Example: biology

MaxiMintegrated Data Sheet

General DescriptionThe MAX30102 is an integrated pulse oximetry and heart-rate monitor module. It includes internal LEDs, photodetectors, optical elements, and low- noise electronics with ambient light rejection. The MAX30102 provides a complete system solution to ease the design-in process for mobile and wearable MAX30102 operates on a single power supply and a separate power supply for the internal LEDs. Communication is through a standard I2C-compatible interface. The module can be shut down through software with zero standby current, allowing the power rails to remain powered at all Wearable Devices Fitness Assistant Devices Smartphones TabletsBenefits and Features Heart-Rate Monitor and Pulse Oximeter Sensor in LED Reflective Solution Tiny x x 14-Pin optical Module Integrated Cover Glass for Optimal, Robust Performance Ultra-Low Power Operation for Mobile Devices Programmable Sample Rate and LED Current for Power Savings Low-Power Heart-Rate Monitor (< 1mW) Ultra-Low Shutdown Current ( A, typ) Fast data Output Capability high Sample Rates Robust Motion Artifact Resilience high SNR -40 C to +85 C Operating Temperature RangeOrdering Information appears at end of data ; Rev 1; 10/18 PHOTO DIODELED DRIVERSRED/IR LEDPACKAGINGHUMAN SUBJECTELECTRICALOPTICALAMBIENT LIGHT18-BIT CURRENT ADCAMBIENT LIGHT CANCELLATIONDIGITAL noise CANCELLATIONDATA F

High SNR -40°C to +85°C Operating Temperature Range Ordering Information appears at end of data sheet. 19-7740; Rev 1; 10/18 PHOTO DIODE LED DRIVERS RED/IR LED PACKAGING HUMAN SUBJECT ELECTRICAL OPTICAL AMBIENT LIGHT 18-BIT CURRENT ADC AMBIENT LIGHT CANCELLATION DIGITAL NOISE CANCELLATION DATA FIFO HOST (AP) DRIVER …

Tags:

  High, Optical, Data, Noise

Information

Domain:

Source:

Link to this page:

Please notify us if you found a problem with this document:

Other abuse

Advertisement

Transcription of MaxiMintegrated Data Sheet

1 General DescriptionThe MAX30102 is an integrated pulse oximetry and heart-rate monitor module. It includes internal LEDs, photodetectors, optical elements, and low- noise electronics with ambient light rejection. The MAX30102 provides a complete system solution to ease the design-in process for mobile and wearable MAX30102 operates on a single power supply and a separate power supply for the internal LEDs. Communication is through a standard I2C-compatible interface. The module can be shut down through software with zero standby current, allowing the power rails to remain powered at all Wearable Devices Fitness Assistant Devices Smartphones TabletsBenefits and Features Heart-Rate Monitor and Pulse Oximeter Sensor in LED Reflective Solution Tiny x x 14-Pin optical Module Integrated Cover Glass for Optimal, Robust Performance Ultra-Low Power Operation for Mobile Devices Programmable Sample Rate and LED Current for Power Savings Low-Power Heart-Rate Monitor (< 1mW) Ultra-Low Shutdown Current ( A, typ) Fast data Output Capability high Sample Rates Robust Motion Artifact Resilience high SNR -40 C to +85 C Operating Temperature RangeOrdering Information appears at end of data ; Rev 1.

