Transcription of Keysight InfiniiVision 3000 X-Series Oscilloscopes …
1 Keysight InfiniiVision 3000 X-Series OscilloscopesUser's Guide2 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's GuideNotices Keysight Technologies, Inc. 2005-2013No part of this manual may be reproduced in any form or by any means (includ ing elec-tronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technolo-gies, Inc. as governed by United States and international copyright Part Number75019-97073Ed itionEighth ed ition, April 2013 Printed in MalaysiaPublished by: Keysight Technologies, Garden of the Gods Road Colorado Springs, CO 80907 USAW arrantyThe material contained in this docu-ment is provided as is, and is subject to being changed, without notice, in future ed itions. Further, to the maxi-mum extent permitted by appl icable law, Keysight disclaims all warranties, either express or impl ied, with regard to this manual and any information con-tained herein, incl ud ing but not l imited to the impl ied warranties of merchant-abil ity and fitness for a particular pur-pose.
2 Keysight shall not be l iable for errors or for incidental or consequential d amages in connection with the fur-nishing, use, or performance of this document or of any information con-tained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the mate-rial in this document that confl ict with these terms, the warranty terms in the separate agreement shall Licenses The hard ware and/or software described in this document are furnished under a license and may be used or copied only in accor-dance with the terms of such Rights LegendIf software is for use in the performance of a Government prime contract or subcon-tract, Software is delivered and licensed as Commercial computer software as defined in DFAR (June 1995), or as a commercial item as defined in FAR (a) or as Restricted computer software as defined in FAR (June 1987) or any equivalent agency regulation or contract clause.
3 Use, d uplication or d isclosure of Software is subject to Keysight Technologies standard commercial license terms, and non-DOD Departments and Agencies of the Government will receive no greater than Restricted Rights as defined in FAR (c)(1-2) (June 1987). Govern-ment users will receive no greater than Lim-ited Rights as defined in FAR (June 1987) or DFAR (b)(2) (November 1995), as applicable in any tech-nical NoticesCAUTIONA CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could resul t in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the ind icated cond itions are fully understood and met. WARNINGA WARNING notice denotes a haz-ard. It calls attention to an operat-ing proced ure, practice, or the l ike that, if not correctly performed or adhered to, could resul t in personal injury or death.
4 Do not proceed beyond a WARNING notice until the ind icated cond itions are fully understood and met. Print History75019-97000, January 201175019-97013, February 201175019-97014, June 201175019-97027, October 201175019-97040, February 201275019-97051, March 201275019-97062, July 201275019-97073, April 2013 TrademarksJava is a trademark of Sun Microsys-tems, , Sun Microsystems, and the Sun Logo are trademarks or registered trademarks of Sun Microsystems, Inc. in the and other InfiniiVision 3000 X-Series Oscilloscopes user 's Guide3 InfiniiVision 3000 X-Series Oscilloscopes At a GlanceThe Keysight InfiniiVision 3000 X-Series Oscilloscopes deliver these features: 100 MHz, 200 MHz, 350 MHz, 500 MHz, and 1 GHz bandwidth models. 2- and 4-channel digital storage oscilloscope (DSO) le 13000 X-Series Model Numbers, Band wid ths, Sample RatesBand wid th100 MHz200 MHz350 MHz500 MHz1 GHzSample Rate (interleaved, non-interleaved)4 GSa/s, 2 GSa/s4 GSa/s, 2 GSa/s4 GSa/s, 2 GSa/s4 GSa/s, 2 GSa/s5 GSa/s, GSa/s2-Channel + 16 Logic Channels MSOMSO-X 3012 AMSO-X 3032 AMSO-X 3052 AMSO-X 3102A4-Channel + 16 Logic Channels MSOMSO-X 3014 AMSO-X 3024 AMSO-X 3034 AMSO-X 3054 AMSO-X 3104A2-Channel DSODSO-X 3012 ADSO-X 3032 ADSO-X 3052 ADSO-X 3102A4-Channel DSODSO-X 3014 ADSO-X 3024 ADSO-X 3034 ADSO-X 3054 ADSO-X 3104A4 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide 2+16-channel and 4+16-channel mixed-signal oscilloscope (MSO) MSO lets you debug your mixed-signal designs using analog signals and tightly correlated digital signals simultaneously.
