CH192CH195CH1CH96NO CONNECTNI PXI/PXIe-2575 Specifications | National Instruments|7 Figure 2 shows the NI 2575 connector pinout for the 1-wire 196 1 2. NI 2575 Connector Pinout for the 1-wire 196 1 topologyNO CONNECTCH166CH71CH165CH70CH164CH69CH163C H68CH162CH67CH161CH66CH160CH65CH159CH64C H158CH63CH157CH62CH156CH61CH155CH60CH193 CH154CH59CH153CH58CH152CH57CH151CH56CH15 0CH55CH149CH54CH148CH53CH147CH52CH146CH5 1CH145CH50CH144CH49CH143CH48150151149152 1481531471541461551451561441571431581421 5914116014016113916213816313716413616513 5166134167133168132169131170130171129172 1281731271741261751251761241771231781221 7912118012018111918211818311718411618511 5186114187113188112189111190110191109192 1081931071941061951051961041971031981021 9920010151505249534854475546564557445843 5942604161406239633864376536663567346833 6932703171307229732874277526762577247823 7922802181208219831884178516861587148813 8912901191109299389479569659749839921001 CH84CH179CH85CH180CH86CH181CH87CH182CH88 CH183CH89CH184CH90CH185CH91CH186CH92CH18 7CH93CH188CH94CH189 COMNO
8 CONNECTCH190CH72CH167CH73CH168CH74CH169C H75CH170CH76CH171CH77CH172CH78CH173CH79C H174CH80CH175CH81CH176CH82CH177CH83CH178 CH192CH36CH131CH37CH132CH38CH133CH39CH13 4CH40CH135CH41CH136CH42CH137CH43CH138CH4 4CH139CH45CH140CH46CH141CH47CH142CH191CH 24CH119CH25CH120CH26CH121CH27CH122CH28CH 123CH29CH124CH30CH125CH31CH126CH32CH127C H33CH128CH34CH129CH35CH130CH195NO CONNECTCH118CH23CH117CH22CH116CH21CH115C H20CH114CH19CH113CH18CH112CH17CH111CH16C H110CH15CH109CH14CH108CH13CH107CH12CH194 CH106CH11CH105CH10CH104CH9CH103CH8CH102C H7CH101CH6CH100CH5CH99CH4CH98CH3CH97CH2C H96CH1CH95CH08| |NI PXI/PXIe-2575 SpecificationsFigure 3 shows the NI 2575 connector pinout for the 2-wire 95 1 3. NI 2575 Connector Pinout for the 2-wire 95 1 topologyNO CONNECTCH71 CH71+CH70 CH70+CH69 CH69+CH68 CH68+CH67 CH67+CH66 CH66+CH65 CH65+CH64 CH64+CH63 CH63+CH62 CH62+CH61 CH61+CH60 CH60+NO CONNECTCH59 CH59+CH58 CH58+CH57 CH57+CH56 CH56+CH55 CH55+CH54 CH54+CH53 CH53+CH52 CH52+CH51 CH51+CH50 CH50+CH49 CH49+CH48 CH48+15015114915214815314715414615514515 6144157143158142159141160140161139162138 1631371641361651351661341671331681321691 3117013017112917212817312717412617512517 6124177123178122179121180120181119182118 1831171841161851151861141871131881121891 1119011019110919210819310719410619510519 6104197103198102199200101515052495348544 7554656455744584359426041614062396338643 7653666356734683369327031713072297328742 7752676257724782379228021812082198318841 7851686158714881389129011911092993894795 69659749839921001CH84+CH84 CH85+CH85 CH86+CH86 CH87+CH87 CH88+CH88 CH89+CH89 CH90+CH90 CH91+CH91 CH92+CH92 CH93+CH93 CH94+CH94 COM+COM NO CONNECTCH72+CH72 CH73+CH73
