Transcription of SPECIFICATIONS NI PXIe-4322
1 SPECIFICATIONSNI PXIe-4322 This document lists SPECIFICATIONS for the NI PXIe-4322 module. These SPECIFICATIONS are typical for the range of 0 C to 55 C unless otherwise stated. The system must be allowed to warm up for 15 minutes to achieve the rated accuracy. All SPECIFICATIONS are subject to change without notice. Visit for the most current SPECIFICATIONS and product the filler panels on all unused slots in your chassis to maintain forced air CharacteristicsNumber of channels .. 8 analog output channelsDAC resolution .. 16 bitsType of R-2 RMonotonicity .. 16 bitsDNL .. 1 LSB maxINL (best fit) .. 4 LSBs maxPower-on output state1 Voltage mode .. 0 VCurrent mode .. 0 mANoteYou can program the power-on output states. Refer to your software documentation for information about programming the power-on output states using NI-DAQmx with LabVIEW or other National Instruments application development environments (ADEs).
2 Power-off output High impedance1 When the output stage powers on, a current glitch occurs for 1 ms peaking at 5 A. When the output stage powers off, a current glitch occurs for 3 ms peaking at 5 | |NI PXIe-4322 120 VDCS hort IndefinitelyUpdate rateMaximum .. 250 kS/s per channelMinimum .. No minimumTiming 50 ppm of sample rateTiming resolution .. 10 nsData transfers .. DMA (scatter-gather), programmed I/OOutput FIFO size .. 8,191 samples shared among channels usedAO waveform modes Nonperiodic waveform Periodic waveform regeneration mode from onboard FIFO Periodic waveform regeneration from host buffer including dynamic updateVoltage ModeOutput voltage range1 Nominal .. 16 VMinimum .. VMaximum .. VCurrent drive .. 20 mA per channel maxOutput impedance .. 25 m Noise (rms)10 Hz to 1 kHz bandwidth .. 30 Vrms10 Hz to 300 kHz bandwidth.
3 250 Vrms10 Hz to 20 MHz bandwidth .. 500 VrmsSlew rate .. 10 V/ sCrosstalkChannel-to-channel @ 10 -100 dBCommon-mode voltage @ 60 Hz .. -120 dB1 Refer to the Increasing Output Voltage Range in Voltage Mode section in the NI PXIe-4322 User Manual for information about how to increase the nominal output voltage range by connecting multiple voltage channels in PXIe-4322 SPECIFICATIONS | National Instruments|3 Settling time1000 pF load, to 1 LSB20 V step .. 20 s1 V step .. 12 V step .. 10 s500 ||100 pF, to 1 LSB20 V step .. 20 sCapacitive drive .. 4500 pFAccuracy1 StabilityGain 7 ppm/ COffset drift .. 25 V/ CAbsolute Voltage Output Accuracy EquationAbsoluteVoltageAccuracy = Output * (GainError) + Range * (OffsetError)Absolute Voltage Output Accuracy ExampleFor a 10 V voltage output, the absolute output accuracy for an external temperature range of 18 C to 28 C is as follows:GainError = = = 10 V * (GainError) + 16 V * (OffsetError) = mV1 Accuracies listed are warranted for the conditions described in the table for up to one year from the module external ConditionsPercent of Output(Gain Error)Percent of Range*(Offset Error)Calibrated, max (0 C to 55 C) , max (0 C to 40 C) , max (23 C 5 C) , typ (23 C 5 C) * Range equals 16 | |NI PXIe-4322 SpecificationsCurrent ModeOutput current range1 Nominal.
4 20 mAMinimum .. mAMaximum .. mACompliance voltage .. 16 V per channel maxOutput impedance .. 100 M Noise (rms)10 Hz to 1 kHz bandwidth .. 50 nA10 Hz to 300 kHz bandwidth .. 600 nASlew rate .. 20 mA/ sCrosstalkChannel-to-channel @ 1 -100 dBCommon-mode voltage @ 60 Hz .. 50 nA/VSettling time100 loadFull-scale step to 2 LSB .. 20 s800 loadFull-scale step to 2 LSB .. 25 s2 mA step to 1 15 sInductive drive .. 10 H1 Refer to the Increasing Output Current Range in Current Mode section of the NI PXIe-4322 User Manual for information about how to increase the nominal output current range by connecting current channels in PXIe-4322 SPECIFICATIONS | National Instruments|5 Accuracy1 StabilityGain 15 ppm/ COffset drift .. 75 nA/ CAbsolute Current Output Accuracy EquationAbsoluteCurrentAccuracy = Output * (GainError) + Range * (OffsetError)Absolute Current Output Accuracy ExampleFor a 10 mA current output, the absolute output accuracy for an external temperature range of 18 C to 28 C is as follows:GainError = = = 10 mA * (GainError) + 20 mA* (OffsetError) = ASynchronizationReference clock source.