5 The 16 digital channels have a 1 GSa/s sample rate ( GSa/s for the 1 GHz models), with a 50 MHz toggle rate. inch WVGA display. Interleaved 2 Mpts or non-interleaved 1 Mpts MegaZoom IV memory for the fastest waveform update rates, uncompromised. Upgradeable to 4 Mpts/2 Mpts. All knobs are pushable for making quick selections. Trigger types: edge, edge then edge, pulse width, pattern, OR, rise/fall time, Nth edge burst, runt, setup & hold, video, and USB. Serial decode/trigger options for: CAN/LIN, FlexRay, I2C/SPI, I2S, UART/RS232, and MIL-STD-1553/ARINC 429. Lister for serial decode Math waveforms: add, subtract, multiply, FFT, d/dt, integrate, and square root. With the DSOX3 ADVMATH option, you get these additional math waveforms: Ax+B, square, absolute value, common logarithm, natural logarithm, exponential, base 10 exponential, low pass filter, high pass filter, magnify, measurement trend, chart logic bus timing, and chart logic bus state.
6 Reference waveform locations (2) for comparing with other channel or math waveforms. Many built-in measurements and a measurement statistics display. Built-in license-enabled waveform generator with: arbitrary, sine, square, ramp, pulse, DC, noise, sine cardinal, exponential rise, exponential fall, cardiac, and gaussian pulse. USB ports make printing, saving and sharing data easy. Optional LAN/VGA module for connecting to a network and displaying the screen on a different monitor. Optional GPIB module. A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help. Complete instructions for using the quick help system are given in Access the Built-In Quick Help"on page more information about InfiniiVision Oscilloscopes , see: " " Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide5 In This GuideThis guide shows how to use the InfiniiVision 3000 X-Series unpacking and using the oscilloscope for the first time, see: Chapter 1, Getting Started, starting on page 23 When d isplaying waveforms and acquired data, see.
7 Chapter 2, Horizontal Controls, starting on page 47 Chapter 3, Vertical Controls, starting on page 61 Chapter 4, Math Waveforms, starting on page 71 Chapter 5, Reference Waveforms, starting on page 99 Chapter 6, Digital Channels, starting on page 103 Chapter 7, Serial Decode, starting on page 121 Chapter 8, Display Settings, starting on page 127 Chapter 9, Labels, starting on page 133 When setting up triggers or changing how data is acquired, see: Chapter 10, Triggers, starting on page 139 Chapter 11, Trigger Mode/Coupling, starting on page 173 Chapter 12, Acquisition Control, starting on page 181 Making measurements and analyzing data: Chapter 13, Cursors, starting on page 199 Chapter 14, Measurements, starting on page 209 Chapter 15, Mask Testing, starting on page 237 Chapter 16, Digital Vol tmeter, starting on page 249 When using the buil t-in license enabled waveform generator, see: Chapter 17, Waveform Generator, starting on page 253 When saving, recalling, or printing, see: Chapter 18, Save/Recall (Setups, Screens, Data), starting on page 269 Chapter 19, Print (Screens), starting on page 281 When using the oscilloscope's utility functions or web interface, see: Chapter 20, Utility Settings, starting on page 287 Chapter 21, Web Interface, starting on page 3076 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide For reference information, see: Chapter 22, Reference, starting on page 323 When using licensed serial bus triggering and decode features, see.
