NI cDAQ -9178

SPECIFICATIONSNI cDAQ -9178NI CompactDAQ Eight-Slot USB ChassisThese specifications are for the national instruments CompactDAQ 9178 chassis only. Thesespecifications are typical at 25 C unless otherwise noted. For the C Series I/O modulespecifications, refer to the documentation for the C Series I/O module you are FIFO size127 samples per sample rate1 Determined by the C Series I/O module accuracy250 ppm of sample of channels supportedDetermined by the C Series I/O module ormodulesAnalog OutputNumber of channels supportedHardware-timed by the C Series I/O module ormodules1 Performance dependent on type of installed C Series I/O module and number ofchannels in the not include group delay.

SPECIFICATIONS NI cDAQ™-9178 NI CompactDAQ Eight-Slot USB Chassis These specifications are for the National Instruments CompactDAQ 9178 chassis only. These

Tags:

  National, Instruments, 1987, National instruments

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Transcription of NI cDAQ -9178

1 SPECIFICATIONSNI cDAQ -9178NI CompactDAQ Eight-Slot USB ChassisThese specifications are for the national instruments CompactDAQ 9178 chassis only. Thesespecifications are typical at 25 C unless otherwise noted. For the C Series I/O modulespecifications, refer to the documentation for the C Series I/O module you are FIFO size127 samples per sample rate1 Determined by the C Series I/O module accuracy250 ppm of sample of channels supportedDetermined by the C Series I/O module ormodulesAnalog OutputNumber of channels supportedHardware-timed by the C Series I/O module ormodules1 Performance dependent on type of installed C Series I/O module and number ofchannels in the not include group delay.

2 For more information, refer to the documentation foreach C Series I/O not include group delay. For more information, refer to the documentation foreach C Series I/O taskDetermined by the C Series I/O module ormodulesMaximum update MS/s (multi-channel, aggregate)..Non-regenerationDetermined by the C Series I/O module accuracy50 ppm of sample nsOutput FIFO regeneration8,191 samples shared among channels samples per waveform modesNon-periodic waveform, periodic waveformregeneration mode from onboard memory,periodic waveform regeneration from hostbuffer including dynamic updateDigital Waveform acquisition (DI) FIFO127 samples per slotWaveform generation (DO) 1 to 42,047 5 to 81,023 samplesNote When modules are installed in slots 1 through 4, FIFO is 2,047 samples perslot for all slots.

3 When any module is installed in slots 5 through 8, FIFO is1,023 samples per slot for all eight input sample clock to to 10 MHzDigital output sample clock from from FIFO0 to 10 to 10 MHz2 | NI cDAQ -9178 Specifications | accuracy50 ppmGeneral-Purpose of measurementsEdge counting, pulse, semi-period, period,two-edge separation, pulse measurementsX1, X2, X4 quadrature encoding withChannel Z reloading; two-pulse applicationsPulse, pulse train with dynamic updates,frequency division, equivalent time base clocks80 MHz, 20 MHz, 100 base clock frequency0 to 20 clock accuracy50 frequency0 to 20 , Source, HW_Arm, Aux, A, B, Z, options for inputsAny module PFI, chassis PFI BNC, analogtrigger, many internal 127-sample FIFOF requency of clocks20 MHz, 10 MHz, 100 to 16 (integers).

4 Base clock accuracy50 chassis PFI BNC or module PFI terminalNI cDAQ -9178 Specifications | national instruments | 3 Module PFI digital input, static digital output, timinginput, and timing output sources4 Many analog input, analog output, counter,digital input, and digital output timing input frequency0 to 20 output frequency0 to 20 MHzChassis PFI input or output frequency1 length3 m (10 ft)..Cable impedance50 ..TRIG 0 (PFI 0), TRIG 1 (PFI 1) stateHigh impedanceTable 1. Input/Output Voltage ProtectionVoltageMinimumMaximum Input-20 V25 VOutput-15 V20 VMaximum operating output low current8 mA output high current-8 mA maximum4 Actual signals available dependent on type of installed C Series I/O beyond those listed under Maximum operating conditions may cause permanent damage tothe | NI cDAQ -9178 Specifications | 2.