5 PXIe_DSTAR< B>, PXI_STAR, PXIe_Clk100, PXI_TRIG< >Digital TriggersSource .. PXI_TRIG< >, PXI_STAR, PXIe_DSTAR< > PFI< >Purpose .. Start Trigger,Pause Trigger1 Accuracies listed are warranted for the conditions described in the table for up to one year from the module external ConditionsPercent of Output(Gain Error)Percent of Range*(Offset Error)Calibrated, max (0 C to 55 C) , max (0 C to 40 C) , max (23 C 5 C) , typ (23 C 5 C) *Range equals 20 | |NI PXIe-4322 SpecificationsPolarity .. Software-selectableDebounce filter settings .. Disable, 90 ns, s, ms,custom intervalClockingSource .. Onboard Clock, PXI_TRIG< >, PXI_STAR, PXIe_DSTAR< >, PXIe_Clk100 (RefClk only)Destination .. Sample Clock,Sample Clock Timebase,Reference ClockPolarity .. Software-selectable (except Reference Clock)Debounce filter settings(Sample Clock only).
6 Disable, 90 ns, s, ms,custom intervalReference clock locking frequenciesNoteNational Instruments does not recommend locking to non-selected Timing SignalsSource .. Start Trigger,Pause Trigger,Sample Clock,various derived timebases and clocksDestination .. PXI_TRIG< > PXIe_DSTARCP olarity .. Software-selectableReference SignalLocking Input Frequency (MHz)1020100 PXIe_DSTAR< > PXI_STAR PXIe_Clk100 PXI_TRIG< > NI PXIe-4322 SPECIFICATIONS | National Instruments|7 Bus InterfaceForm factor .. x1 PXI Express peripheral module, specification rev compliantSlot compatibility .. x1 and x4 PXI Express or PXI Express hybrid slotsDMA channels .. 1 analog outputNI PXIe-4322 modules may be installed in PXI Express slots or PXI Express hybrid warm-up 15 minutesCalibration interval .. 1 yearPower Requirements+ V .. 800 mA+12 V .. 700 mAPhysical RequirementsDimensions.
7 Standard 3U PXIe, 16 10 cm ( in.) 161 g ( oz)I/O connector .. 96-pin male DIN 41612/IEC 60603-2 connectorCautionClean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to SpecificationsMaximum 2,000 m (800 mbar), at 25 C ambient temperaturePollution Degree .. 2 Indoor use only8| |NI PXIe-4322 SpecificationsOperating EnvironmentAmbient temperature range .. 0 C to 55 C(Tested in accordance with IEC-60068-2-1and IEC-60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.)Relative humidity range .. 10% to 90%, noncondensing(Tested in accordance with IEC-60068-2-56.)Storage EnvironmentAmbient temperature range .. -40 C to 71 C(Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)
8 Relative humidity range .. 5% to 95% noncondensing(Tested in accordance with IEC-60068-2-56.)Shock and VibrationOperating shock .. 30 g peak, half-sine, 11 ms pulse(Tested in accordance with IEC-60068-2-27. Meets MIL-PRF-28800F Class 2 limits.)Random vibrationOperating .. 5 Hz to 500 Hz, grmsNon-operating .. 5 Hz to 500 Hz, grms(Tested in accordance with IEC-60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)Safety VoltageConnect only voltages that are within the following limits:Between any two terminalsof an isolated channel .. 120 VDCI solationChannel to channelContinuous .. 300 Vrms, Measurement Category II (Basic)Withstand .. 1,500 Vrms,verified by 5 s dielectric withstand testNI PXIe-4322 SPECIFICATIONS | National Instruments|9 Channel to earth 300 Vrms, Measurement Category IIWithstand.
9 3,000 Vrms,verified by a 5 s dielectric withstand testMeasurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for or 230 V for not use for measurements within Measurement Categories III or protection provided by the NI PXIe-4322 can be impaired if it is used in a manner not described in this When hazardous voltages (>30 Vpk/60 VDC) are present on any terminal, safety low-voltage ( 30 Vpk/60 VDC) cannot be connected to any other not supply hazardous voltages (>30 Vpk/60 VDC) to the terminal block without the terminal block being connected to the NI product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use.
10 IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 NoteFor UL and other safety certifications, refer to the product label or the Online Product Certification CompatibilityThis product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions EN 55022 (CISPR 22): Class A emissions EN 55024 (CISPR 24): Immunity AS/NZS CISPR 11: Group 1, Class A emissions AS/NZS CISPR 22: Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions10| |NI PXIe-4322 SpecificationsNoteIn the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis EMC declarations and certifications, and additional information, refer to the Online Product Certification ComplianceThis product meets the essential requirements of applicable European Directives as follows: 2006/95/EC; Low-Voltage Directive (safety) 2004/108/EC; Electromagnetic Compatibility Directive (EMC)Online Product CertificationRefer to the product Declaration of Conformity (DoC) for additional regulatory compliance information.