2 10/18 PHOTO DIODELED DRIVERSRED/IR LEDPACKAGINGHUMAN SUBJECTELECTRICALOPTICALAMBIENT LIGHT18-BIT CURRENT ADCAMBIENT LIGHT CANCELLATIONDIGITAL noise CANCELLATIONDATA FIFOHOST (AP) I2 CDRIVERHARDWARE FRAMEWORKAPPLICATIONSCOVER GLASSMAX30102 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthSystem DiagramClick here for production status of specific part to GND .. to + to PGND .. to + + to to + Other Pins to GND .. to + Output Short-Circuit Current Duration ..ContinuousContinuous Input Current into Any Terminal .. 20mA ESD, Human Body Model (HBM) .. Immunity .. 250mAContinuous Power Dissipation (TA = +70 C) OESIP (derate C above +70 C) ..440mWOperating Temperature Range .. -40 C to +85 CJunction Temperature ..+90 CSoldering Temperature (reflow) ..+260 C Storage Temperature Range ..-40 C to +105 C (VDD = , VLED+ = , TA = -40 C to +85 C, unless otherwise noted. Typical values are at TA = +25 C) (Note 1)PARAMETERSYMBOLCONDITIONSMINTYPMAXUNIT SPOWER SUPPLYP ower-Supply VoltageVDDG uaranteed by RED and IR count Supply Voltage VLED+ to PGNDVLED+Guaranteed by PSRR of LED CurrentIDDSpO2 and HR mode, PW = 215 s, 50sps6001200 AIR only mode, PW = 215 S, 50sps6001200 Supply Current in ShutdownISHDNTA = +25 C, MODE = Integrated 2 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthPackage thermal resistances were obtained using the method described in JEDEC specification JESD51-7, using a four-layer board.

3 For detailed information on package thermal considerations, refer to beyond those listed under Absolute Maximum Ratings may cause permanent damage to the device. These are stress ratings only; functional operation of the device at these or any other conditions beyond those indicated in the operational sections of the specifications is not implied. Exposure to absolute maximum rating conditions for extended periods may affect device Maximum RatingsElectrical CharacteristicsFor the latest package outline information and land patterns (footprints), go to Note that a + , # , or - in the package code indicates RoHS status only. Package drawings may show a different suffix character, but the drawing pertains to the package regardless of RoHS InformationPACKAGE TYPE: 14 OESIPP ackage Code F143A5MK+1 Outline Number21-1048 Land Pattern Number90-0602 THERMAL RESISTANCE, FOUR-LAYER BOARDJ unction to Ambient ( JA)180 C/WJunction to Case ( JC) 150 C/W(VDD = , VLED+ = , TA = -40 C to +85 C, unless otherwise noted.)

4 Typical values are at TA = +25 C) (Note 1)PARAMETERSYMBOLCONDITIONSMINTYPMAXUNIT SPULSE OXIMETRY/HEART-RATE SENSOR CHARACTERISTICSADC Resolution18bitsRed ADC Count(Note 2)REDCLED1_PA = 0x0C, LED_PW = 0x01, SPO2_SR = 0x05,ADC_RGE = 0x0065536 CountsIR ADC Count(Note 2)IRCLED2_PA = 0x0C, LED_PW = 0x01, SPO2_SR = 0x05 ADC_RGE = 0x0065536 CountsDark Current CountLED_DCCLED1_PA = LED2_PA = 0x00,LED_PW = 0x03, SPO2_SR = 0x01 ADC_RGE = of FSDC Ambient Light RejectionALRADC counts with finger on sensor under direct sunlight (100K lux), ADC_RGE = 0x3, LED_PW = 0x03, SPO2_SR = 0x01 Red LED2 CountsIR LED2 CountsADC Count PSRR (VDD) < VDD < ,LED_PW = 0x01, SPO2_SR = of FSFrequency = DC to 100kHz, 100mVP-P10 LSBADC Count PSRR(LED Driver Outputs) < VLED+, < , LED1_PA = LED2_PA = 0x0C, LED_PW = 0x01, SPO2_SR = of FSFrequency = DC to 100kHz, 100mVP-P10 LSBADC Clock Integration TimeINTLED_PW = 0x0069 sLED_PW = 0x01118 LED_PW = 0x02215 LED_PW = 0x03411 Slot Timing (Timing Between Sequential Channel Samples; , Red Pulse Rising Edge To IR Pulse Rising Edge)INTLED_PW = sLED_PW = = = GLASS CHARACTERISTICS (Note 3)Hydrolytic Resistance ClassPer DIN ISO 719 HGB Integrated 3 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthElectrical Characteristics (continued)(VDD = , VLED+ = , TA = -40 C to +85 C, unless otherwise noted.