5 DC Input CharacteristicsVoltageMinimumMaximum Positive going VNegative going VTable 3. DC Output CharacteristicsVoltageConditionsMinimumM aximum VSourcing 100 V-Sourcing 2 V-Sourcing V-LowSinking 100 VSinking 2 VSinking VDigital chassis PFI BNC or module PFI for most input functionStart Trigger, Reference Trigger, PauseTrigger, Sample Clock, Sample output functionStart Trigger, Pause Trigger, Sample Clock,Sample Clock functionGate, Source, HW_Arm, Aux, A, B, Z,Up_DownNI cDAQ -9178 Specifications | national instruments | 5 Module I/O power-onModule-dependent. Refer to the documentationfor each C Series I/O The chassis may revert the input/output of the modules to their power-onstate when the USB cable is RequirementsCaution You must use a national Electric Code (NEC) Class 2 power source withthe NI cDAQ -9178 Some C Series I/O modules have additional power requirements.

6 For moreinformation about C Series I/O module power requirements, refer to thedocumentation for each C Series I/O Sleep mode for C Series I/O modules is not supported in the NI voltage range9 to 30 required input power615 input connector2 positions mm pitch pluggable screwterminal with screw locks similar to input mating connectorSauro CTF020V8, Phoenix Contact 1714977,or consumption from USB, to V500 A maximum6 Includes maximum 1 W module load per slot across rated temperature and | NI cDAQ -9178 Specifications | specificationUSB data stream types availableAnalog input, analog output, digital input,digital output, counter/timer input,counter/timer output, NI-XNET7 Note If you are connecting the NI cDAQ -9178 chassis to a USB hub, the hub mustbe externally (unloaded)Approx.

7 878 g ( oz)..Dimensions (unloaded) mm mm mm(10 in. x in. in.) Refer to thefollowing you need to clean the chassis, wipe it with a dry a session is active, CAN or LIN (NI-XNET) C Series modules use a total of two data streamsregardless of the number of NI-XNET modules in the cDAQ -9178 Specifications | national instruments | 7 Figure 1. NI cDAQ -9178 mm( in.) mm( in.)INPUT9-30 V 15 W MAX POWERREADYACTIVEVCTRIG 0 TRIG mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) mm( in.)NI cDAQ-9178NI mm( in.) mm( in.) mm( in.) mm( in.) mm( in.) temperature8-20 C to 55 C (IEC-60068-2-1 andIEC-60068-2-2).

8 Storage temperature-40 C to 85 C (IEC-600068-2-1 andIEC-60068-2-2)..Ingress protectionIP 308 When operating the NI cDAQ -9178 in temperatures below 0 C, you must use thePS-15 power supply or another power supply rated for below 0 | NI cDAQ -9178 Specifications | humidity10 to 90% RH, noncondensing(IEC-60068-2-56)..Storage humidity5 to 95% RH, noncondensing(IEC-60068-2-56)..Pollution Degree (IEC 60664) altitude5,000 mIndoor use and VibrationTo meet these specifications, you must panel mount the NI cDAQ -9178 system, use anNI locking USB cable, and affix ferrules to the ends of the terminal shock30 g peak, half-sine, 11 ms pulse (Tested inaccordance with IEC 60068-2-27.)

9 Test profiledeveloped in accordance withMIL-PRF-28800F.)Random to 500 Hz, to 500 Hz, grms (Tested in accordancewith IEC 60068-2-64. Non-operating testprofile exceeds the requirements ofMIL PRF-28800F, Class 3.)SafetyThis product meets the requirements of the following standards of safety for electricalequipment for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 Note For UL and other safety certifications, refer to the product label or the OnlineProduct Certification cDAQ -9178 Specifications | national instruments | 9 Electromagnetic CompatibilityThis product meets the requirements of the following EMC standards for electrical equipmentfor measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions.

10 Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions AS/NZS CISPR 11: Group 1, Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissionsNote In the United States (per FCC 47 CFR), Class A equipment is intended foruse in commercial, light-industrial, and heavy-industrial locations. In Europe,Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intendedfor use only in heavy-industrial Group 1 equipment (per CISPR 11) is any industrial, scientific, or medicalequipment that does not intentionally generate radio frequency energy for thetreatment of material or inspection/analysis For EMC declarations and certifications, refer to the Online ProductCertification ComplianceThis product meets the essential requirements of applicable European Directives as follows: 2006/95/EC; Low-Voltage Directive (safety) 2004/108/EC.