Target-Pro 2 Analyzer and SamplerOperation ManualOperating PrecautionsBattery The Target-Pro 2 uses a nickel-cadmium (NiCd) it for at least 24 hours before first use. For optimum performance, fully discharge and rechargethe battery once a month: Leave the Target-Pro 2switched on until the LCD screen goes blank, then switchoff the Target-Pro 2, connect the charging adapter, andallow it to charge the battery for 24 hours. To prolong battery life, switch off the Analyzer when it isnot in CleaningDo not clean the Target-Pro 2 or Bottle Sampler with ace-tone or other solvents that are not compatible with nitrileseals. The recommended cleaning fluid for internal flushingis mineral Vickers Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E July 2004 Target-Pro 2 Laser Particle AnalyzerUser Guide: September 2004 ContentsOperating Precautions2 Introduction3 ISO Cleanliness Code3 Target-Pro 2 Operation3 Key and Connector Diagrams3 Online Operation3 Analyzer Set-up4 Test Set-up4 Running a Test5 Shutting Down5 Continuous Sampling6 7 External Wiring Details8 Memory Recall8 Battery Charging / Print
6 EATON Vickers Target-Pro 2 Analyzer and Sampler Operation Manual V-PP-MC-0002-E July 2004 Continuous Sampling—Set Intervals The …