5 Typical values are at TA = +25 C) (Note 1)PARAMETERSYMBOLCONDITIONSMINTYPMAXUNIT SIR LED CHARACTERISTICS (Note 3)LED Peak Wavelength PILED = 20mA, TA = +25 C870880900nmFull Width at Half Max ILED = 20mA, TA = +25 C30nmForward VoltageVFILED = 20mA, TA = +25 PowerPOILED = 20mA, TA = +25 LED CHARACTERISTICS (Note 3)LED Peak Wavelength PILED = 20mA, TA = +25 C650660670nmFull Width at Half Max ILED = 20mA, TA = +25 C20nmForward VoltageVFILED = 20mA, TA = +25 PowerPOILED = 20mA, TA = +25 CHARACTERISTICS (Note 3)Spectral Range of Sensitivity (QE > 50%)QE: Quantum Efficiency600900nmRadiant Sensitive of Radiant Sensitive AreaL x x x mmINTERNAL DIE TEMPERATURE SENSORT emperature ADC Acquisition TimeTTTA = +25 C29msTemperature Sensor AccuracyTATA = +25 C 1 CTemperature Sensor Minimum RangeTMIN-40 CTemperature Sensor Maximum RangeTMAX85 CDIGITAL INPUT CHARACTERISTICS: SCL, SDAI nput high VoltageVIHVDD = x VDDVI nput Low VoltageVILVDD = x VDDVH ysteresis Leakage CurrentIINVIN = GND or VDD (STATIC) 1 ADIGITAL OUTPUT CHARACTERISTICS: SDA, INTO uput Low VoltageVOLISINK = Integrated 4 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthElectrical Characteristics (continued)(VDD = , VLED+ = , TA = -40 C to +85 C, unless otherwise noted.

6 Typical values are at TA = +25 C) (Note 1)Note 1: All devices are 100% production tested at TA = +25 C. Specifications over temperature limits are guaranteed by Maxim Integrated s bench or proprietary automated test equipment (ATE) 2: Specifications are guaranteed by Maxim Integrated s bench characterization and by 100% production test using proprietary ATE setup and 3: Guaranteed by design and characterization. Not tested in final 1. I2C-Compatible Interface Timing DiagramPARAMETERSYMBOLCONDITIONSMINTYPMA XUNITSI2C TIMING CHARACTERISTICS (SDA, SDA, INT) (Note 3)I2C Write AddressAEHexI2C Read AddressAFHexSerial Clock FrequencyfSCL0400kHzBus Free Time Between STOP and START sHold Time (Repeated) START ConditiontHD; sSCL Pulse-Width sSCL Pulse-Width sSetup Time for a Repeated START ConditiontSU; sData Hold TimetHD;DAT0900nsData Setup TimetSU;DAT100nsSetup Time for STOP ConditiontSU; sPulse Width of Suppressed SpiketSP050nsBus CapacitanceCB400pFSDA and SCL Receiving Rise TimetR20 + and SCL Receiving Fall TimetRF20 + Transmitting Fall TimetTF300nsSDASCLtHD,STASTART CONDITIONtRtFtLOWtSU,DATtHD,DATtSU,STAtH D,STAREPEATED START CONDITIONtSPtSU, Integrated 5 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthElectrical Characteristics (continued)(VDD = , VLED+ = , TA = +25 C, RST, unless otherwise noted.

7 0102030405060012345IR LED CURRENT (mA)VLEDVOLTAGE (V)IR LED SUPPLY HEADROOMtoc02 ILED= 50mAILED= CURRENT (mA)SUPPLY VOLTAGE (V)VDDSUPPLY CURRENT VOLTAGEtoc 03 SHUTDOWNMODENORMALOPERATION0500010000150 0020000250003000035000400004500050000051 01520 COUNTS(SUM)DISTANCE (mm)DCCOUNTS vs. DISTANCE FORWHITE high -IMPACT STYRENE CARDtoc 04 REDIRMODE = SPO2 andHRADCRES= 18 BITsADCSR = 100 SPSADCFULLSCALE= CURRENT (uA)TEMPERATURE ( C)VDDSHUTDOWN CURRENTvs. TEMPERATUREtoc05 VDD CURRENT ( A)TEMPERATURE ( C)VLEDSHUTDOWN CURRENTvs. TEMPERATUREVLED = toc06 VLED = LED CURRENT (mA)VLEDVOLTAGE (V)RED LED SUPPLY HEADROOMtoc01 ILED= 50mAILED= 20mA-20020406080100120500600700800 NORMALIZED POWER (%)WAVELENGTH(nm)RED LED SPECTRUMAT TA= +30 Ctoc 080204060801001207008009001000 NORMALIZED POWER (%)WAVELENGTH(nm)IR LED SPECTRUMAT TA= +30 Ctoc 09 Maxim Integrated Pulse Oximeter and Heart-Rate Sensor for Wearable HealthTypical Operating Characteristics(VDD = , VLED+ = , TA = +25 C, RST, unless otherwise noted.)