9 CH74+CH74 CH75+CH75 CH76+CH76 CH77+CH77 CH78+CH78 CH79+CH79 CH80+CH80 CH81+CH81 CH82+CH82 CH83+CH83 NO CONNECTCH36+CH36 CH37+CH37 CH38+CH38 CH39+CH39 CH40+CH40 CH41+CH41 CH42+CH42 CH43+CH43 CH44+CH44 CH45+CH45 CH46+CH46 CH47+CH47 NO CONNECTCH24+CH24 CH25+CH25 CH26+CH26 CH27+CH27 CH28+CH28 CH29+CH29 CH30+CH30 CH31+CH31 CH32+CH32 CH33+CH33 CH34+CH34 CH35+CH35 NO CONNECTNO CONNECTCH23 CH23+CH22 CH22+CH21 CH21+CH20 CH20+CH19 CH19+CH18 CH18+CH17 CH17+CH16 CH16+CH15 CH15+CH14 CH14+CH13 CH13+CH12 CH12+NO CONNECTCH11 CH11+CH10 CH10+CH9 CH9+CH8 CH8+CH7 CH7+CH6 CH6+CH5 CH5+CH4 CH4+CH3 CH3+CH2 CH2+CH1 CH1+CH0 CH0+NI PXI/PXIe-2575 Specifications | National Instruments|9 Figure 4 shows the NI 2575 connector pinout for the 2-wire 98 1 4. NI 2575 Connector Pinout for the 2-wire 98 1 topologyNO CONNECTCH71 CH71+CH70 CH70+CH69 CH69+CH68 CH68+CH67 CH67+CH66 CH66+CH65 CH65+CH64 CH64+CH63 CH63+CH62 CH62+CH61 CH61+CH60 CH60+CH95 CH59 CH59+CH58 CH58+CH57 CH57+CH56 CH56+CH55 CH55+CH54 CH54+CH53 CH53+CH52 CH52+CH51 CH51+CH50 CH50+CH49 CH49+CH48 CH48+15015114915214815314715414615514515 6144157143158142159141160140161139162138 1631371641361651351661341671331681321691 3117013017112917212817312717412617512517 6124177123178122179121180120181119182118 1831171841161851151861141871131881121891 1119011019110919210819310719410619510519 6104197103198102199200101515052495348544 7554656455744584359426041614062396338643 7653666356734683369327031713072297328742 7752676257724782379228021812082198318841 7851686158714881389129011911092993894795 69659749839921001CH84+CH84 CH85+CH85 CH86+CH86 CH87+CH87 CH88+CH88 CH89+CH89 CH90+CH90 CH91+CH91 CH92+CH92 CH93+CH93 CH94+CH94 COM+COM CH95+CH72+CH72 CH73+CH73 CH74+CH74
10 CH75+CH75 CH76+CH76 CH77+CH77 CH78+CH78 CH79+CH79 CH80+CH80 CH81+CH81 CH82+CH82 CH83+CH83 CH97+CH36+CH36 CH37+CH37 CH38+CH38 CH39+CH39 CH40+CH40 CH41+CH41 CH42+CH42 CH43+CH43 CH44+CH44 CH45+CH45 CH46+CH46 CH47+CH47 CH96+CH24+CH24 CH25+CH25 CH26+CH26 CH27+CH27 CH28+CH28 CH29+CH29 CH30+CH30 CH31+CH31 CH32+CH32 CH33+CH33 CH34+CH34 CH35+CH35 CH97 NO CONNECTCH23 CH23+CH22 CH22+CH21 CH21+CH20 CH20+CH19 CH19+CH18 CH18+CH17 CH17+CH16 CH16+CH15 CH15+CH14 CH14+CH13 CH13+CH12 CH12+CH96 CH11 CH11+CH10 CH10+CH9 CH9+CH8 CH8+CH7 CH7+CH6 CH6+CH5 CH5+CH4 CH4+CH3 CH3+CH2 CH2+CH1 CH1+CH0 CH0+10| |NI PXI/PXIe-2575 SpecificationsNoteFor topology-specific connection information, refer to your device in the NI Switches for more information About the following using either of the LFH200 cables in Table 1 with the NI 2575 in the 2-wire 98 1 topology, CH95, CH96, and CH97 will have lower RF performance than the other 95 channels because they are not in twisted pairs in the cable.