8 Chapter 23, CAN/LIN Triggering and Serial Decode, starting on page 343 Chapter 24, FlexRay Triggering and Serial Decode, starting on page 359 Chapter 25, I2C/SPI Triggering and Serial Decode, starting on page 369 Chapter 26, I2S Triggering and Serial Decode, starting on page 387 Chapter 27, MIL-STD-1553/ARINC 429 Triggering and Serial Decode, starting on page 397 Chapter 28, UART/RS232 Triggering and Serial Decode, starting on page 413 NOTEA bbreviated instructions for pressing a series of keys and softkeysInstructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:Press [Key1]> Softkey2 > keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located d irectly below the oscilloscope d InfiniiVision 3000 X-Series Oscilloscopes user 's Guide7 ContentsInfiniiVision 3000 X-Series Oscilloscopes At a Glance / 3In This Guide / 51 Getting StartedInspect the Package Contents / 23 Install the Optional LAN/VGA or GPIB Module / 26 Til t the Oscilloscope for Easy Viewing / 26 Power-On the Oscilloscope / 27 Connect Probes to the Oscilloscope / 28 Maximum input voltage at analog inputs / 28 Do not float the oscilloscope chassis / 29 Input a Waveform / 29 Recall the Defaul t Oscilloscope Setup / 29 Use Auto Scale / 30 Compensate Passive Probes / 32 Learn the Front Panel Controls and Connectors / 33 Front Panel Overlays for Different Languages / 40 Learn the Rear Panel Connectors / 41 Learn the Oscilloscope Display / 43 Access the
9 Buil t-In Quick Help / 448 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide2 Horizontal ControlsTo ad just the horizontal (time/d iv) scale / 48To ad just the horizontal delay (position) / 49 Panning and Zooming Single or Stopped Acquisitions / 50To change the horizontal time mode (Normal, XY, or Roll) / 50XY Time Mode / 51To d isplay the zoomed time base / 54To change the horizontal scale knob's coarse/fine ad justment setting / 55To position the time reference (left, center, right) / 56 Searching for Events / 56To set up searches / 57To copy search setups / 57 Navigating the Time Base / 58To navigate time / 58To navigate search events / 58To navigate segments / 593 Vertical ControlsTo turn waveforms on or off (channel or math) / 62To ad just the vertical scale / 63To ad just the vertical position / 63To specify channel coupling / 63To specify channel input impedance / 64To specify band wid th limiting / 65To change the vertical scale knob's coarse/fine ad justment setting / 65To invert a waveform / 65 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide9 Setting Analog Channel Probe Options / 66To specify the channel units / 67To specify the probe attenuation / 67To specify the probe skew / 68To calibrate a probe / 684 Math WaveformsTo d isplay math waveforms / 71To perform transforms or fil ters on an arithmetic operation / 72To ad just the math waveform scale and offset / 73 Units for Math Waveforms / 73 Math Operators / 74 Add or Subtract / 74 Mul tiply or Divide / 75 Math Transforms / 76
10 Differentiate / 77 Integrate / 78 FFT Measurement / 81 Square Root / 87Ax + B / 88 Square / 89 Absolute Value / 90 Common Logarithm / 90 Natural Logarithm / 91 Exponential / 91 Base 10 Exponential / 91 Math Fil ters / 92 High Pass and Low Pass Fil ter / 92 Math Visualizations / 93 Magnify / 9410 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's GuideMeasurement Trend / 94 Chart Logic Bus Timing / 96 Chart Logic Bus State / 975 Reference WaveformsTo save a waveform to a reference waveform location / 99To d isplay a reference waveform / 100To scale and position reference waveforms / 101To ad just reference waveform skew / 101To d isplay reference waveform information / 102To save/recall reference waveform files to/from a USB storage device / 1026 Digital ChannelsTo connect the d igital probes to the device under test / 103 Probe cable for digital channels / 104 Acquiring waveforms using the d igital channels / 107To d isplay d igital channels using AutoScale / 107 Interpreting the d igital waveform d isplay / 108To change the d isplayed size of the d igital channels / 109To switch a single channel on or off / 109To switch all d igital channels on or off / 109To switch groups of channels on or off / 110To change the logic threshold for d igital channels / 110To reposition a d igital channel / 111To display digital channels as a bus / 111 Keysight InfiniiVision 3000 X-Series Oscilloscopes user 's Guide11 Digital channel signal fidelity: Probe impedance and ground ing / 114 Input Impedance / 114 Probe Ground ing / 116 Best Probing Practices /