8 860870880890900910-50050100150 PEAK WAVELENGTH (nm)TEMPERATURE (deg C)IR LED PEAK WAVELENGTHvs. TEMPERATUREtoc11 LED CURRENT (mA)FORWARD VOLTAGE (V)RED LED FORWARD AT TA= +25 Ctoc12 MODE = FLEX LEDADC RES = 18 BITSADC SR = 100 SPSADC FULL SCALE = CURRENT (mA)FORWARD VOLTAGE (V)IR LED FORWARD AT TA= +25 Ctoc13 MODE = FLEX LEDADC RES = 18 BITSADC SR = 100 SPSADC FULL SCALE = 2048nA-70-60-50-40-30-20-100101001000100 00100000 MAGNITUDE (dB)FREQUENCY (Hz)PW=69 sPW= 118 sPW= 215 sPW= 411 sAMBIENT LIGHT CANCELLATION PASSBAND CHARACTERISTICStoc EFFICIENCYWAVELENGTH (nm)PHOTODIODE QUANTUM EFFICIENCY vs. WAVELENGTHtoc15645650655660665670675-500 50100150 PEAK WAVELENGTH (nm)TEMPERATURE ( C)RED LED PEAK WAVELENGTHvs. TEMPERATUREtoc10 LED CURRENT:10mA20mA30mA50mAMODE = FLEX LEDADC RES = 18 BITSADC SR = 400 SPSADC FULL SCALE = 2048nAMaxim Integrated Pulse Oximeter and Heart-Rate Sensor for Wearable HealthTypical Operating Characteristics (continued)PINNAMEFUNCTION1, 5, 6, 7, 8, Connection.

9 Connect to PCB pad for mechanical Clock Input3 SDAI2C data , Bidirectional (Open-Drain)4 PGNDP ower Ground of the LED Driver Blocks9 VLED+LED Power Supply (anode connection). Use a bypass capacitor to PGND for best +11 VDDA nalog Power Supply Input. Use a bypass capacitor to GND for best Ground 13 INTA ctive-Low Interrupt (Open-Drain). Connect to an external voltage with a pullup +9 VLED+ Integrated 8 MAX30102 high -Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable HealthPin DescriptionPin ConfigurationDetailed DescriptionThe MAX30102 is a complete pulse oximetry and heart-rate sensor system solution module designed for the demanding requirements of wearable devices. The device maintains a very small solution size without sacrificing optical or electrical performance. Minimal external hardware components are required for integration into a wearable MAX30102 is fully adjustable through software registers, and the digital output data can be stored in a 32-deep FIFO within the IC.

10 The FIFO allows the MAX30102 to be connected to a microcontroller or processor on a shared bus, where the data is not being read continuously from the MAX30102 s SubsystemThe SpO2 subsystem of the MAX30102 contains ambient light cancellation (ALC), a continuous-time sigma-delta ADC, and a proprietary discrete time filter. The ALC has an internal Track/Hold circuit to cancel ambient light and increase the effective dynamic range. The SpO2 ADC has programmable full-scale ranges from 2 A to 16 A. The ALC can cancel up to 200 A of ambient internal ADC is a continuous time oversampling sigma-delta converter with 18-bit resolution. The ADC sampling rate is The ADC output data rate can be programmed from 50sps (samples per second) to SensorThe MAX30102 has an on-chip temperature sensor for calibrating the temperature dependence of the SpO2 sub-system. The temperature sensor has an inherent resolu-tion of device output data is relatively insensitive to the wavelength of the IR LED, where the Red LED s wave-length is critical to correct interpretation of the data .